Journal of Innovation in Cardiac Rhythm Management
Articles Articles 2011 August

The Use of Cryoballoon Technology in Catheter Ablation of Atrial Fibrillation: State of the Art

DOI: 10.19102/icrm.2011.020 802

FRANCESCO PERNA, MD and MOUSSA MANSOUR, MD

Heart Center, Massachusetts General Hospital, Boston, MA

PDF Download PDF
tweeter Follow Us >>

ABSTRACT.Pulmonary vein (PV) isolation using radiofrequency (RF) energy has been recognized as the cornerstone of catheter ablation for atrial fibrillation (AF). However, potentially life-threatening complications can occur during RF ablation of AF. Cryoballoon PV isolation has been recently introduced as an alternative to RF ablation. This treatment modality has been shown to be effective in the treatment of paroxysmal AF, but its usefulness in the management of persistent AF still needs further investigation. Phrenic nerve paralysis is the most frequent complication observed with the use of this technology, but it is fortunately reversible in the majority of cases.

KEYWORDS.catheter ablation of atrial fibrillation, cryoablation, cryoballoon, cryoenergy, cryothermal energy, cryothermia.

Dr. Mansour reports he is a consultant of Medtronic, Inc.
Manuscript received June 8, 2011, final version accepted June 20, 2011.
Address correspondence to: Moussa Mansour, MD, 55 Fruit Street GRB-109, Boston, MA 02114. E-mail: mmansour@partners.org

Introduction

Catheter ablation has become one of the primary treatments of symptomatic drug-refractory atrial fibrillation (AF).1,2 Although ablation techniques vary among centers, it is generally accepted that pulmonary vein isolation (PVI) represents the cornerstone of catheter ablation of AF.37 Achieving PVI with focal radiofrequency (RF) catheters is a time-consuming and technically challenging process which requires a high level of expertise. Moreover, this treatment modality is associated with a complication rate of 2–5% according to some studies.811 As a result, novel tools and techniques, including various balloon ablation catheters using different energy sources, have been introduced into clinical practice with the aim of providing a safer, faster, more standardized, and more effective PVI procedure.1216 So far the greatest clinical experience with balloon ablation technologies has been with the cryoballoon.

Principles of cryothermal energy

RF tissue heating results in tissue destruction, an intense inflammatory response, and the risk of thrombus formation requiring anticoagulation during and after the procedure to prevent stroke. Moreover, delivery of RF lesions within venous structures such as the coronary sinus or pulmonary veins (PVs) is associated with a risk of venous stenosis, endoluminal thrombosis, and perforation, possibly leading to tamponade or damage to adjacent arterial structures.17 Cryothermal energy is an alternative energy source that has been developed to overcome some of the disadvantages of RF ablation such as tissue disruption by excessive heating and generation of inhomogeneous lesions.18 The immediate cause of injury is the deleterious effect of freezing and rewarming cycles on the cells. The delayed effects consist of progressive failure of the microcirculation and ultimately vascular stasis. Previous studies imply that cryosurgery also triggers apoptosis.19 Cryolesions are sharply demarcated and the margins are infiltrated by fibrin, collagen stranding, and capillary growth.20 The final phase in the evolution of a stable cryolesion is seen within 2–4 weeks. All this results in preservation of tissue architecture and endocardial contours and minimal thrombus formation, thus making cryogenic lesions homogeneous, sharply defined, and with a low arrhythmogenic potential.18,21,22

Pulmonary vein isolation using cryoballoon technology

Features of the cryoballoon strategy

Cryoablation has some potential advantages for AF ablation. This ablation energy has a lower thrombogenicity and has never been associated with atrioesophageal fistula (AEF), even though this complication is theoretically possible since esophageal lesions have been detected in animal studies using cryoablation.22,23 Catheter point-by-point cryoablation is an effective method to create electrical PVI with clinically satisfactory results and no risk of pulmonary vein (PV) stenosis.24 However, procedure times are long, due to the longer individual ablation applications required to produce equivalent lesions, considering that one single ablation point takes about 4 min, making focal cryoablation for AF impractical.

