Journal of Innovation in Cardiac Rhythm Management
Articles Articles 2011 June

Telemonitoring in Patients with Congestive Heart Failure and Indication for ICD-Cardiac Resynchronization Therapy: TRIAGE-CRT

DOI: 10.19102/icrm.2011.020606

1NEAL A. CHATTERJEE, MD, 2JOHN J. GALLAGHER, MD, 3JOHN IP, MD, 4JOSEPH POKU, MD, 5RAJESH MALIK, MD, 6JOSEPH AKAR, MD, 7KEVIN MITCHELL, RN, BS, 1JAGMEET SINGH, MD, PhD ON BEHALF OF THE TRIAGE-CRT INVESTIGATORS

1Division of Cardiology, Department of Medicine, Massachusetts General Hospital, Boston, MA
2Cardiology Consultants, Spartanburg, SC
3Thoracic and Cardiovascular Institute, Lansing, MI
4Georgia Arrhythmia Consultants, Macon, GA
5Pee Dee Cardiology, Florence, SE
6Yale University School of Medicine, New Haven, CT
7Biotronik Inc., Lake Oswego, OR

PDF Download PDF
tweeter Follow Us >>

ABSTRACT.Aims: Results from previous studies suggest that remote monitoring may benefit patients with heart failure (HF). The TRIAGE-CRT study prospectively evaluates the combined use of cardiac resynchronization therapy defibrillator (CRT-D)-based home monitoring (HM) in conjunction with weight and blood pressure (BP) electronic telemonitoring (ETM) in patients with HF. Methods and Results: 66 patients were enrolled and followed up at 3 and 6 months post implant with interim follow-up as necessary. Weight and BP were evaluated by weekly telemonitoring starting within 7–14 days post enrollment. Enrolled patients had depressed left ventricle (LV) function (age 65 ± 11 years, left ventricular ejection fraction (LVEF) 24 ± 7%, systolic resting BP 120 ± 18 mm Hg, diastolic resting BP 71 ± 15 mm Hg, etiology of HF: ischemic 58%, non-ischemic 42%). The average number of days with successful telemonitoring transmission was significantly higher for HM (135.4 ± 6.6) when compared to weight (117.7 ± 6.2) and BP (111.8 ± 6.2) ETM (p for comparison = 0.010 and 0.011, respectively). Transmission success, defined as the ratio of actual per expected daily transmission, was also higher for HM (0.806 ± 0.030) versus weight (0.683 ± 0.035) and BP (0.685 ± 0.037) (p for comparison = 0.008 and 0.011, respectively). Wireless transfer of CRT device event notifications resulted in more frequent actionable visits compared to BP or weight transmissions. Conclusions: TRIAGE-CRT demonstrates that transfer of CRT-D-based diagnostics was more frequent than weight and BP transmissions. Continuous HM in the CRT patient enables timely intervention, facilitating both remote and office-based interventions.

KEYWORDS.cardiac resynchronization therapy, heart failure, telemonitoring.

Dr. Chatterjee reports no disclosures; Dr. Gallagher reports that he receives research support from Biotronik and serves on a local institutional review board; Dr. Ip reports that he receives research grants from St. Jude Medical, Biotronik, Medtronic and Atricure, serving on advisory boards for St. Jude Medical, Biotronik and Atricure, and receiving lecture fees for Biotronik, Medtronic, Astra Zeneca and Atricure; Dr. Poku reports serving as a speaker for Sanofi Aventis and Boehringer Ingelheim; Dr. Malik reports serving as a consultant for Sorin and St. Jude Medical; Dr. Akar reports receiving research support from Boston Scientific Corp., serving as a consultant to Biosense-Webster and Biotronik, and as a speaker for St. Jude Medical; K. Mitchell is a paid employee of Biotronik, however, no bonuses are tied to clinical study outcomes; Dr. Singh reports receiving research grants from St. Jude Medical, Medtronic Inc., Boston Scientific Corp. and Biotronik, serving on advisory boards or as a consultant for Boston Scientific Corp., CardioInsight. Thoratec Inc., St. Jude Medical and Medtronic and receiving lecture fees from Medtronic Inc., Biotronik, Guidant Corp., St. Jude Medical and Sorin Group. In addition to the individual disclosures noted above, the institutions of several authors (Drs. Chatterjee, Gallagher, Akar, Poku, Ip, Malik, and Singh) participated as an enrolling site in TRIAGE-CRT and received research support from Biotronik for conduct of the trial.
Manuscript received March 31, 2011, final version accepted May 19, 2011.

