Journal of Innovation in Cardiac Rhythm Management
Articles Articles 2011 August

Clinical Applications of Remote Implantable Cardioverter-Defibrillator Monitoring: Current Status and Future Directions

DOI: 10.19102/icrm.2011.020804

1JACOB C. JENTZER, MD and 2JOHN H. JENTZER, MD

1Duke University Medical Center, Durham, NC
2Dixie Regional Medical Center, St. George, UT

PDF Download PDF
tweeter Follow Us >>

ABSTRACT.Implantable cardioverter-defibrillator and cardiac resynchronization devices are integral for reducing mortality in many heart failure patients. Monitoring of these devices can be facilitated using wireless remote transmission. Remote monitoring can lead to early detection of important events such as device or lead malfunction, pacing abnormalities, arrhythmias, and device-delivered antiarrhythmia therapy. Studies of remote monitoring show a reduction in the response time to clinical events. Remote monitoring may safely reduce the need for frequent office device checks and facilitate triage of device-detected events. Monitoring of several device-detected parameters can predict heart failure decompensation, and interventions based on this information may prevent heart failure hospitalization. Ongoing clinical studies will further elucidate the benefits of an expanding array of remote monitoring possibilities.

KEYWORDS.ambulatory monitoring, artificial cardiac pacing, cardiac arrhythmias, implantable defibrillators, heart failure.

The authors report no conflicts of interest for the published content.
Manuscript submitted May 23, 2011, final version accepted June 6, 2011.
Address correspondence to: John Jentzer, MD, Southwest Cardiology Associates, Dixie Regional Medical Center, 1380 East Medical Center Drive, St. George, UT 84790. E-mail: john.jentzer@imail.org

Introduction

Implantable cardioverter-defibrillators (ICDs) reduce all-cause mortality by preventing sudden death in patients with a history of sustained ventricular arrhythmias1 and in patients with left ventricular (LV) systolic dysfunction at high risk of ventricular arrhythmias.2 Addition of cardiac resynchronization therapy (CRT) to ICDs (CRT-D) via biventricular pacing provides a further reduction in mortality in populations with LV systolic dysfunction, QRS prolongation and moderate-to-severe heart failure (HF) symptoms.3,4 In patients with LV systolic dysfunction, prolonged QRS and mild-to-moderate HF symptoms, CRT-D therapy reduces the incidence of HF hospitalizations, improves indices of LV function and remodeling, and may reduce mortality beyond ICD alone.5 In 2007, over 234,00 ICDs and over 148,000 CRT-Ds were implanted in North America.6 Monitoring of these devices has placed an increasing burden on the electrophysiology device community due to regularly scheduled in-office ICD follow-ups which usually do not result in clinical intervention.79 Expert consensus recommends ICD follow-up every 3–6 months,6 but clinically important events between clinic visits may be silent until the next scheduled device interrogation. Remote monitoring (RM) via periodic wireless downloading of device data can provide more frequent surveillance of devices, allowing for early problem identification and clinical intervention.10 RM may safely allow reduced frequency of in-office device checks,9 which could translate into reductions in healthcare costs.1114

Remote monitoring basics

RM of ICD and CRT-D devices has become an accepted6 and evolving strategy for device surveillance since its clinical introduction by Biotronik in 2000.1517 Each major device manufacturer has a proprietary wireless RM system: Biotronik (Lake Oswego,OR) Home Monitoring, Boston Scientific (St. Paul, MN) Latitude, Medtronic (Mounds View, MN) CareLink, and St. Jude (St. Paul, MN)Merlin.net (previously HouseCall Plus). These systems share the capability to transmit device diagnostic information automatically to a local receiver which sends this information transtelephonically to a central server, allowing access by providers who can receive clinical alerts.1517 Depending on the system, device data can be downloaded automatically each day or downloads may only occur when triggered by the patient or a device alert.16 Remote device reprogramming is not currently available on any of these systems. Further details regarding the technical specifications of the different RM systems are reviewed elsewhere.1517 Patterned after remote pacemaker monitoring of battery voltage, magnet rate, and current heart rhythm, modern ICD and CRT-D devices monitor a much broader array of parameters. Pacing and shock lead impedance, device-triggered alerts, sensing function, programmed device parameters, percent pacing and sensing in each chamber, and review of detected arrhythmias and therapies with stored or real-time electrograms can be monitored (Table 1).15 Device-based parameters and algorithms for HF monitoring are being developed for incorporation into clinical decision making.18 We hope that this evolving RM armamentarium may translate into better patient outcomes and health care savings.

Table 1: Remotely monitored device parameters

crm-02-08-430-t1.jpg

Remote monitoring of ICD and CRT-D pacemaker function

Analogous to standard pacemaker evaluation, RM can confirm proper pacing function of ICD and CRT-D devices and can identify the need for in-office reprogramming to optimize device function. Abnormal pacing function can contribute to worsening HF in ICD and CRT-D patients, and remote monitoring will disclose the amount of right ventricular (RV) pacing or biventricular pacing. Excessive RV pacing (above 40%) is associated with increased mortality and HF events in patients with ICDs19 and may predispose to ventricular arrhythmias20 and attenuated ICD benefit.21 Excessive RV pacing can often be minimized by lengthening the programmed AV delay or use of pacing-minimization device algorithms, but may require device upgrade to include CRT, particularly in patients who become pacing dependent.22

Inadequate percentage of biventricular pacing (below 93%) can eliminate the clinical benefits of CRT.23,24 RM can detect drops in percent biventricular pacing, warranting a full evaluation to identify the etiology.25 Loss of LV pacing capture in CRT-D devices due to LV lead malfunction is suggested by a dramatic change in the LV pacing resistance, which must be confirmed with formal pacing threshold testing in a device clinic. An excessively long programmed AV delay may allow intrinsic conduction through the native conduction system to pre-empt biventricular pacing, especially during exertion. Evaluation of electrograms with their marker channels will disclose ventricular sensing rather than biventricular pacing. Atrial tachyarrhythmias (ATs), supraventricular tachyarrhythmias (SVTs) or sinus tachycardia with ventricular rates exceeding the upper rate limit of the pacemaker will inhibit biventricular pacing,25 as can frequent ventricular premature beats. In CRT-D patients with atrial fibrillation, AV nodal ablation may be required to yield the full benefit of biventricular pacing.26