As a result a cryoballoon device (Arctic Front, Medtronic, Cryocath, Quebec, Canada) has been developed which allows cinrcumfererntial PVI.19,20,2527 In contrast to sequential point-by-point ablation using a cryo or an RF focal catheter, the cryoballoon technique allows simultaneous energy delivery at the left atrium (LA)–PV junction circumferentially, often resulting in isolation of the PV with a single cryothermal energy application.27,28

Cryoballoon technology

The CryoConsole (Medtronic) houses the coolant as well as electrical and mechanical. The refrigerant nitrous oxide is delivered into the inner balloon where it undergoes a liquid-to-gas phase change, resulting in an inner balloon cooling temperature of approximately –80°C. A central lumen is used for the insertion of the guidewire and injection of contrast medium for PV angiograms. Using the “over-the-wire” technique in conjunction with the 12-F steerable sheath (FlexCath, Medtronic), the balloon can be navigated to each PV ostium.28

The catheter is available in 23 mm and 28 mm diameter sizes, as appropriate for the diameter of each PV. Preprocedural (computed tomography/magnetic resonance imaging) or intraprocedural (angiography, intracardiac ultrasound) imaging can be helpful to correctly size the balloon to the PV ostia. Many operators however advocate using the 28-mm balloon for most pulmonary veins. After transseptal puncture, the catheter is deployed to the LA. A stiff guidewire is placed deep into a branch of the PV to achieve stability, and the balloon is then positioned in front of the PV ostium. In case of multiple freezes, it is possible to select different branches with the guidewire at every freeze to achieve a better contact of the balloon with the entire contour of the antrum. In order to avoid any mechanical damage, special care must be taken to inflate the balloon outside the PV. After inflation, the catheter is applied to the vein and the optimum position in the PV ostium is usually confirmed by PV angiography. It is important for isolation success to achieve complete occlusion of the vein. It is believed that blood FLow alongside the balloon (peri-balloon leaks) and incomplete vein occlusion account for higher temperatures which have been deemed responsible for reversible lesions.27

Occlusion-monitoring strategies

The minimal temperature achieved during cryoballoon ablation and the slope of rewarming, acquired using a temperature sensor located in the back of the balloon together with the CryoConsole software, are correlated with depth of the lesions, but no precise cut-off value has been correlated with effective PVI.20,26 Transesophageal echocardiography (TEE) allows monitoring of balloon positioning and PV occlusion via color Doppler. After inflation, the balloon is advanced, aiming for PV flow disappearance as documented by color Doppler. TEE-guided assessment of PV occlusion during cryoballoon ablation has been proven feasible, effective and able to predict effective PV electrical isolation with a positive predictive value of 98%.29 However, performing TEE-guided cryoballoon PVI requires additional well-trained personnel and the use of general anesthesia. Siklody and colleagues showed that occlusion and electrical isolation of the PVs during cryoballoon ablation can be predicted by a PV wedge pressure curve recorded at the tip of the catheter;30 one limitation of this technique is that no pressure curve can be registered beneath 0°C, so that exclusive pressure monitoring would fail to detect late balloon dislocations occurring during cryoablation. Intracardiac echography (ICE) can be very helpful for adequate balloon sizing and might confirm occlusion of the PV not only before the initial freeze, but also while performing maneuvers and techniques of repositioning during the freezing period to help optimize contact.31

Ablation techniques with the cryoballoon

After confirmation of PV occlusion, a cryoablation is applied for 240–300 s for each vein. With optimal positioning, it has been demonstrated that electrical isolation can be obtained acutely with the delivery of a single, 240-s lesion,32 but bonus lesions are often performed to ensure a better PVI rate. Conduction recovery after cryoballoon PVI has been observed to occur with a higher incidence at inferior poles of the veins.33 As a result, every effort should be done to obtain good inferior contact. When occlusion is not as desired, the wire can be changed to a different side branch. Flexion of the sheath can also be attempted to ensure better occlusion. Special ablation techniques have been developed for different veins in order to optimize PV occlusion and isolation28 and these are detailed below:

  • the “hockey stick” technique and the “pull-down” technique allow balloon to tissue contact at the inferior PV circumference of the inferior PVs by changing its position before or during the freezing period;
  • the “big loop” technique helps to obtain right inferior PV occlusion if a large inferior gap is present (Figure 1).

crm-02-08-421-f1.jpg

Figure 1: Big loop technique. Right anterior view showing the cryoballoon catheter in the right inferior pulmonary vein.

