Journal of Innovation in Cardiac Rhythm Management
Articles Articles 2012 June

Successful Cardiac Resynchronization Defibrillator Placement in a Patient with Tetralogy of Fallot, Widened QRS Duration, and Mechanical Tricuspid Valve Replacement Using Dual-Site Left Ventricular Venous Lead Placement

DOI: 10.19102/icrm.2012.030605

JONATHAN SALCEDO, MD and RAHUL N. DOSHI, MD, FHRS

University of California Irvine, CA

PDF Download PDF
tweeter Follow Us >>

ABSTRACT.Mechanical tricuspid valves present a challenging scenario when patients develop a need for cardiac pacing or defibrillator placement. Total transvenous lead placement has proven to be both safe and efficacious. A 46-year-old male with a history of tetralogy of Fallot and mechanical tricuspid replacement for past endocarditis developed worsening fatigue and shortness of breath. He was found to have developed depressed left ventricular systolic function, widened QRS, and atrial flutter with controlled response and intermittent atrioventricular block. After successful right atrial flutter ablation, pacing leads placed in the anterior interventricular and posterolateral branch veins achieved cardiac resynchronization, thus avoiding having to cross the tricuspid prosthesis. An azygous defibrillator coil and subcutaneous array were utilized for defibrillator therapy and demonstrated adequate defibrillator thresholds. One month after the procedure, the patient demonstrated recovery of his left ventricular systolic function and remains asymptomatic 1 year after device placement with normal function. This represents a novel way of achieving both cardiac resynchronization and defibrillator therapy with a mechanical tricuspid valve.

KEYWORDS.congenital heart disease, cardiac resynchronization therapy, tricuspid valve prosthesis.

The authors report no conflicts of interest for the published content.
Manuscript received February 21, 2012, final version accepted March 13, 2012.

Address correspondence to: Rahul N. Doshi, MDDirector, FHRS of Electrophysiology and Assistant Professor of Medicine, University of California, Irvine Fullerton Cardiovascular Medical Group, 2240 N. Harbor Blvd Suite 200, Fullerton, CA 92835. E-mail: rahuldoshimd@mac.com

Introduction

Patients with tricuspid valve replacement pose a difficult dilemma if they require permanent ventricular pacing cardiac, given the challenges in placing a right ventricular endocardial lead placement across the prosthesis. However, the increasing use of coronary sinus lead placement has provided a safe and effective alternative for various conditions requiring single-lead ventricular pacing, bifocal-lead pacing, and even implantable cardioverter-defibrillator (ICD) therapy. We now present a case report of a patient who showed clinical improvement and echocardiographic evidence of recovered systolic left ventricular (LV) function after utilizing the anterior interventricular and lateral cardiac veins for cardiac resynchronization therapy (CRT).

Case report and procedure

A 46-year-old male with a history of tetralogy of Fallot status post traditional childhood repair and a history of infective endocarditis requiring aortic, tricuspid, and pulmonic valve replacements 16 years prior to presentation was referred for fatigue and dyspnea. At the time of his valve surgery, a pacemaker was placed in the subcostal position, and apparently this was abandoned as the patient presumably resumed anterograde conduction. On electrocardiogram he was found to have bifascicular block (right bundle branch block and left posterior inferior hemiblock, QRS duration 178 ms) and atrial flutter with controlled ventricular response. A prior Holter recording revealed a ventricular rate in the 40s–90s bpm with intermittent atrioventricular (AV) block). A transthoracic echocardiogram was thus performed and showed reduced LV systolic function with an ejection fraction (LVEF) of 30%.

He was therefore referred for electrophysiological evaluation, ablation, and possible CRT/ICD therapy. Intracardiac recordings revealed typical counterclockwise atrial flutter with concealed entrainment within the tricuspid valve isthmus. Ablation of the isthmus was successful and uneventful with a return to sinus bradycardia and continued bifascicular block and long first-degree AV block. The following day, an attempt at CRT and ICD implantation commenced. Via his left axillary vein, a coronary sinus lumen catheter was utilized to selectively engage the azygous vein. A defibrillator coil (Medtronic model 69373A, 65C, Medtronic Inc, MN) was then deployed into the azygous vein via a long peel-away sheath and Amplatz wire.

Subsequently, a coronary guide sheath and inner lumen catheter was deployed to the coronary sinus, where contrast venography was performed to locate the anterior interventricular vein and posterolateral branch vein. A LV coronary sinus lead (Guidant Easytrak 2 model 4543, Boston Scientific, St. Paul, MN) was then introduced to the anterior interventricular vein (Figure 1) and secured in place after routine testing. Another coronary guide sheath and inner lumen catheter was then used to sub-selectively engage the posterolateral branch vein. A second LV lead was deployed and secured after testing (Guidant Easytrak 2 model 4542, Boston Scientific), thus achieving a biventricular pacing system with a lead on the anterior septum and a lead on the lateral wall. An active fixation atrial lead (Boston Scientific Model 4136) was next placed in the right atrium along the lateral wall. A biventricular ICD device was connected to the lead system (Boston Scientific model N119). When the single azygous vein coil in the distal coil port of the ICD system failed to defibrillate the patient at maximum output, a subcutaneous array (Guidant Model 0085, St. Paul, MN) was placed along the left anterior axillary line, tunneled to the pocket, and connected to the proximal coil port of the device. This resulted in successful defibrillation in a reverse polarity configuration. The shocking coil impedance for the system was 45 ohms. The defibrillation threshold was between 14 and 20 joules, and therefore the device was programmed at 31 joules for all first shocks.

crm-03-06-829-f1.jpg

Figure 1: Posterior-anterior (a) and lateral chest radiograph post procedure. The large black arrows identify the left ventricular leads in the anterior interventricular vein and posterolateral branch vein. The large white arrow depicts the azygos coil. The open arrows indicate the coils for the subcutaneous array. The thin black arrow marks the right atrial lead. Note the abandoned pacer in the subcostal position from the patient’s prior surgery.