Remote monitoring of ICD lead integrity

Oversensing of non-physiological signals is an important non-arrhythmic cause of inappropriate ICD shocks, generally due to lead fracture, insulation breaks, or lead–header connection problems.27 The majority of ICD system complications are related to lead integrity failure, which continues to be a significant clinical problem.28 Rates of modern ICD lead failure are relatively low (0.6%/year), but can be higher with certain high-risk ICD leads such as the Medtronic Sprint Fidelis lead (3.8%/year, hazard ratio 6.4);29,30 yearly lead failure rates appear to accelerate over time.31 Changes in lead impedance and short sensing intervals from lead fracture can be identified earlier with RM, allowing timely clinical intervention.32,33 Figure 1a illustrates an abrupt increase in ICD lead pacing impedance due to a lead fracture. Electrograms and their marker channels downloaded remotely from Latitude (Figure 1b) reveal oversensing of noise and confirm the diagnosis. Changes in lead impedance alone are not adequately sensitive for lead fracture,3335 because oversensing is often the first manifestation of lead fracture.3537 Device reprogramming with a downloadable lead integrity algorithm may improve lead fracture detection by identifying oversensing and increases in impedance.33 This algorithm can prevent many inappropriate ICD shocks in patients with Sprint Fidelis leads,33 but remains unable to completely prevent inappropriate ICD shocks in patients with fractured leads.36 Further analysis of downloaded device data can differentiate lead fracture from connection problems and normal lead function in patients with high-impedance lead integrity algorithm alerts.37 Lead malfunction may also cause ventricular undersensing, which can prevent delivery of appropriate therapy for a ventricular arrhythmia. RM may detect more than 90% of lead-related complications and reduce symptomatic lead failure episodes by half,32 while reducing the time to recognition of lead complications.38 Continuous follow-up in a device clinic plus RM is recommended for patients implanted with a high-risk lead.39 Oversensing of physiologic signals, such as T-wave oversensing (Figure 2) or sensing of myopotentials40 may be corrected with a change in device sensitivity or may require lead repositioning or implantation of a new lead.

crm-02-08-430-f1.jpg

Figure 1: (a) Ventricular pacing impedance histogram downloaded from Boston Scientific Latitude showing an abrupt rise in right ventricular (RV) pacing impedance (top) in November 2010 at the time of lead fracture. (b) Electrograms downloaded from Boston Scientific Latitude showing right atrium (top), right ventricle (middle) and far-field right ventricle (bottom). Noise on the RV sensing electrogram (middle) is indicative of RV lead fracture, leading to oversensing interpreted by the device as ventricular tachyarrhythmias and treated by antitachycardia pacing (V-Detect, end of strip).

crm-02-08-430-f2.jpg

Figure 2: Electrograms from Biotronik Home Monitoring showing far-field ventricular (top), right atrium (middle) and right ventricle (bottom). VF markers overlie the T wave, indicating T wave oversensing. VS indicates ventricular sensing; VF indicates ventricular sensing in the ventricular fibrillation zone; Det VF indicates ventricular fibrillation detection.

Remote monitoring of arrhythmias

RM provides an effective means to evaluate device-detected arrhythmias and device-delivered therapy such as antitachycardia pacing (ATP), low-energy cardioversion, and defibrillation. RM systems allow accurate evaluation of stored electrograms taken during symptomatic or asymptomatic episodes of arrhythmias or device therapy.41 Electrogram analysis helps determine the appropriateness and effectiveness of device-delivered therapy and can identify device or lead malfunctions and adverse effects related to device therapy. If device therapy was not effective, a change in device programming, medications, or hardware configuration may be necessary. Changes in ICD programming or antiarrhythmic medication may require formal in-hospital evaluation of cardioversion and defibrillation thresholds and ATP efficacy, such as with device-based non-invasive programmed stimulation.

Ventricular arrhythmias and ICD shocks

Although ICDs reduce mortality by preventing death due to ventricular arrhythmias,42 patients receiving ICD shock have a higher risk of both arrhythmic and pump failure death than patients without ICD shocks.43 Patients with more frequent ICD shocks have a higher mortality than patents with fewer ICD shocks.44 The mortality risk is greater for appropriate ICD shocks (treated ventricular arrhythmias) than for inappropriate ICD shocks,43,45 which are most often due to AT or other SVTs and less often due to ventricular oversensing (typically from lead failure).35 Inappropriate ICD shocks due to lead complications may not pose an excess mortality risk.7 Early detection is paramount because ICD shocks correlate with HF instability, often requiring adjustment of medications, clinical assessment and/or device reprogramming.46 Strategies to avoid inappropriate ICD therapy due to SVT include antiarrhythmic agents, catheter ablation, or device programming of atrial fibrillation and sinus tachycardia discriminators.35,47 ATP can be effective in more than 70% of ventricular tachycardia (VT) episodes, safely preventing unnecessary ICD shocks.48,49 Unlike ICD shocks, VT episodes terminated by ATP may not confer an adverse prognosis.44 ATP can occasionally be associated with proarrhythmia such as acceleration of VT rate,49 in which case ATP should be eliminated. Frequent episodes of VT may require pharmacologic or ablative intervention, particularly when device therapy is ineffective.50,51 Figure 3 shows electrograms from an episode of unsuccessful ATP therapy for sustained VT, which prompted referral for VT ablation that successfully eliminated most of the patient’s ventricular arrhythmia burden.

crm-02-08-430-f3.jpg

Figure 3: Electrograms (right atrium, top and right ventricle, bottom) downloaded from Medtronic CareLink demonstrating ventricular tachyarrhythmia that is not terminated with ATP (VT Rx 1 Burst), indicated by persistent tachycardia sensing after antitachycardia pacing. TS indicates tachycardia sensing; TP indicates tachycardia pacing.