When PVI is not achievable with a single cryoballoon, a double-balloon strategy (with two different sizes) or a point-to-point cryotip catheter has been used to complete lesions.

Clinical use and results – success rate of the cryoballoon catheter

The acute PVI rate is relatively high with the cryoballoon, often exceeding 95%.28,31,34,35 The main reason for the inability to achieve complete isolation is anatomic features such as oval PV ostia or inferior veins inserting onto the LA with a sharp angulation.

At the present time, cryoballoon ablation of paroxysmal AF has the promise to be an effective approach and appears to have similar success rates as RF ablation after 6 months,28 with midterm success rates ranging from 55% to 86% with respect to AF-free survival.28,30,3537 RF and cryoballoon strategies appeared to be equivalent in terms of practicability in a case–control study conducted by Linhart and colleagues,38 but AF recurrence and AF episode burden were lower after cryoablation (55% versus 45% of freedom from AF during a 6-month follow-up). The success rate was higher for a combination approach (cryoballoon plus point-to-point cryoablation) than for balloon-only ablation. The differences in success rates did not reach statistical significance. In the first published series of 57 patients treated with cryoballoon PVI, freedom from AF after 3 months was reached in 60%.27 Neumann et al achieved maintenance of sinus rhythm in 74% of patients with paroxysmal AF and 42% with persistent AF without antiarrhythmic drug therapy after cryoballoon ablation of 346 patients with a mean follow-up of 12 months.25 In a selected population of patients with paroxysmal AF and normally sized left atria, the 6-month success rate of the cryoballoon AF ablation approach reached 86%.34 The difference in the success rate among different groups might be explained by different followup strategies; if intensive follow-up is performed, the overall success rate of the first PV ablation appears to be quite low.

Cryoballoon PVI is relatively easy to perform and presents a steep learning curve. Klein et al34 observed that fluoroscopy times significantly decreased after only five or six procedures to a mean of about 36 min.

The cryoballoon system has been approved for clinical use in Europe and the United States and has been investigated in two randomized, multicenter trials comparing its safety and efficacy with that of antiarrhythmic drugs and RF ablation, respectively.36,39 In the STOP-AF trial, investigators enrolled 245 patients with paroxysmal AF at 26 centers with the purpose of comparing the cryoballoon ablation technique with antiarrhythmic drug therapy. During a 12-month follow-up, 69.9% of patients treated with cryoablation were free from AF compared with 7.3% of patients treated with antiarrhythmic drugs. Among the successfully ablated patients, 58% were free from AF at 1 year without the use of antiarrhythmic drug therapy, and 60% were free from AF after a single ablation procedure.36 The Freeze AF trial is underway and aims to compare the safety and efficacy of cryoballoon with RF ablation.39

Future studies will have to show whether this new device is also cost-effective as first-line treatment of paroxysmal AF without any prior antiarrhythmic medication and whether cryoballoon ablation is also able to achieve a similar substrate modification compared with RF energy in patients with persistent AF.34

Table 1 summarizes the results of the most important clinical trials investigating cryoballoon ablation of AF.

Table 1: Studies investigating cryoablation of atrial fibrillation

crm-02-08-421-t1.jpg

Complications associated to cryoballoon ablation of atrial fibrillation

Cryothermal energy is known to be safer than other energy sources.22 The most widely known complication of cryoballoon ablation is phrenic nerve palsy (PNP). The occurrence of PNP when using cryoballoon ranges from 2% to 11%.25,27,28,38 It has been seen more frequently with the use of balloon ablation catheters irrespective of the energy source14,35 and with ablation in the right superior PV. High-output phrenic nerve pacing during ablation of the right superior PV can considerably reduce this risk, through early detection of phrenic nerve injury and promptly discontinuing ablation. This kind of complication is usually not life-threatening; moreover, when the palsy had occurred, it predominantly regressed during follow-up.37 The small balloon/PV ostium ratio has been observed to favor phrenic nerve injury. It is believed that the smaller balloon can reach relatively deep into the PV and achieve closer proximity to the phrenic nerve. However, a PNP rate of 11% has been described in a study that investigated PV ablation with a single 28-mm cryoballoon approach.28 These findings indicate that a balloon position deep in the vein in close proximity to the phrenic nerve is responsible for the damage, and that the actual balloon size might only be a subordinated factor. Modification of balloon anatomy would be of some help to avoid inadvertent shift of the balloon into a more distal position.

