DOI: 10.19102/icrm.2016.070502
HERMINE R. POGHOSYAN, MD1,2, JOHN V. WYLIE, MD, FACC1 and MICHAEL V. ORLOV, MD1,3
1St. Elizabeth’s Medical Center of Boston, MA
2Astghik Medical Center, Yerevan, Armenia
3The Miriam and Rhode Island Hospitals, Brown University School of Medicine, Providence, RI
Download PDF |
|
ABSTRACT. A 54-year-old male with ischemic cardiomyopathy and cardiac resynchronization therapy defibrillator implantation (St. Jude Medical, St. Paul, MN) 6 months prior to presentation and with a history of ventricular tachycardia (VT) storm suddenly received multiple implantable cardioverter-defibrillator (ICD) shocks. Device interrogation showed a normally functioning biventricular ICD. The initial supraventricular tachycardia appears to be a sinus tachycardia. A premature ventricular complex during tachycardia and the atrioventricular (AV) Interval Delta parameter value of 60 ms led to a diagnosis of VT resulting in inappropriate therapies. The AV Interval Delta is helpful for discrimination during dual tachycardia, but utilization of every discriminator algorithm in a particular patient should be adapted to the specific arrhythmic substrate of the patient.
KEYWORDS. AV Interval Delta, cardiac resynchronization therapy defibrillator, discriminator algorithm, post-shock electrograms (EGM) changes, Quadra Assura.
The authors have no conflicts of interest relevant to this study to disclose.
Manuscript received March 16, 2016, Final version accepted April 19, 2016.
Address correspondence to: Hermine Poghosyan, MD, Astghik Medical Center, 28/1 Daniel Varujan St. 0038 Yeravan, Armenia. E-mail: epdoctorh@gmail.com
A 54-year-old man with ischemic cardiomyopathy was implanted with a cardiac resynchronization therapy defibrillator (CRT-D, Quadra Assura, St. Jude Medical, IESD – Implantable Electronic Systems Division, Sylmar, CA) for secondary prevention of sudden cardiac death. The patient had a history of ventricular tachycardia (VT) storm and acute amiodarone toxicity. He previously had three ablations of recurrent symptomatic monomophic VT. The most recent ablation, 4 months prior to presentation, was effective with no recurrent VT when off antiarrhythmic medications. The patient dramatically improved and his device remained programmed in DDI pacing mode because of suspicion for proarrhythmia related to left ventricular pacing and resolution of heart failure symptoms. He reported mild palpitations and suddenly received multiple implantable cardioverter-defibrillator (ICD) shocks while climbing stairs one evening. He had not taken his daily β-blocker (300 mg/day metoprolol), and the last dose had been taken 36 hours previously. There were 14 shocks delivered over 30 minutes (Figure 1).
Figure 1: Programmed supraventricular tachycardia/ventricular tachycardia (SVT/VT) discriminators and therapy. Notably, in the V = A field, the “If All” criterion is on. If the rhythm is 1:1 or V = A, but the difference between the AV intervals is greater than the setting for the AV Interval Delta parameter (in this case 60 ms), then the rhythm is diagnosed as VT and additional discrimination using the Morphology and Sudden Onset/Chamber Onset criteria are not evaluated. If the difference is less than the AV Interval Delta setting, the rhythm is analyzed using these criteria further for differentiation of VT/SVT. If AV Interval Delta = OFF, then Morphology and Sudden Onset/Chamber Onset are always evaluated. |
Device interrogation showed a normally functioning biventricular ICD programmed to DDI mode at a rate of 60 bpm with an AV interval of 350 ms and three tachycardia zones: VF zone (222 bpm) with antitachycardia pacing during charging, VT-2 zone (169 bpm), and a VT-1 zone (123 bpm) with supraventricular tachycardia (SVT) discriminators ON. The slowest VT zone was used due to the presence of slow VT prior to the last ablation. The underlying rhythm was sinus rhythm with firstdegree atrioventricular (AV) block and no retrograde ventriculoatrial conduction (VA). Programmed SVT/VT discriminators and therapy settings are seen in Figure 1. Review of the arrhythmia log showed episodes of tachycardia falling into the VT-1 zone at rates ranging from 127 to 136 bpm (Figure 2a). There was a 1:1 AV/VA relationship, and the tachyarrhythmia was diagnosed as SVT by the Rate Branch algorithm (V=A); Onset and Morphology Criteria (before the first therapy) and therapy was withheld. Subsequently, because of programming of the AV Interval Delta, the patient received multiple VT therapies resulting in a change in EGM morphology (Figure 2a,b). The change in morphology led to further therapies because the Morphology Criteria diagnosed VT. Therapies did not affect the tachycardia, which eventually slowed down below the rate detection.