Atrial arrhythmias

Symptomatic and asymptomatic AT are common in HF patients, and may be identified during the first post-implant year in up to one-fourth of CRT-D recipients who are in sinus rhythm at the time of implantation.52,53 Knowledge regarding the incidence, duration, frequency, and ventricular rate response of AT can be gleaned easily using RM (Figure 4).52 AT episodes are one of the most common clinical alerts identified by RM,10,54 and device-detected AT episodes predict worse outcomes in ICD and CRT-D patients.55 AT can interfere with proper device functioning by preventing optimal biventricular pacing25,26 or causing inappropriate ICD shocks.35 RM may allow earlier intervention for new or worsening device-detected AT,56 but it remains unclear if early detection of asymptomatic AT improves clinical outcomes. Device-detected atrial high-rate episodes correspond to AT and predict increased stroke risk.57 Mathematical modeling suggests that early initiation of anticoagulation based on asymptomatic device-detected AT could prevent embolism and stroke,58 and an ongoing randomized trial will directly test this hypothesis.59

crm-02-08-430-f4.jpg

Figure 4: Electrograms downloaded from Boston Scientific Latitude in right atrium (top), right ventricle (middle) and ventricular far-field (bottom), showing 2:1 conducted atrial flutter. AS indicates atrial sensing; AF indicates atrial sensing in atrial fibrillation zone; VP-MT indicates ventricular pacing at maximum tracking rate.

Clinical trials of remote device monitoring

Whereas transtelephonic pacemaker monitoring appears useful primarily for battery status determination,60 recent clinical trials using RM of ICD and CRT-D devices demonstrate earlier identification of clinical events10 and safe reduction in clinic visits.9 The CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) trial randomized 1,997 Medtronic ICD and CRT-D patients to standard in-office assessment with or without RM, and showed decreased time to clinical decision making by 17 days (more than 75%) and reduced healthcare utilization days with RM.10 The majority of remotely detected clinical events in CONNECT were prolonged AT episodes, consistent with prior studies.54 The TRUST (Lumos-T Safely Reduces Routine Office Device Follow-up) trial randomized 1,339 Biotronik ICD patients to standard office-only follow-up every 3 months or less frequent office visits plus RM and demonstrated a 45% reduction of in-office visits and a reduction in time to respond to clinical alerts using RM, without increasing adverse events.9 TRUST confirmed the safety of using RM in lieu of more frequent office visits for device checks, showing that 85% of follow-ups between 3 and 12 months after implantation can be accomplished remotely and that more than 90% of scheduled clinic device checks did not require intervention. These studies confirm the results of smaller studies suggesting the ability of RM to reduce response time to clinical events54,61 and reduce in-hospital follow-up visits.14,62 Earlier identification of clinical events by RM did not translate into an improvement in clinical outcomes in either CONNECT10 or TRUST.9 Small studies suggest that a reduction in clinic visits by RM may reduce healthcare costs,1214 particularly for patients with longer driving distances,11 but we await formal economic analysis from larger trials for confirmation. Further cost savings are possible if early detection of clinical events by RM leads to interventions preventing hospitalization, as suggested in CONNECT by a reduced number of hospital days.10 The observational ALTITUDE study compared 69,556 Boston Scientific ICD and CRT-D patients who underwent RM with 116,222 similar patients who received in-office device follow-up only, and identified a 50% mortality reduction for those patients undergoing RM, with better survival at 1 and 5 years.7 The non-randomized design of ALTITUDE limits conclusions about causality, but baseline demographics did not differ between the two groups, so a dramatic imbalance in risk factor burden would be required to explain the survival difference. ALTITUDE is the only published study showing improved clinical outcomes associated with RM, although smaller studies suggest improvements in clinical care56,63 or patient satisfaction and quality of life64 without clear outcome benefits.

Remote monitoring of heart failure status

The majority of ICD and CRT-D devices are implanted in patients with HF, a group at high risk of hospitalization.65 Weight gain and worsening of HF symptoms may occur late66 and have poor sensitivity for predicting HF hospitalization.60,67 Several device-measured parameters (Table 2) can predict HF decompensation, allowing RM to facilitate early clinical intervention and potentially avert hospitalization.14,15,18,68 Intrathoracic impedance (measured between the RV lead and the device generator) varies inversely with lung fluid content, such that a decline in intrathoracic impedance suggests worsening pulmonary congestion.69 Intrathoracic impedance correlates inversely with pulmonary capillary wedge pressure,70 RV pressures,71 and natriuretic peptide levels.72 Declines in intrathoracic impedance can also occur due to device pocket complications, pneumonia, or pleural effusion.69 A decline in intrathoracic impedance may occur nearly 2 weeks before worsening symptoms and weight gain preceding a HF exacerbation.70 Declines in intrathoracic impedance may be associated with worsening atrial73 and ventricular arrhythmias.74 Intrathoracic impedance and AT demonstrate a bidirectional relationship, in which reduced intrathoracic impedance predicts more AT and increasing AT burden predicts a reduction in intrathoracic impedance.73 The Medtronic OptiVol feature calculates a positive numeric OptiVol index to facilitate recognition of a sustained reduction in intrathoracic impedance over time, measured in ohm-days (Figure 5a); a remote or audible alert can be sent when a preset threshold is crossed.69 Heart rate variability and patient activity level also decline as much as 2 weeks before HF hospitalization (Figure 5b).68,75,76 The Medtronic Cardiac Compass alert feature integrates the OptiVol index with patient activity, nocturnal heart rate, and heart rate variability as well as clinical events warranting provider notification such as increasing AT, reduction in biventricular pacing, or ICD shocks.77

crm-02-08-430-f5.jpg

Figure 5: (a) Remote monitoring of OptiVol using Medtronic CareLink. Intrathoracic impedance (bottom) declines in mid-June, leading to a OptiVol (top) threshold crossing event in late June, successfully treated by uptitration of diuretics. A similar episode occurs at the beginning of August, leading to hospitalization. (b) Remote monitoring of patient activity level (top) and heart rate variability (third from top) downloaded from Medtronic CareLink. Patient activity and heart rate variability decline coincident with decline in intrathoracic impedance and OptiVol threshold crossing.