Despite minor esophageal injuries that have been observed with balloon cryoablation,40 no AEF has occurred to date. In contrast, the incidence of AEF after RF ablation for AF is 0.04%.41 However, the total number of patients who underwent PVI so far using cryoballoon catheter is too small to conclude whether serious esophageal lesions can be excluded from the list of complications following cryoballoon ablation.

Even though PV stenosis had not been observed in earlier studies, the STOP-AF trial showed an incidence of PV stenosis as high as 3.1.36 This risk might be reduced by utilizing a 28-mm balloon, which is more likely to allow antral extra-ostial ablation.

Further improvements to the cryoballoon technique

Combined cryoballoon–RF approach

Clinical success of cryoballoon PVI for treatment of paroxysmal AF has not been achieved in patients with persistent AF. In fact, the clinical success rate in one study was only 39% among patients with persistent AF,25 likely because of the need for additional atrial substrate modification in this subgroup. Extensive substrate modification using focal cryoablation catheters is limited by a prolonged application time and the inability to drag the catheter during ablation due to cryocatheter adherence to tissue. In order to overcome this limitation, a combined approach has been described using a cryoballoon for PVI and a focal RF catheter for LA substrate modification. It was found that this approach is feasible, improves intraprocedural termination of AF compared with cryoballoon PVI alone, and is associated with favorable short- and mid-term maintenance of sinus rhythm (86% arrhythmia-free survival at 6 months) compared with that of cryoballoon PVI alone for persistent AF.42

Real-time recording of electrograms during cryoballoon ablation

Cryoballoon technique requires either two transseptal punctures to accommodate both an ablation catheter and a PV mapping catheter or a single transseptal puncture with the operator interchanging the cryoballoon catheter and mapping catheter. The use of a cryoballoon catheter equipped with a 6-pole micromapping spiral catheter (Promap, ProRhyrthm Inc., Ronkonkoma, NY), inserted through its central lumen for the purpose of mapping and ablation during PVI procedures, has been proven to be both feasible and effective.32,43 The mapping catheter introduced into the central lumen can be used as a guidewire as well as a recording device, thus allowing the real-time recording of PV electrograms and making possible to monitor the isolation of the vein during the freeze. One limitation of this novel device is that sometimes the mapping catheter has to be positioned deep inside the target vein during ablation to achieve complete occlusion of the vein and has to be withdrawn immediately after each ablation attempt to test the PV potential. With this approach, the operator does not either need to interchange the two catheters (cryoballoon and circular mapping catheter) nor to perform double transseptal punctures.

Conclusion

Cryoballoon ablation of paroxysmal AF is a novel promising technique. Clinical trials are needed to study its efficacy in patients with persistent AF. The most frequent complication of this approach is phrenic nerve paralysis, which has been shown to be reversible in most cases. New developments like modifications of the balloon anatomy or over-the-wire techniques to enable circular PV potential recordings during cryoablation will contribute to a further simplification of the ablation procedure and an optimization of its safety profile.