Address correspondence to: Jagmeet Singh, MD, Cardiac Arrhythmia Service, GRB 109 Massachusetts General Hospital Heart Center, Harvard Medical School, 55 Fruit Street, Boston, MA 02412, USA. E-mail: jsingh@partners.org

Introduction

Device-based interventions, including cardiac resynchronization therapy (CRT) and implantable cardioverter-defibrillators (ICD) are an important adjunct to traditional medical therapy for a subset of patients with heart failure (HF). In patients with systolic dysfunction and evidence of mechanical dyssynchrony, cardiac resynchronization therapy significantly improves survival and quality of life.1,2 The addition of defibrillation capabilities (CRT-D) further improves symptoms and diminishes the risk of death.35

However, despite improved survival associated with medical and device-based therapeutics, rising costs and increasing rates of hospital readmission frame a growing need for improved longitudinal management of patients with HF.6 Among several proposed strategies, the role of remote telemonitoring remains controversial.7 In patients with systolic dysfunction, previous prospective studies and two recent meta-analyses suggest remote monitoring of weight, blood pressure (BP) and symptoms, reduces the incidence of HF hospitalization, duration of hospital stay and all-cause mortality.812 In contrast, a more recent randomized trial found no effect of weight and symptom-based telemonitoring on mortality or incidence of HF hospitalization.13

In addition to weight and BP, electrophysiological parameters (e.g. atrial fibrillation-burden, ventricular arrhythmias, heart rate variability) may be more sensitive indicators of clinical decompensation in HF.14 This improved sensitivity could be of particular importance to the advanced HF population, including those with CRT/CRT-D devices.

The role of electrophysiological remote monitoring in the out-patient management of patients with CRT-devices has yet to be established. TRIAGE-CRT, a single-arm, multicenter feasibility study evaluated the transmission efficacy of and correlation between CRT-device-based home Monitoring (HM) and weight and BP electronic telemonitoring (ETM) in patients with HF.

Methods

Patient population

Sixty-six patients were enrolled at 12 centers between November 2006 and January 2008. Patients selected from the investigators' general patient population were enrolled within 45 days of CRT-D implantation (Kronos LV-T or Lumax HF-T, Biotronik, Germany). Eligible patients were adult, ambulatory, and weighed ≤400 lbs. Patients must have been able to follow and comply with the study-related procedures, including sufficient cognitive and reading skills to operate the weight and BP ETM system (Carematix Inc., Chicago, IL) and have access to a home landline telephone connection to transmit HM and ETM system information. Exclusion criteria consisted of participating in another cardiovascular clinical study and a life expectancy of <6 months. The institutional review board at each center approved the protocol, and all patients gave written informed consent.

Study procedures

At enrollment, inclusion and exclusion criteria were verified and clinical status was assessed including weight, BP, antiarrhythmic and HF medications. After enrollment, the center was required to register the subject on the HM website, order the ETM system, and educate the patient on the use of the HM and the ETM systems.

HM event notifications included HF diagnostics as well as monitoring of device integrity (Table 1). Additional notifications were activated for ventricular tachycardia (VT), ventricular fibrillation (VF), supraventricular tachycardia (SVT) and ineffective shock events. For ETM notifications, the investigator was required to activate a maximum weight and BP threshold, both of which were left to provider discretion (i.e. individualized to each patient) (recommended: 140 mm Hg; maximum number of readings: three). Measurements exceeding the indicated maximum weight and BP would be highlighted on the ETM system website for better identification and trigger an alert message via e-mail. In addition to the maximum weight alert, the ETM system activated an alert message if the patients weight increased or decreased by > 2 lbs/day or > 5 lbs/week (Table 1).

Table 1: Activated telemonitoring notifications

crm-02-06-347-t1.jpg

Within 7–14 days of HM registration and ETM system ordering, the investigational centers were required to perform weekly HM and ETM system website evaluations. Each site assigned dedicated personnel (research or device nurses) trained in both HM and ETM systems. When events from either system were flagged, the research personnel notified physician investigators or HF nurse practitioners designated as sub-investigators. The research coordinator at each site participated in direct telephone contact with each research subject. The two website evaluations were conducted on the same day and included evaluating HM event notifications, changes in HM-parameters of clinical relevance, ETM alert messages, clinically relevant weight or BP changes, changes in event trigger settings for HM or ETM, and any actions taken (e.g. subject instructed via phone to change HF medications, antiarrhythmics, dietary plan, oxygen therapy, or if an office visit was scheduled). Additional HM/ETM website evaluations may have been performed at any time with similar documentation to the weekly HM/ETM website evaluations.