Table 2: Device parameters predicting heart failure decompensation

crm-02-08-430-t2.jpg

Clinical studies of remote heart failure monitoring

A standard OptiVol threshold of 60 ohm-days has up to 60–76% sensitivity for detecting future HF hospitalizations in CRT-D patients,67,70,78 and many missed HF events are associated with a subthreshold decline in intrathoracic impedance. Each OptiVol threshold crossing per year predicts a 36% increased risk of HF hospitalization.76 The FAST (Fluid Accumulation Status Trial) study demonstrated that an OptiVol threshold of 60 ohm-days had superior sensitivity (76% versus 23%) and fewer false positives than daily weight monitoring for prediction of HF worsening in 156 CRT-D patients.67 The SENSE-HF study of 501 ICD and CRT-D patients showed low sensitivity (42%) and positive predictive value (38%) of the standard 60 ohm-day OptiVol threshold for predicting HF hospitalizations, especially within the first 6 months after device implantation.79 False-positive OptiVol threshold crossings occur at a rate of 0.25–1.9 events per year, including episodes in which medication titration avoided hospitalization and lesser degrees of decompensation not leading to hospitalization.67,68,78 The majority of OptiVol threshold crossings at 60 ohm-days may not be associated with a clinical HF event.67,68,76,79 Higher OptiVol thresholds of 100–120 ohm-days have been proposed to improve specificity,77,80 and refinements in the OptiVol algorithm may increase specificity, reduce false-positive threshold crossing events, and improve positive predictive value of OptiVol.81

Given these limitations in sensitivity and specificity, intrathoracic impedance and OptiVol cannot be used as a sole means of patient evaluation, but may supplement daily weight monitoring and standard clinical assessment to facilitate telemonitoring programs.82,83 The observational DECODE (Decompensation Detection)study of 699 Boston Scientific CRT-D patients found that adding heart rate variability and lead (not intrathoracic) impedance monitoring to standard telemonitoring yielded poor sensitivity (<50%) for predicting HF hospitalizations, with frequent false-positive alerts (two per patient-year).84 An observational study of 532 Medtronic CRT patients showed that intrathoracic impedance monitoring using OptiVol was associated with a reduced risk of HF hospitalization.68 The observational PARTNERS-HF (Program to Access and Review Trending Information and Evaluate Correlation to Symptoms in Patients With Heart Failure) study of 694 Medtronic CRT-D patients showed that the remote Cardiac Compass alert predicted a fivefold increased 1-month risk of HF hospitalization with improved sensitivity over the OptiVol index alone.77 Further research is needed to demonstrate that RM of intrathoracic impedance and other parameters can prevent HF events, and ongoing randomized studies will help to clarify this question.85

Remote hemodynamic monitoring by devices

Dedicated implantable devices can remotely monitor intracardiac hemodynamics in HF patients. Parallel to declines in intrathoracic impedance,70 intracardiac pressures rise before weight gain or symptom worsening preceding HF decompensation,86 allowing RM of hemodynamic parameters to facilitate early intervention.18 Investigational wireless implantable devices can measure pulmonary artery pressure (CardioMEMS Champion)87 or left atrial pressure (St. Jude HeartPOD),86 or estimate pulmonary capillary wedge pressure from RV hemodynamics (Medtronic Chronicle).88 Physician-directed medication titration based on device hemodynamic readings may prevent HF hospitalizations.8688 The randomized COMPASS-HF trial of 274 advanced HF patients found that Chronicle-guided patient management produced a non-significant 21% reduction in HF events.88 HeartPOD-guided medication titration increased utilization of evidence-based HF therapies and improved functional class.86 The randomized CHAMPION trial of 550 advanced HF patients showed that Champion-guided patient management reduced total HF hospitalizations by 36%.87 Although not available on current ICD and CRT-D devices, attempts are being made to integrate intracardiac pressure monitoring into the next generation of devices.89

Integrating remote monitoring into clinical practice

A number of potential logistical and technical pitfalls limit the clinical application of RM.16 Currently available RM systems generally transmit via telephone landline only, but mobile phone compatibility is available for some models. Transmission failure can occur, but technical success rates generally exceed 90%.9 Patients must be actively involved in RM for best results, and device manufacturer representatives can often facilitate patient training and enrollment. Remote device alerts may occur in the majority of patients within 18 months after implantation,90 and frequent alerts for non-critical events may add to physician workload. Remote alert management may require individualized patient alert settings and/or a nurse-run triage system, and a specific plan for device alert management is recommended,16,62 This approach may reduce the number of alerts reaching the responsible physician by 90%, leading to a minimal increase in physician workload estimated at approximately 15 min per week per 100 patients.62,90,91 Physicians could potentially be liable if critical RM data are not acted on in a timely manner, so direct-to-physician alerts for critical events may be preferred. False-positive OptiVol alerts can occur at a rate approaching two per patient per year,67 but OptiVol monitoring can be successfully integrated into a multidisciplinary heart failure disease management program without excessive provider burden.83 Integrating RM data into electronic medical records facilitates documentation and coordination of clinician responses to remote alerts, and this capability is available on some systems.12,13,16 Reimbursement can be available for RM encounters as often as every 3 months in place of reimbursement for an office visit; more frequent RM encounters are generally not reimbursed unless a change in therapy occurs. RM may allow a reduced frequency of scheduled in-office device checks, but it cannot currently replace all in-office device checks or obviate the need for patient–clinician encounters. While the majority of routine device checks do not yield relevant findings requiring intervention,7,9 unscheduled visits are much more likely to require intervention or hospitalization.8,62 RM provides an opportunity for early triage of clinically relevant events, allowing prompt in-office follow-up for significant events and allowing fewer critical alerts to be managed without face-to-face contact unless necessary.

Conclusion

With the proliferation of ICD and CRT-D devices and the expanding HF population, RM of these devices may become increasingly important for patient care. RM can allow early warning of lead malfunction, pacing abnormalities, arrhythmias, and HF deterioration to allow timely intervention. RM may safely allow less frequent in-office device checks, facilitating care of patients living far away from their providers and potentially reducing healthcare costs. The evolving HF monitoring features of ICD and CRT-D devices are an exciting advance that may simplify clinical management of a complicated and challenging patient group. We hope that ongoing clinical studies of RM will clearly document clinical benefits and cost savings to justify the widespread proliferation of this technology.