References

  1. Blomstrom-Lundqvist C, Scheinman MM, Aliot EM, Alpert JS, Calkins H, Camm AJ, et al. ACC/AHA/ESC guidelines for the management of patients with supraventricular arrhythmias--executive summary: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop Guidelines for the Management of Patients With Supraventricular Arrhythmias). Circulation 2003; 108:1871–909. [CrossRef] [PubMed]
  2. Fisher JD, Spinelli MA, Mookherjee D, Krumerman AK, Palma EC. Atrial fibrillation ablation: reaching the mainstream. Pacing Clin Electrophysiol 2006; 29:523–537. [CrossRef] [PubMed]
  3. Chen SA, Hsieh MH, Tai CT, Tsai CF, Prakash VS, Yu WC, et al. Initiation of atrial fibrillation by ectopic beats originating from the pulmonary veins: electrophysiological characteristics, pharmacological responses, and effects of radiofrequency ablation. Circulation 1999; 100:1879–1886. [CrossRef] [PubMed]
  4. Haissaguerre M, Jais P, Shah DC, Takahashi A, Hocini M, Quiniou G, et al. Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. N Engl J Med 1998; 339:659–666. [CrossRef] [PubMed]
  5. Marrouche NF, Dresing T, Cole C, Bash D, Saad E, Balaban K, et al. Circular mapping and ablation of the pulmonary vein for treatment of atrial fibrillation: impact of different catheter technologies. J Am Coll Cardiol 2002; 40:464–474. [CrossRef] [PubMed]
  6. Oral H, Scharf C, Chugh A, Hall B, Cheung P, Good E, et al. Catheter ablation for paroxysmal atrial fibrillation: segmental pulmonary vein ostial ablation versus left atrial ablation. Circulation 2003; 108:2355–2360. [CrossRef] [PubMed]
  7. Ouyang F, Bansch D, Ernst S, Schaumann A, Hachiya H, Chen M, et al. Complete isolation of left atrium surrounding the pulmonary veins: new insights from the double-Lasso technique in paroxysmal atrial fibrillation. Circulation 2004; 110:2090–2096. [CrossRef] [PubMed]
  8. Calkins H, Sousa J, el-Atassi R, Rosenheck S, de Buitleir M, Kou WH, et al. Diagnosis and cure of the Wolff-Parkinson-White syndrome or paroxysmal supraventricular tachycardias during a single electrophysiologic test. N Engl J Med 1991; 324:1612–1618. [CrossRef] [PubMed]
  9. Jackman WM, Beckman KJ, McClelland JH, Wang X, Friday KJ, Roman CA, et al. Treatment of supraventricular tachycardia due to atrioventricular nodal reentry, by radiofrequency catheter ablation of slow-pathway conduction. N Engl J Med 1992; 327:313–318. [CrossRef] [PubMed]
  10. Jackman WM, Wang XZ, Friday KJ, Roman CA, Moulton KP, Beckman KJ, et al. Catheter ablation of accessory atrioventricular pathways (Wolff-Parkinson-White syndrome) by radiofrequency current. N Engl J Med 1991; 324:1605–1611. [CrossRef] [PubMed]
  11. Langberg JJ, Chin M, Schamp DJ, Lee MA, Goldberger J, Pederson DN, et al. Ablation of the atrioventricular junction with radiofrequency energy using a new electrode catheter. Am J Cardiol 1991; 67:142–147. [PubMed]
  12. Mansour M, Forleo GB, Pappalardo A, Heist EK, Avella A, Laurenzi F, et al. Initial experience with the Mesh catheter for pulmonary vein isolation in patients with paroxysmal atrial fibrillation. Heart Rhythm 2008; 5:1510–1516. [CrossRef] [PubMed]
  13. Nakagawa H, Antz M, Wong T, Schmidt B, Ernst S, Ouyang F, et al. Initial experience using a forward directed, high-intensity focused ultrasound balloon catheter for pulmonary vein antrum isolation in patients with atrial fibrillation. J Cardiovasc Electrophysiol 2007; 18:136–144. [CrossRef] [PubMed]
  14. Natale A, Pisano E, Shewchik J, Bash D, Fanelli R, Potenza D, et al. First human experience with pulmonary vein isolation using a through-the-balloon circumferential ultrasound ablation system for recurrent atrial fibrillation. Circulation 2000; 102:1879–1882. [CrossRef] [PubMed]