Figure 2: (a) Ventricular tachycardia (VT) detected by the atrioventricular (AV) interval algorithm. The initial far field EGM morphology was the same as sinus rhythm. In the EGM strip the AV Interval Delta discriminator forces VT diagnosis. Redetection of supraventricular tachycardia occurred until critically timed premature ventricular contraction (PVC) caused short AV intervals of 98 ms and 100 ms. The 16 AV intervals prior to redetection are evaluated. The difference between the second longest AV interval (160 ms) and second shortest AV interval (100 ms) is 60 ms, which is the threshold for VT diagnosis. PVCs are indicated by the red boxes. (b) Changing of far field and local EGM morphology following shock delivery during this clinical event. |
The annual rate of inappropriate shocks has fallen from 37–50% for SVT alone in early studies to 1–5% for all causes in modern clinical trials. Most trial data concern the programming for primary prevention (e.g. PREPARE, MADIT-RIT).1–4 There are only a few trials (e.g. the subset of the ADVANCE III study) that have evaluated the programming approach in secondary prevention patients.4 According to the recent HRS Consensus Statement for secondary prevention patients, discrimination algorithms should be programmed to include rhythms with rates faster than 200 bpm and potentially up to 230 bpm (unless contraindicated).1 The ventricular rate alone is a mandatory discriminator, and all other discriminators can be considered individually. Each of SVT discriminators may have limitations that should be carefully understood when programming ICD therapies, especially for lower treatment zones.
Several interval-based discriminators focus on differences in sinus tachycardia. Sudden onset was one of the first single-chamber, interval-based discriminators. Onset discriminators classify the rhythm only once, and thus cannot correct misclassifications. Current devices provide algorithms based on the PR–RP interval relationship to avoid inappropriate therapy. Consistent PR patterns usually indicate SVT. Each manufacturer provides their own building blocks for discrimination of tachycardias with the atrial rate equal to ventricular rate. Medtronic provides PR Logic (Medtronic Inc., St. Paul, MN), Biotronik provides SMART Detection® algorithm (Biotronik, Berlin, Germany), and St. Jude Medical provides AV Rate Branch + (St. Jude Medical, St. Paul, MN), etc. In the Medtronic and Guidant algorithms, incremental addition of discriminators increases specificity and decreases sensitivity of classification. In the St. Jude Medical ICDs, discriminators may be combined using either the “ANY“ or “ALL” operators. Measurement of AV intervals is part of the AV Rate Branch + building block in St. Jude Medical devices (AV Detection Enhance™, St. Paul, MN). This algorithm counts the AV intervals before detection (up to 16 AV intervals if in VT Detection) and calculates the difference between the second longest and second shortest AV interval. If the rhythm is 1:1 or V = A, but the difference between the AV intervals is greater than the setting for the AV Interval Delta parameter (in this case 60 ms), then the rhythm is diagnosed as VT and additional discrimination using the Morphology and Sudden Onset/Chamber Onset criteria are not evaluated.5 If the difference is less than the AV Interval Delta setting, the rhythm is analyzed using these criteria further for differentiation of VT/SVT. If AV Interval Delta = OFF, then Morphology and Sudden Onset/Chamber Onset are always evaluated (Figure 3). In this case the initial SVT appears to be sinus tachycardia. The AV Interval Delta parameter forced a VT diagnosis; therefore, inappropriate therapy was delivered. The AV Interval Delta is intended to facilitate identification of a dual tachycardia with isorhythmic AV dissociation by detecting subtle changes in the PR intervals. The Morphology Discrimination was not a party to this. In this case, a relatively late cycle premature ventricular contraction (PVC), occurring after the native P wave during the tachycardia caused PR interval irregularity, resulted in VT detection with subsequent therapy (Figure 4).
Figure 3: Rate Branch/AV Interval Delta in these episodes.2 |
Figure 4: Diagnosis summary of the event shows that the AV Interval Delta and Morphology algorithm indicated ventricular tachycardia (VT), and even Onset indicated supraventricular tachycardia (SVT); therapy in VT-1 zone was delivered. When the AV Interval Delta exceeded the programmed value and a diagnosis of VT was made, the further use of SVT discriminators was automatically discontinued by the device. |
This case represents a situation where the programmed AV Interval Delta resulted in inappropriate ICD shocks after a PVC altered the AV interval. After resumption of b-blocker therapy along with deactivating the AV Interval Delta discrimination feature, and increasing the lower detection rate of the VT-1 zone, no further arrhythmia or device therapies were noted during follow-up.
What can cause morphology changes in the situation in Figure 2b?
No discriminator algorithm functions with 100% sensitivity and specify. The discriminator algorithms utilized in a particular patient should be adapted to the specific arrhythmic substrate of the patient. Each of the SVT discriminators have limitations that should be carefully understood when programming ICD therapies, especially for lower treatment zones. Detailed knowledge of the patient’s arrhythmia history along with the subtleties of specific ICD algorithms may help in individualization of the device programming. When tachycardias are recorded by the device and after therapies are delivered, careful review of stored episodes should be performed to further tailor programmed settings for discrimination of SVT and VT.1
|