References

  1. Connolly SJ, Hallstrom AP, Cappato R, et al. Meta-analysis of the implantable cardioverter defibrillator secondary prevention trials. AVID, CASH and CIDS studies. Antiarrhythmics vs Implantable Defibrillator study. Cardiac Arrest Study Hamburg. Canadian Implantable Defibrillator Study. Eur Heart J 2000; 21:2071–2078. [CrossRef] [PubMed]
  2. Theuns DA, Smith T, Hunink MG, Bardy GH, Jordaens L. Effectiveness of prophylactic implantation of cardioverter-defibrillators without cardiac resynchronization therapy in patients with ischaemic or non-ischaemic heart disease: a systematic review and meta-analysis. Europace 2010; 12:1564–1570. [CrossRef] [PubMed]
  3. Bristow MR, Saxon LA, Boehmer J, et al. Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N Engl J Med 2004; 350:2140–2150. [CrossRef] [PubMed]
  4. McAlister FA, Ezekowitz J, Hooton N, et al. Cardiac resynchronization therapy for patients with left ventricular systolic dysfunction: a systematic review. JAMA. 2007; 297:2502–2514. [CrossRef] [PubMed]
  5. Al-Majed NS, McAlister FA, Bakal JA, Ezekowitz JA. Meta-analysis: cardiac resynchronization therapy for patients with less symptomatic heart failure. Ann Intern Med 2011; 154:401–412. [CrossRef] [PubMed]
  6. Wilkoff BL, Auricchio A, Brugada J, et al. HRS/EHRA expert consensus on the monitoring of cardiovascular implantable electronic devices (CIEDs): description of techniques, indications, personnel, frequency and ethical considerations. Heart Rhythm 2008; 5:907–925. [PubMed]
  7. Saxon LA, Hayes DL, Gilliam FR, et al. Long-term outcome after ICD and CRT implantation and influence of remote device follow-up: the ALTITUDE survival study. Circulation 2010; 122:2359–2367. [CrossRef] [PubMed]
  8. Heidbuchel H, Lioen P, Foulon S, et al. Potential role of remote monitoring for scheduled and unscheduled evaluations of patients with an implantable defibrillator. Europace 2008; 10:351–357. [CrossRef] [PubMed]
  9. Varma N, Epstein AE, Irimpen A, Schweikert R, Love C. Efficacy and safety of automatic remote monitoring for implantable cardioverter-defibrillator follow-up: the Lumos-T Safely Reduces Routine Office Device Follow-up (TRUST) trial. Circulation 2010; 122:325–332. [CrossRef] [PubMed]
  10. Crossley GH, Boyle A, Vitense H, Chang Y, Mead RH. The CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) trial: the value of wireless remote monitoring with automatic clinician alerts. J Am Coll Cardiol 2011; 57:1181–1189. [CrossRef] [PubMed]
  11. Fauchier L, Sadoul N, Kouakam C, et al. Potential cost savings by telemedicine-assisted long-term care of implantable cardioverter defibrillator recipients. Pacing Clin Electrophysiol 2005; 28(Suppl 1):S255–259. [CrossRef] [PubMed]
  12. Raatikainen MJ, Uusimaa P, van Ginneken MM, Janssen JP, Linnaluoto M. Remote monitoring of implantable cardioverter defibrillator patients: a safe, time-saving, and cost-effective means for follow-up. Europace 2008; 10:1145–1151. [CrossRef] [PubMed]
  13. Bikou O, Licka M, Kathoefer S, Katus HA, Bauer A. Cost savings and safety of ICD remote control by telephone: a prospective, observational study. J Telemed Telecare 2010; 16:403–408. [CrossRef] [PubMed]
  14. Elsner CH, Sommer P, Piorkowski C, et al. A prospective multicenter comparison trial of home monitoring against regular follow-up in MADIT II patients: additional visits and cost impact. Computers Cardiol 2006; 33:241–244.
  15. Andrikopoulos G, Tzeis S, Theodorakis G, Vardas P. Monitoring capabilities of cardiac rhythm management devices. Europace 2010; 12:17–23. [CrossRef] [PubMed]
  16. Heidbuchel H. Telemonitoring of implantable cardiac devices: hurdles towards personalised medicine. Heart 2011; 97:931–939. [CrossRef] [PubMed]
  17. Orlov MV, Szombathy T, Chaudhry GM, Haffajee CI. Remote surveillance of implantable cardiac devices. Pacing Clin Electrophysiol 2009; 32:928–939. [CrossRef] [PubMed]
  18. Samara MA, Wilson Tang WH. Device monitoring strategies in acute heart failure syndromes. Heart Fail Rev 2011; (in press). [CrossRef] [PubMed]
  19. Sharma AD, Rizo-Patron C, Hallstrom AP, et al. Percent right ventricular pacing predicts outcomes in the DAVID trial. Heart Rhythm 2005; 2:830–834. [CrossRef] [PubMed]
  20. Gardiwal A, Yu H, Oswald H, et al. Right ventricular pacing is an independent predictor for ventricular tachycardia/ventricular fibrillation occurrence and heart failure events in patients with an implantable cardioverter-defibrillator. Europace 2008; 10:358–363. [CrossRef] [PubMed]
  21. Barsheshet A, Moss AJ, McNitt S, et al. Long-term implications of cumulative right ventricular pacing among patients with an implantable cardioverter-defibrillator. Heart Rhythm 2011; 8:212–218. [CrossRef] [PubMed]
  22. Martinelli Filho M, de Siqueira SF, Costa R, et al. Conventional versus biventricular pacing in heart failure and bradyarrhythmia: the COMBAT study. J Card Fail 2010; 16:293–300. [CrossRef] [PubMed]
  23. Koplan BA, Kaplan AJ, Weiner S, Jones PW, Seth M, Christman SA. Heart failure decompensation and all-cause mortality in relation to percent biventricular pacing in patients with heart failure: is a goal of 100% biventricular pacing necessary? J Am Coll Cardiol 2009; 53:355–360. [CrossRef] [PubMed]
  24. Hayes DL, Boehmer JP, Day J, et al. Cardiac resynchronization therapy and the relationship of percent biventricular pacing to symptoms and survival. Heart Rhythm 2011; (in press). [CrossRef] [PubMed]