The expected days of transmission varied by the status of HM and/or ETM transmission. In patients with successful transmissions, the observation periods for weight, BP and HM started with the respective first transmission. This was done to take into account that it usually took longer for the ETM system to be ordered and installed than for automatic HM transmissions. In patients where neither system transmitted any data, the date of enrollment was taken as the start date of the observation period for all systems. If only one remote system transmitted data (e.g. HM or weight and BP data only), there was no first transmission date available for the non-transmitting system. In such cases, the first transmission date from the corresponding, successfully transmitting system was assumed as the start date of the observation period for the non-transmitting system. Each observation period ended with the date when the patient exited or completed the study.

Required office follow-ups occurred at 3 months and 6 months post implant to assess the patient's weight, BP, and NYHA-class. In addition, the CRT-system was evaluated and changes in HF treatment, HF medications, and antiarrhythmics were recorded, as well as any adverse events. The investigational center was required to document all interim office follow-ups and hospitalizations recording the same information as required for the scheduled follow-ups. Days of data transmissions and observation periods were evaluated individually for each system (weight, BP, and HM) and patient. Days of weight and BP transmissions were retrieved from the Carematix web database server. Days of HM data transmissions were retrieved from an ORACLE database containing the HM raw data.

In patients with successful transmissions, the observation periods for weight, BP, and HM started with the respective first transmission. Each observation period ended with the date when the patient exited or completed the study at the 6-month follow-up.

Results

Study population

Patient characteristics of the 66 patients included in this study are summarized in Table 2. Mean implant duration prior to enrollment was 6.1 ± 1.6 months and all patients had depressed LV function, with a mean LVEF of 24 ± 7%. The mean systolic resting BP was 120 ± 18 mm Hg and mean diastolic pressure was 71 ± 15 mm Hg. Pertinent comorbidities were present in 52 subjects, the most common being hypertension (42 subjects, 81%) and renal insufficiency (17 subjects, 33%). In total, 13 patients exited the study prior to the 6-month office follow-up (seven withdrew, one lost to follow-up, four deaths). Of the four deaths, one was non-cardiac, one was sudden cardiac, and two were secondary to decompensated HF.

Table 2: Patient Characteristics

crm-02-06-347-t2.jpg

Transmission success rates and patient compliance

The average number of successful transmissions and rates of transmission success (expressed as the ratio of actual per expected days of transmissions) are listed in Table 3. During the study period, the average number of days of successful transmissions was significantly lower for weight (111.7 ± 6.2 days) and BP (111.8 ± 6.2 days) when compared with transmission of HM parameters (135.4 ± 6.6 days) (p for comparison = 0.010 and 0.011, respectively). Expressed as a ratio of successful transmission days to the expected days of transmission (see Methods), rates of transmission success were significantly lower for weight (0.683 ± 0.035) and BP (0.685 ± 0.037) when compared to HM (0.806 ± 0.030) (p for comparison = 0.08 and 0.011, respectively).

Table 3: Compliance rates and transmission success of HM and ETM

crm-02-06-347-t3.jpg

Of the 66 patients, four did not transmit any weight or BP data during the observation period, two patients did not transmit any HM data, and one patient did not transmit any weight, BP, or HM data.

Telemonitoring triggers initiating office follow-ups

Ninety-two (6%) of the 1,473 required telemonitoring website evaluations performed led to physician intervention in the form of: medication change (n=26; 28%), scheduling urgent office visits (n=19; 21%), dietary plan change (n=3; 3%), or other actions (e.g. device reprogramming) (n=44; 48%). Seventeen unscheduled office visits in 12 patients were initiated by HM event notifications, independent from office visits scheduled following required telemonitoring evaluations. There were no unscheduled office visits initiated by alerts from weight and BP telemonitoring. The most common trigger for an unscheduled, HM-triggered office visit was “low percentage of CRT pacing” (n=6; 35%). Other reasons included VT/VF/AF episodes (n=4; 24%) and high PVC counts (n=1; 6%). In four cases, no particular HM event notification was specified. Actions taken at the unscheduled HM-initiated office visits included device reprogramming (n=8; 47%), medication changes (n=7; 41%) and hospital admission (n=2; 12%).

Compliance

All sites in the protocol were required to perform weekly HM and ETM evaluations, in addition to scheduled 3- and 6-month follow-up evaluations. Of the total 1,577 required follow-ups, there was a compliance rate of 96% (1,509/1,577) during the course of the study.