  15. Lemery R, Veinot JP, Tang AS, Green M, Farr N, Baxter L, et al. Fiberoptic balloon catheter ablation of pulmonary vein ostia in pigs using photonic energy delivery with Diode laser. Pacing Clin Electrophysiol 2002; 25:32–36. [CrossRef] [PubMed]
  16. Reddy VY, Houghtaling C, Fallon J, Fischer G, Farr N, Clarke J, et al. Use of a diode laser balloon ablation catheter to generate circumferential pulmonary venous lesions in an open-thoracotomy caprine model. Pacing Clin Electrophysiol 2004; 27:52–57. [CrossRef] [PubMed]
  17. Epstein MR, Knapp LD, Martindill M, Lulu JA, Triedman JK, Calkins H, et al. Embolic complications associated with radiofrequency catheter ablation. Atakr Investigator Group. Am J Cardiol 1996; 77:655–658. [CrossRef] [PubMed]
  18. Rodriguez LM, Leunissen J, Hoekstra A, Korteling BJ, Smeets JL, Timmermans C, et al. Transvenous cold mapping and cryoablation of the AV node in dogs: observations of chronic lesions and comparison to those obtained using radiofrequency ablation. J Cardiovasc Electrophysiol 1998; 9:1055–1061. [CrossRef] [PubMed]
  19. Avitall B, Urboniene D, Rozmus G, Lafontaine D, Helms R, Urbonas A. New cryotechnology for electrical isolation of the pulmonary veins. J Cardiovasc Electrophysiol 2003; 14:281–286. [CrossRef] [PubMed]
  20. Garan A, Al-Ahmad A, Mihalik T, Cartier C, Capuano L, Holtan D, et al. Cryoablation of the pulmonary veins using a novel balloon catheter. J Interv Card Electrophysiol 2006; 15:79–81. [CrossRef] [PubMed]
  21. Avitall B, Lafontaine D, Rozmus G, Adoni N, Le KM, Dehnee A, et al. The safety and efficacy of multiple consecutive cryo lesions in canine pulmonary veins-left atrial junction. Heart Rhythm 2004; 1:203–209. [CrossRef] [PubMed]
  22. Khairy P, Chauvet P, Lehmann J, Lambert J, Macle L, Tanguay JF, et al. Lower incidence of thrombus formation with cryoenergy versus radiofrequency catheter ablation. Circulation 2003; 107:2045–2050. [CrossRef] [PubMed]
  23. Gaita F, Paperini L, Riccardi R, Ferraro A. Cryothermic ablation within the coronary sinus of an epicardial posterolateral pathway. J Cardiovasc Electrophysiol 2002; 13:1160–1163. [CrossRef] [PubMed]
  24. Tse HF, Reek S, Timmermans C, Lee KL, Geller JC, Rodriguez LM, et al. Pulmonary vein isolation using transvenous catheter cryoablation for treatment of atrial fibrillation without risk of pulmonary vein stenosis. J Am Coll Cardiol 2003; 42:752–758. [CrossRef] [PubMed]
  25. Neumann T, Vogt J, Schumacher B, Dorszewski A, Kuniss M, Neuser H, et al. Circumferential pulmonary vein isolation with the cryoballoon technique results from a prospective 3-center study. J Am Coll Cardiol 2008; 52:273–278. [CrossRef] [PubMed]
  26. Sarabanda AV, Bunch TJ, Johnson SB, Mahapatra S, Milton MA, Leite LR, et al. Efficacy and safety of circumferential pulmonary vein isolation using a novel cryothermal balloon ablation system. J Am Coll Cardiol 2005; 46:1902–1912. [CrossRef] [PubMed]
  27. Van Belle Y, Janse P, Rivero-Ayerza MJ, Thornton AS, Jessurun ER, Theuns D, et al. Pulmonary vein isolation using an occluding cryoballoon for circumferential ablation: feasibility, complications, and short-term outcome. Eur Heart J 2007; 28:2231–2237. [CrossRef] [PubMed]
  28. Chun KR, Schmidt B, Metzner A, Tilz R, Zerm T, Koster I, et al. The ‘single big cryoballoon’ technique for acute pulmonary vein isolation in patients with paroxysmal atrial fibrillation: a prospective observational single centre study. Eur Heart J 2009; 30:699–709. [CrossRef] [PubMed]
  29. Siklody CH, Minners J, Allgeier M, Allgeier HJ, Jander N, Weber R, et al. Cryoballoon pulmonary vein isolation guided by transesophageal echocardiography: novel aspects on an emerging ablation technique. J Cardiovasc Electrophysiol 2009; 20:1197–1202. [CrossRef] [PubMed]