  25. Leclercq C. Problems and troubleshooting in regular follow-up of patients with cardiac resynchronization therapy. Europace 2009; 11(Suppl 5):v66–71. [CrossRef] [PubMed]
  26. Wilton SB, Leung AA, Ghali WA, Faris P, Exner DV. Outcomes of cardiac resynchronization therapy in patients with versus those without atrial fibrillation: A systematic review and meta-analysis. Heart Rhythm 19 2011; (in press). [CrossRef] [PubMed]
  27. Swerdlow CD, Friedman PA. Advanced ICD troubleshooting: Part I. Pacing Clin Electrophysiol 2005; 28:1322–1346. [CrossRef] [PubMed]
  28. Maisel WH, Hauser RG. Proceedings of the ICD Lead Performance Conference. Heart Rhythm 2008; 5:1331–1338. [CrossRef] [PubMed]
  29. Hauser RG, Hayes DL. Increasing hazard of Sprint Fidelis implantable cardioverter-defibrillator lead failure. Heart Rhythm 2009; 6:605–610. [CrossRef] [PubMed]
  30. Borleffs CJ, van Erven L, van Bommel RJ, et al. Risk of failure of transvenous implantable cardioverter-defibrillator leads. Circ Arrhythm Electrophysiol 2009; 2:411–416. [CrossRef] [PubMed]
  31. Kleemann T, Becker T, Doenges K, et al. Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years. Circulation. 2007; 115:2474–2480. [CrossRef] [PubMed]
  32. Spencker S, Coban N, Koch L, Schirdewan A, Muller D. Potential role of home monitoring to reduce inappropriate shocks in implantable cardioverter-defibrillator patients due to lead failure. Europace 2009; 11:483–488. [CrossRef] [PubMed]
  33. Swerdlow CD, Gunderson BD, Ousdigian KT, et al. Downloadable algorithm to reduce inappropriate shocks caused by fractures of implantable cardioverter-defibrillator leads. Circulation 2008; 118:2122–2129. [CrossRef] [PubMed]
  34. Kallinen LM, Hauser RG, Lee KW, et al. Failure of impedance monitoring to prevent adverse clinical events caused by fracture of a recalled high-voltage implantable cardioverter-defibrillator lead. Heart Rhythm 2008; 5:775–779. [CrossRef] [PubMed]
  35. Tzeis S, Andrikopoulos G, Kolb C, Vardas PE. Tools and strategies for the reduction of inappropriate implantable cardioverter defibrillator shocks. Europace 2008; 10:1256–1265. [CrossRef] [PubMed]
  36. Kallinen LM, Hauser RG, Tang C, et al. Lead integrity alert algorithm decreases inappropriate shocks in patients who have Sprint Fidelis pace-sense conductor fractures. Heart Rhythm 2010; 7:1048–1055. [CrossRef] [PubMed]
  37. Swerdlow CD, Sachanandani H, Gunderson BD, Ousdigian KT, Hjelle M, Ellenbogen KA. Preventing overdiagnosis of implantable cardioverter-defibrillator lead fractures using device diagnostics. J Am Coll Cardiol 2011; 57:2330–2339. [CrossRef] [PubMed]
  38. Varma N, Michalski J, Epstein AE, Schweikert R. Automatic remote monitoring of implantable cardioverter-defibrillator lead and generator performance: the Lumos-T Safely RedUceS RouTine Office Device Follow-Up (TRUST) trial. Circ Arrhythm Electrophysiol 2010; 3:428–436. [CrossRef] [PubMed]
  39. Maisel WH, Hauser RG, Hammill SC, et al. Recommendations from the Heart Rhythm Society Task Force on Lead Performance Policies and Guidelines: developed in collaboration with the American College of Cardiology (ACC) and the American Heart Association (AHA). Heart Rhythm 2009; 6:869–885. [CrossRef] [PubMed]
  40. Kowalski M, Ellenbogen KA, Wood MA, Friedman PL. Implantable cardiac defibrillator lead failure or myopotential oversensing? An approach to the diagnosis of noise on lead electrograms. Europace 2008; 10:914–917. [CrossRef] [PubMed]
  41. Perings C, Bauer WR, Bondke HJ, et al. Remote monitoring of implantable-cardioverter defibrillators: results from the Reliability of IEGM Online Interpretation (RIONI) study. Europace 2011; 13:221–229. [CrossRef] [PubMed]
  42. Packer DL, Prutkin JM, Hellkamp AS, et al. Impact of implantable cardioverter-defibrillator, amiodarone, and placebo on the mode of death in stable patients with heart failure: analysis from the sudden cardiac death in heart failure trial. Circulation 2009; 120:2170–2176. [CrossRef] [PubMed]
  43. Poole JE, Johnson GW, Hellkamp AS, et al. Prognostic importance of defibrillator shocks in patients with heart failure. N Engl J Med 2008; 359:1009–1017. [CrossRef] [PubMed]
  44. Sweeney MO, Sherfesee L, DeGroot PJ, Wathen MS, Wilkoff BL. Differences in effects of electrical therapy type for ventricular arrhythmias on mortality in implantable cardioverter-defibrillator patients. Heart Rhythm 2010; 7:353–360. [CrossRef] [PubMed]
  45. Daubert JP, Zareba W, Cannom DS, et al. Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors, and survival impact. J Am Coll Cardiol 2008; 51:1357–1365. [CrossRef] [PubMed]
  46. Braunschweig F, Boriani G, Bauer A, et al. Management of patients receiving implantable cardiac defibrillator shocks: recommendations for acute and long-term patient management. Europace 2010; 12:1673–1690. [CrossRef] [PubMed]
  47. Theuns DA, Rivero-Ayerza M, Boersma E, Jordaens L. Prevention of inappropriate therapy in implantable defibrillators: a meta-analysis of clinical trials comparing single-chamber and dual-chamber arrhythmia discrimination algorithms. Int J Cardiol 2008; 125:352–357. [CrossRef] [PubMed]