Discussion

The TRIAGE-CRT feasibility study assessed the role of traditional weight and BP telemonitoring in combination with device-based HM in patients with CRT-D devices. This is the first report directly comparing the efficacy of traditional monitoring parameters (weight, BP) to device-based monitoring in the CRT population. Our results demonstrate significantly higher rates of transmission of HM parameters (normalized transmission success: 80.6%) when compared to transmission of weight and BP telemonitoring data (normalized transmission success: 68.3% and 68.5%, respectively). HM parameters were also more likely to trigger an office visit than telemonitoring parameters.

The effectiveness of a remote monitoring intervention depends on several important factors; namely, the population selected for monitoring, the monitoring parameter used as a proxy for clinical decompensation, the intensity of monitoring (continuous versus intermittent), patient compliance with the monitoring intervention, adequate physician or caregiver review of monitoring data, and effective communication of potential therapeutic changes between patient and provider.7 Previously successful telemonitoring strategies in patients with HF used traditional weight and symptom-based algorithms in patients with evidence of systolic dysfunction and Class II–IV symptoms (e.g. WHARF, SPAN-CHF, TEN-HMS).811 These telemonitoring studies required a significant nurse/physician monitoring infrastructure. Rates of patient compliance ranged from 88% to 90% at 6 months (WHARF, TEN-HMS), to 63% at 14 months (TEN-HMS). In contrast, the recent Tele-HF trial, which also used weight and symptom-based monitoring, found no reduction in mortality or HF hospitalization at 6 months.13 Sub-group analysis did not find any marginal benefit for patients stratified by systolic function or HF class. In this latter study, patient adherence to the telemonitoring intervention was 55% at 6 months. Consistent with previous studies, we found rates of transmission for weight and BP data of approximately 68% at 6 months. The transmission rate (81%) of HM parameters in our study was significantly higher than weight/BP transmission and similar in magnitude to compliance rates of previously successful telemonitoring interventions.

In addition to patient compliance, another critical determinant of telemonitoring efficacy is the sensitivity of the monitoring parameters used as a proxy for clinical decompensation. Traditional telemonitoring strategies have used weight and symptom-based triggers.11 More recent invasive hemodynamic data, however, suggest that perturbations of weight may not correlate with the physiologically relevant changes in intra-cardiac pressures that typically precede clinical decompensation.14 In contrast to weight-based changes, electrophysiological perturbations in patients with CRT devices (e.g. transthoracic impedance, physical activity log, heart rate variability) may precede clinical symptoms by several days, better identifying patients at risk for decompensation.15 By example, in the recent PARTNERS-HF study, the addition of CRT-based triggers to a model including age, medications, gender, and diabetic status, improved prognostic accuracy in predicting HF hospitalization.16 In addition, device-specific triggers (e.g. percentage of CRT pacing) may be particularly useful in preventing decompensation in the subset of patients with HF undergoing CRT.

In our cohort, all HM-associated office visits were associated with actionable changes in management. It is nevertheless important to note that any potential benefit from earlier detection of decompensation may be counterbalanced by the risk of overly sensitive triggers and unnecessary office visits. Finally, telemonitoring efficacy is also related to the ease of telemonitoring data review. Our finding of 96% physician compliance with the monitoring intervention suggests that the HM system is a practical and usable telemonitoring tool. Of note, given the method of enrollment, the degree of compliance may have been influenced by selection bias.

TRIAGE-CRT shows that device-related transmission of electrophysiological data is feasible and more successful than transmission of traditional monitoring parameters in patients with HF and CRT-D devices. The recently published TRUST study in patients with ICDs showed similar feasibility and superiority of monitoring when compared to conventional office follow-up.17 Indeed, although retrospective analysis of the ALTITUDE cohort suggests a possible survival benefit for remote monitoring in patients with CRT devices,18 prospective data are lacking. Hemodynamic monitoring using device-based sensor strategies, coupled with the HM described here, could serve as a better indicator of clinical decline when compared to traditional markers such as weight and BP. As a feasibility study, these data cannot establish whether the superior transmission of electrophysiological parameters reduces the incidence of HF decompensation and/or mortality.

Limitations

The small size of our study limited our ability to perform relevant comparative analyses with an ETM-only control group. As TRIAGE-CRT was a feasibility study, it was under-powered to generate meaningful comparisons with historical controls for clinical end-points including HF-hospitalizations and mortality. In addition, as discussed above, HM triggers may precipitate unnecessary office visits. The relative economic burden of unnecessary visits, balanced against the potential benefit of reduced HF hospitalizations, remains an open question in HF telemonitoring for patients with CRT-D devices.