  30. Siklody CH, Minners J, Allgeier M, Allgeier HJ, Jander N, Keyl C, et al. Pressure-guided cryoballoon isolation of the pulmonary veins for the treatment of paroxysmal atrial fibrillation. J Cardiovasc Electrophysiol 2010; 21:120–125. [CrossRef] [PubMed]
  31. Nolker G, Heintze J, Gutleben KJ, Muntean B, Putz V, Yalda A, et al. Cryoballoon Pulmonary Vein Isolation Supported by Intracardiac Echocardiography: Integration of a Nonfluoroscopic Imaging Technique in Atrial Fibrillation Ablation. J Cardiovasc Electrophysiol 2010 Dec; 21(12):1325-30. [CrossRef] [PubMed]
  32. Chun KR, Furnkranz A, Metzner A, Schmidt B, Tilz R, Zerm T, et al. Cryoballoon pulmonary vein isolation with real-time recordings from the pulmonary veins. J Cardiovasc Electrophysiol 2009; 20:1203–1210. [CrossRef] [PubMed]
  33. Furnkranz A, Chun KR, Nuyens D, Metzner A, Koster I, Schmidt B, et al. Characterization of conduction recovery after pulmonary vein isolation using the “single big cryoballoon” technique. Heart Rhythm 2010; 7:184–190. [CrossRef] [PubMed]
  34. Klein G, Oswald H, Gardiwal A, Lusebrink U, Lissel C, Yu H, et al. Efficacy of pulmonary vein isolation by cryoballoon ablation in patients with paroxysmal atrial fibrillation. Heart Rhythm 2008; 5:802–806. [CrossRef] [PubMed]
  35. Antz M, Chun KR, Ouyang F, Kuck KH. Ablation of atrial fibrillation in humans using a balloon-based ablation system: identification of the site of phrenic nerve damage using pacing maneuvers and CARTO. J Cardiovasc Electrophysiol 2006; 17:1242–1245. [CrossRef] [PubMed]
  36. Packer D, Irwin JM, Champagne J, Guerra PG, Dubuc M, Wheelan KR, et al. Cryoballoon ablation of pulmonary veins for paroxysmal atrial fibrillation: first results of the North American Arctic Front STOP-AF pivotal trial [Abstract]. J Am Coll Cardiol 2010; 55:E3015–3016. [CrossRef] [PubMed]
  37. Sacher F, Monahan KH, Thomas SP, Davidson N, Adragao P, Sanders P, et al. Phrenic nerve injury after atrial fibrillation catheter ablation: characterization and outcome in a multicenter study. J Am Coll Cardiol 2006; 47:2498–2503. [CrossRef] [PubMed]
  38. Linhart M, Bellmann B, Mittmann-Braun E, Schrickel JW, Bitzen A, Andrie R, et al. Comparison of cryoballoon and radiofrequency ablation of pulmonary veins in 40 patients with paroxysmal atrial fibrillation: a case-control study. J Cardiovasc Electrophysiol 2009; 20:1343–1348. [CrossRef] [PubMed]
  39. Luik A, Merkel M, Hoeren D, Riexinger T, Kieser M, Schmitt C. Rationale and design of the FreezeAF trial: a randomized controlled noninferiority trial comparing isolation of the pulmonary veins with the cryoballoon catheter versus open irrigated radiofrequency ablation in patients with paroxysmal atrial fibrillation. Am Heart J 2010; 159:555–560 e1. [CrossRef] [PubMed]
  40. Furnkranz A, Chun KR, Metzner A, Nuyens D, Schmidt B, Burchard A, et al. Esophageal endoscopy results after pulmonary vein isolation using the single big cryoballoon technique. J Cardiovasc Electrophysiol 2010 Aug 1; 21(8):869-74. [CrossRef] [PubMed]
  41. Cappato R, Calkins H, Chen SA, Davies W, Iesaka Y, Kalman J, et al. Updated worldwide survey on the methods, efficacy, and safety of catheter ablation for human atrial fibrillation. Circ Arrhythm Electrophysiol. 2010 Feb 1; 3(1):32-8. [CrossRef] [PubMed]
  42. Mansour M, Forleo GB, Pappalardo A, Barrett C, Heist EK, Avella A, et al. Combined use of cryoballoon and focal open-irrigation radiofrequency ablation for treatment of persistent atrial fibrillation: results from a pilot study. Heart Rhythm 2010; 7:452–458. [CrossRef] [PubMed]
  43. Tang M, Kriatselis C, Nedios S, Ye G, Roser M, Fleck E, et al. A novel cryoballoon technique for mapping and isolating pulmonary veins: a feasibility and efficacy study. J Cardiovasc Electrophysiol 2010; 21:626–631. [CrossRef] [PubMed]