Postoperatively, the patient did very well and was discharged without any incident. Follow-up 1 month later revealed that the patient was asymptomatic and able to perform all his baseline activities. Echocardiogram at that time showed that his LV function had recovered with an ejection fraction of 65%. One-year post procedure, the patient continues to show normal defibrillator function and shocking impedance. He remains asymptomatic with normal LV function and normal prosthetic valve function.

Discussion

Pacing leads placed across prosthetic tricuspid valves can lead to valve damage, acute valve failure, and even death, especially with mechanical tilting-valve replacements. Epicardial lead systems provide an alternate means of pacing;1 however, this involves a more invasive procedure including general anesthesia, thoracotomy, and increased procedural risk. Bai et al 2 first described utilizing coronary veins for LV leads, and since then multiple case reports have shown the efficacy of total transvenous lead placement in tricuspid repair patients with a variety pacing needs.

Lopez and Mihalick3 have demonstrated successful pacing and ICD lead placement in patients with a classic Fontan procedure, tetralogy of Fallot with-third degree AV block,4 and Ebstein’s anomaly,5 all with tricuspid obstacles to endocardial placement. In the patient with the previous Fontan operation and tricuspid atresia, they placed the pacing lead in the anterolateral vein and the ICD lead in the coronary sinus. They utilized the middle cardiac vein for ICD lead placement and a lateral vein for the pacing lead in the patient with tetralogy of Fallot and third degree AV block who had prior prosthetic tricuspid valve place. The same configuration for ICD and pacing was employed as well in the patient with Ebstein’s anomaly and significant congestive heart failure.

In two other patients, one with Ebstein’s anomaly and second-degree AV block6 and another patient with rheumatic tricuspid valve disease and high-grade AV block,7 Lopez and Mihalick 3 placed pacing leads in the anterior interventricular vein and posterolateral veins, similar to the configuration we used in our patient. In both patients, they demonstrated using tissue Doppler echocardiography of the septal and lateral LV walls that there was no significant delay to the peak systolic myocardial velocity. In other words, mechanical LV desynchronization was avoided with this approach.

To our knowledge, our case report demonstrates for the first time successful recovery of LV function after cardiac resynchronization therapy in a patient with a mechanical prosthetic tricuspid valve using a bifocal transvenous LV lead placement system and the combination of an azygous coil and subcutaneous array to achieve defibrillation. The anterior interventricular vein and lateral vein locations for pacing leads appears to optimize the timing of septal and lateral wall contraction, respectively, thus making this a viable alternative to the traditional CRT approach employing right ventricular pacing and a single coronary sinus lead. In addition, this is the first report utilizing an entirely transvenous approach for biventricular defibrillator placement utilizing both an azygos vein coil and a subcutaneous array. This approach is a viable long-term solution to a patient with a mechanical tricuspid valve replacement and cardiomyopathy requiring biventricular ICD therapy.

References

  1. Ricciardi D, DE Asmundis C, Czapla J, LA Meir M, Brugada P, Wellens F. Complete epicardial resynchronization device implantation in a patient who underwent a replacement of mitral and tricuspid valve. Pacing Clin Electrophysiol 2011; doi:10.1111/j.1540-8159.2011.03143.x. [CrossRef] [PubMed]
  2. Bai Y, Strathmore N, Mond H, Grigg L, Hunt D. Permanent ventricular pacing via the great cardiac vein. Pacing Clin Electrophysiol 1994; 17(Pt 1):678–683. [CrossRef] [PubMed]
  3. Lopez JA, Mihalick M. Successful use of a transvenous dual-chamber automatic implantable cardiac defibrillator after a classic Fontan operation for tricuspid atresia. Pacing Clin Electrophysiol 2006; 29:1449–1451. [CrossRef] [PubMed]
  4. Lopez JA, Lufschanowski R. A novel approach to transvenous dual-chamber pacing lead placement and cardiac defibrillator implantation after tricuspid valve replacement. J Cardiovasc Electrophysiol 2008; 19:873–875. [CrossRef] [PubMed]
  5. Lopez JA. Total transvenous approach to pacing and defibrillation after Ebstein’s anomaly. Ann Thorac Surg 2009; 87:303–305. [CrossRef] [PubMed]
  6. Lopez JA, Leachman DR. Successful use of transvenous atrial and bifocal left ventricular pacing in Ebstein’s anomaly after tricuspid prosthetic valve surgery. Ann Thorac Surg 2007; 83:1183–1185. [CrossRef] [PubMed]
  7. Lopez JA, Lufschanowski R. Transvenous bifocal left ventricular pacing after mechanical prosthetic tricuspid valve replacement with use of echocardiography to optimize pacing parameters. J Interv Card Electrophysiol 2007; 18:233–237. [CrossRef] [PubMed]