  48. Wathen MS, DeGroot PJ, Sweeney MO, et al. Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneous rapid ventricular tachycardia in patients with implantable cardioverter-defibrillators: Pacing Fast Ventricular Tachycardia Reduces Shock Therapies (PainFREE Rx II) trial results. Circulation 2004; 110:2591–2596. [CrossRef] [PubMed]
  49. Wathen MS, Sweeney MO, DeGroot PJ, et al. Shock reduction using antitachycardia pacing for spontaneous rapid ventricular tachycardia in patients with coronary artery disease. Circulation 2001; 104:796–801. [CrossRef] [PubMed]
  50. Ferreira-Gonzalez I, Dos-Subira L, Guyatt GH. Adjunctive antiarrhythmic drug therapy in patients with implantable cardioverter defibrillators: a systematic review. Eur Heart J 2007; 28:469–477. [CrossRef] [PubMed]
  51. Aliot EM, Stevenson WG, Almendral-Garrote JM, et al. EHRA/HRS Expert Consensus on Catheter Ablation of Ventricular Arrhythmias: developed in a partnership with the European Heart Rhythm Association (EHRA), a Registered Branch of the European Society of Cardiology (ESC), and the Heart Rhythm Society (HRS); in collaboration with the American College of Cardiology (ACC) and the American Heart Association (AHA). Heart Rhythm 2009; 6:886–933. [CrossRef] [PubMed]
  52. Leclercq C, Padeletti L, Cihak R, et al. Incidence of paroxysmal atrial tachycardias in patients treated with cardiac resynchronization therapy and continuously monitored by device diagnostics. Europace 2010; 12:71–77. [CrossRef] [PubMed]
  53. Marijon E, Jacob S, Mouton E, et al. Frequency of atrial tachyarrhythmias in patients treated by cardiac resynchronization (from the Prospective, Multicenter Mona Lisa Study). Am J Cardiol 2010; 106:688–693. [CrossRef] [PubMed]
  54. Lazarus A. Remote, wireless, ambulatory monitoring of implantable pacemakers, cardioverter defibrillators, and cardiac resynchronization therapy systems: analysis of a worldwide database. Pacing Clin Electrophysiol 2007; 30(Suppl 1):S2–S12. [CrossRef] [PubMed]
  55. Santini M, Gasparini M, Landolina M, et al. Device-detected atrial tachyarrhythmias predict adverse outcome in real-world patients with implantable biventricular defibrillators. J Am Coll Cardiol 2011; 57:167–172. [CrossRef] [PubMed]
  56. Ricci RP, Morichelli L, Santini M. Remote control of implanted devices through Home Monitoring technology improves detection and clinical management of atrial fibrillation. Europace 2009; 11:54–61. [CrossRef] [PubMed]
  57. Glotzer TV, Daoud EG, Wyse DG, et al. The relationship between daily atrial tachyarrhythmia burden from implantable device diagnostics and stroke risk: the TRENDS study. Circ Arrhythm Electrophysiol 2009; 2:474–480. [CrossRef] [PubMed]
  58. Ricci RP, Morichelli L, Gargaro A, Laudadio MT, Santini M. Home monitoring in patients with implantable cardiac devices: is there a potential reduction of stroke risk? Results from a computer model tested through monte carlo simulations. J Cardiovasc Electrophysiol 2009; 20:1244–1251. [CrossRef] [PubMed]
  59. Ip J, Waldo AL, Lip GY, et al. Multicenter randomized study of anticoagulation guided by remote rhythm monitoring in patients with implantable cardioverter-defibrillator and CRT-D devices: Rationale, design, and clinical characteristics of the initially enrolled cohort The IMPACT study. Am Heart J 2009; 158:364–370 e361. [CrossRef] [PubMed]
  60. Crossley GH, Chen J, Choucair W, et al. Clinical benefits of remote versus transtelephonic monitoring of implanted pacemakers. J Am Coll Cardiol 2009; 54:2012–2019. [CrossRef] [PubMed]
  61. De Ruvo E, Gargaro A, Sciarra L, et al. Early detection of adverse events with daily remote monitoring versus quarterly standard follow-up program in patients with CRT-D. Pacing Clin Electrophysiol 2011; 34:208–216. [CrossRef] [PubMed]
  62. Ricci RP, Morichelli L, Santini M. Home monitoring remote control of pacemaker and implantable cardioverter defibrillator patients in clinical practice: impact on medical management and health-care resource utilization. Europace 2008; 10:164–170. [CrossRef] [PubMed]
  63. Santini M, Ricci RP, Lunati M, et al. Remote monitoring of patients with biventricular defibrillators through the CareLink system improves clinical management of arrhythmias and heart failure episodes. J Interv Card Electrophysiol 2009; 24:53–61. [CrossRef] [PubMed]
  64. Al-Khatib SM, Piccini JP, Knight D, Stewart M, Clapp-Channing N, Sanders GD. Remote monitoring of implantable cardioverter defibrillators versus quarterly device interrogations in clinic: results from a randomized pilot clinical trial. J Cardiovasc Electrophysiol 2010; 21:545–550. [CrossRef] [PubMed]
  65. Hunt SA, Abraham WT, Chin MH, et al. 2009 focused update incorporated into the ACC/AHA 2005 Guidelines for the Diagnosis and Management of Heart Failure in Adults: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines: developed in collaboration with the International Society for Heart and Lung Transplantation. Circulation 2009; 119:e391–479. [CrossRef] [PubMed]
  66. Wolfel EE. Can we predict and prevent the onset of acute decompensated heart failure? Circulation 2007; 116:1526–1529. [CrossRef] [PubMed]
  67. Abraham WT, Compton S, Haas G, et al. Intrathoracic Impedance vs Daily Weight Monitoring for Predicting Worsening Heart Failure Events: Results of the Fluid Accumulation Status Trial (FAST). Congest Heart Fail 2011; 17:51–55. [CrossRef] [PubMed]
  68. Catanzariti D, Lunati M, Landolina M, et al. Monitoring intrathoracic impedance with an implantable defibrillator reduces hospitalizations in patients with heart failure. Pacing Clin Electrophysiol 2009; 32:363–370. [CrossRef] [PubMed]