Conclusion

In conclusion, TRIAGE-CRT demonstrates that transmission of CRT-D based diagnostics was more frequent than weight and BP transmissions. Device-based diagnostics have a higher propensity to initiate an office visit when compared to BP and weight transmission. Automated HM in the CRT patient enables timely intervention, facilitating both remote and office-based interventions. Future prospective studies are needed to establish the potential clinical and economic benefits of telemonitoring in patients with advanced HF and CRT-D devices.

References

  1. Abraham WT, Fisher WG, Smith AL, et al. Cardiac resynchronization in chronic heart failure. N Engl J Med 2002; 346:1845–1853. [CrossRef] [PubMed]
  2. Cleland JG, Daubert JC, Erdmann E, et al. The effect of cardiac resynchronization on morbidity and mortality in heart failure. N Engl J Med 2005; 352:1539–1549. [CrossRef] [PubMed]
  3. Bristow MR, Saxon LA, Boehmer J, et al. Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N Engl J Med 2004; 350:2140–2150. [CrossRef] [PubMed]
  4. Moss AJ, Hall WJ, Cannom DS. Cardiac-resynchronization therapy for the prevention of heart-failure events. N Engl J Med 2009; 361:1329–1338. [CrossRef] [PubMed]
  5. Tang ASL, Wells GA, Talajic MT, et al. Cardiac-resynchronization therapy for mild-to-moderate heart failure. N Engl J Med 2010; 363:2385–2395. [CrossRef] [PubMed]
  6. Bueno H, Ross JS, Wang Y, et al. Trends in length of stay and short-term outcomes among Medicare patients hospitalized for heart failure, 1993–2006. JAMA 2010; 303:2141–2147. [CrossRef] [PubMed]
  7. Desai AS and Stevenson LW. Connecting the circle from home to heart-failure disease management. N Engl J Med 2010; 363:2301–2309. [CrossRef] [PubMed]
  8. Kimmelstiel C, Levine D, Perry K, et al. Randomized, controlled evaluation of short- and long-term benefits of heart failure disease management within a diverse provider network (The SPAN-CHF trial). Circulation 2004; 110: 1450–1455. [CrossRef] [PubMed]
  9. Goldberg LR, Piette JD, Walsh MN, et al. Randomized trial of a daily electronic home monitoring system in patients with advanced heart failure: the Weight Monitoring in Heart Failure (WHARF) trial. Am Heart J 2003; 146:705–712. [CrossRef] [PubMed]
  10. Cleland JG, Louis AA, Rigby AS, Janssens U, Balk AH. Noninvasive home telemonitoring for patients with heart failure at high risk of recurrent admission and death (the Trans-European Network-Home-Care Management System (TEN-HMS) Study. J Am Coll Cardiol 2005; 45:1654–1664. [CrossRef] [PubMed]
  11. Inglis SC, Clark RA, McAlister FA, et al. Structured telephone support or telemonitoring programmes for patients with chronic heart failure. Cochrane Database Syst Rev 2010; 8:CD007228. [CrossRef] [PubMed]
  12. Klersy C, De Silvestri A, Gabutti G, Regoli F, Auricchio A. A meta-analysis of remote monitoring of heart failure patients. J Am Coll Cardiol 2009; 54:1683–1694. [CrossRef] [PubMed]
  13. Chaudhry SI, Mattera JA, Curtis JP, et al. Telemonitoring in patients with heart failure. N Engl J Med 2010; 363:2301–2309. [CrossRef] [PubMed]
  14. Zile MR, Bennett TD, St John Sutton M, et al. Transition from chronic compensated to acute decompensated heart failure: pathophysiological insights obtained from continuous monitoring of intracardiac pressures. Circulation 2008; 118:1433–1441. [CrossRef] [PubMed]
  15. Adamson PB, Smith AL, Abraham WT, Kleckner KJ, Stadler RW, Shih A. Continuous autonomic assessment in patients with symptomatic heart failure: prognostic value of heart rate variability measured by an implanted cardiac resynchronization device. Circulation 2004; 110:2389–2394. [CrossRef] [PubMed]
  16. Whellan DJ, Ousdigian KT, Al-Khatib SM, et al. Combined heart failure device diagnostics identifying patients at higher risk of subsequent heart failure hospitalizations: results from PARTNERS HF study. J Am Coll Cardiol 2010; 27:55:1803–1810. [CrossRef] [PubMed]
  17. Varma N, Epstein AE, Irimpen A, Schweikert R, Love C; TRUST Investigators. Efficacy and safety of automatic remote monitoring for implantable cardioverter-defibrillator follow-up: the Lumos-T Safely Reduces Routine Office Device Follow-up (TRUST) trial. Circulation 2010; 122:325–332. [CrossRef] [PubMed]