  69. Wang L. Fundamentals of intrathoracic impedance monitoring in heart failure. Am J Cardiol 2007; 99(10A):3G–10G. [CrossRef] [PubMed]
  70. Yu CM, Wang L, Chau E, et al. Intrathoracic impedance monitoring in patients with heart failure: correlation with fluid status and feasibility of early warning preceding hospitalization. Circulation 2005; 112:841–848. [CrossRef] [PubMed]
  71. Vanderheyden M, Houben R, Verstreken S, et al. Continuous monitoring of intrathoracic impedance and right ventricular pressures in patients with heart failure. Circ Heart Fail 2010; 3:370–377. [CrossRef] [PubMed]
  72. Tomasi L, Zanotto G, Zanolla L, et al. Physiopathologic correlates of intrathoracic impedance in chronic heart failure patients. Pacing Clin Electrophysiol 2011; 34:407–413. [CrossRef] [PubMed]
  73. Jhanjee R, Templeton GA, Sattiraju S, et al. Relationship of paroxysmal atrial tachyarrhythmias to volume overload: assessment by implanted transpulmonary impedance monitoring. Circ Arrhythm Electrophysiol 2009; 2:488–494. [CrossRef] [PubMed]
  74. Ip JE, Cheung JW, Park D, et al. Temporal associations between thoracic volume overload and malignant ventricular arrhythmias: a study of intrathoracic impedance. J Cardiovasc Electrophysiol 2011; 22:293–299. [CrossRef] [PubMed]
  75. Adamson PB, Smith AL, Abraham WT, et al. Continuous autonomic assessment in patients with symptomatic heart failure: prognostic value of heart rate variability measured by an implanted cardiac resynchronization device. Circulation 2004; 110:2389–2394. [CrossRef] [PubMed]
  76. Perego GB, Landolina M, Vergara G, et al. Implantable CRT device diagnostics identify patients with increased risk for heart failure hospitalization. J Interv Card Electrophysiol 2008; 23:235–242. [CrossRef] [PubMed]
  77. Whellan DJ, Ousdigian KT, Al-Khatib SM, et al. Combined heart failure device diagnostics identify patients at higher risk of subsequent heart failure hospitalizations: results from PARTNERS HF (Program to Access and Review Trending Information and Evaluate Correlation to Symptoms in Patients With Heart Failure) study. J Am Coll Cardiol 2010; 55:1803–1810. [CrossRef] [PubMed]
  78. Vollmann D, Nagele H, Schauerte P, et al. Clinical utility of intrathoracic impedance monitoring to alert patients with an implanted device of deteriorating chronic heart failure. Eur Heart J 2007; 28:1835–1840. [CrossRef] [PubMed]
  79. Conraads VM, Tavazzi L, Santini M, et al. Sensitivity and positive predictive value of implantable intrathoracic impedance monitoring as a predictor of heart failure hospitalizations: the SENSE-HF trial. Eur Heart J 2011; (in press). [CrossRef] [PubMed]
  80. Ypenburg C, Bax JJ, van der Wall EE, Schalij MJ, van Erven L. Intrathoracic impedance monitoring to predict decompensated heart failure. Am J Cardiol 2007; 99:554–557. [CrossRef] [PubMed]
  81. Sarkar S, Hettrick DA, Koehler J, et al. Improved Algorithm to Detect Fluid Accumulation via Intrathoracic Impedance Monitoring in Heart Failure Patients with Implanable Devices. Journal of Cardiac Failure 2011; (in press). [CrossRef] [PubMed]
  82. Hasan A, Paul V. Telemonitoring in chronic heart failure. Eur Heart J 2011; article in press. [CrossRef] [PubMed]
  83. Mullens W, Oliveira LP, Verga T, Wilkoff BL, Wilson Tang WH. Insights from internet-based remote intrathoracic impedance monitoring as part of a heart failure disease management program. Congest Heart Fail 2010; 16:159–163. [CrossRef] [PubMed]
  84. Ewald GA, Gilliam FR, Sweeney RJ. Automated HF Decompensation Detection: Results from the Decompensation Detection Study (DECODE) [abstract]. Journal of Cardiac Failure 2009; 15(Suppl):S122.
  85. Braunschweig F, Ford I, Conraads V, et al. Can monitoring of intrathoracic impedance reduce morbidity and mortality in patients with chronic heart failure? Rationale and design of the Diagnostic Outcome Trial in Heart Failure (DOT-HF). Eur J Heart Fail 2008; 10:907–916. [CrossRef] [PubMed]
  86. Ritzema J, Troughton R, Melton I, et al. Physician-directed patient self-management of left atrial pressure in advanced chronic heart failure. Circulation 2010; 121:1086–1095. [CrossRef] [PubMed]
  87. Abraham WT, Adamson PB, Bourge RC, et al. Wireless pulmonary artery haemodynamic monitoring in chronic heart failure: a randomised controlled trial. Lancet 2011; 377(9766):658–666. [CrossRef] [PubMed]
  88. Bourge RC, Abraham WT, Adamson PB, et al. Randomized controlled trial of an implantable continuous hemodynamic monitor in patients with advanced heart failure: the COMPASS-HF study. J Am Coll Cardiol 2008; 51:1073–1079. [CrossRef] [PubMed]
  89. Adamson PB, Conti JB, Smith AL, et al. Reducing events in patients with chronic heart failure (REDUCEhf) study design: continuous hemodynamic monitoring with an implantable defibrillator. Clin Cardiol 2007; 30:567–575. [CrossRef] [PubMed]
  90. Nielsen JC, Kottkamp H, Zabel M, et al. Automatic home monitoring of implantable cardioverter defibrillators. Europace 2008; 10:729–735. [CrossRef] [PubMed]
  91. Theuns DA, Rivero-Ayerza M, Knops P, Res JC, Jordaens L. Analysis of 57,148 transmissions by remote monitoring of implantable cardioverter defibrillators. Pacing Clin Electrophysiol 2009; 32(Suppl 1):S63–65. [CrossRef] [PubMed]