Staphylococcus Capitis Endocarditis: Living with S.Capitis

DOI: 10.19102/icrm.2012.031007

1EZEQUIEL ZAIDEL, MD, 1DARIO DI TORO, MD, 1LUCÍA KAZELIÁN, MD, 1ROBERTO O. NEME, MD, 2MARÍA I. LESPADA, MD, 3FLORENCIA ARCONDO and 3MARCELO ZYLBERMAN, MD

1Department of Cardiology, Hospital Cosme Argerich, Buenos Aires, Argentina
2Section of Infectious Diseases, Hospital Cosme Argerich, Buenos Aires, Argentina
3Department of Medicine, Hospital Cosme Argerich, Buenos Aires, Argentina

PDF Download PDF
tweeter Follow Us >>

ABSTRACT.We report the case of a 76-year-old white male who underwent Dual Chamber transvenous pacemaker placement in 1999 due to sick sinus disease. He had a history of Chagas disease and chronic renal failure. He underwent a generator replacement in 2008. He presented with a febrile syndrome with probable respiratory focus. He had been prescribed antibiotics, but this treatment was not successful; he was therefore referred to our institution. The patient continued to have mild fever for several days after hospitalization. The blood cultures drawn on admission grew Staphylococcus capitis. A transesophageal echocardiogram was obtained due to suspicions of vegetation on the pacemaker leads. The patient commenced cefalotin, and lead extraction was indicated. During removal of the catheter, the ventricular lead was broken; it was therefore abandoned in the right ventricle. Owing to the high risk of conventional surgery, the lead was abandoned and suppressive antibiotic therapy with cefalexin was indicated. An epicardical catheter was then implanted. The patient was discharged under suppressive antibiotic therapy with cefalexin, and during 2-year follow-up he remained without symptoms or fever.

KEYWORDS.Chagas disease, infective endocarditis, permanent pacemaker, Staphylococcus capitis.

The authors report no conflicts of interest for the published content.
Manuscript received July 15, 2012, final version accepted August 9, 2012.

Address correspondence to: Darío Di Toro, MD, Segurola 4589, Capital Federal CP 1419, Buenos Aires, Argentina. E-mail: ditorodario@yahoo.com.ar

Introduction

The use of pacemakers and defibrillators has been increasing over time in Argentina.1 This increase in the number of devices has led to an increase in device-associated endocarditis. The estimated incidence of infection is between 0.13% and 19.9% for pacemakers. Endocarditis associated with devices constitutes 10% of all device infections.2 Endocarditis is associated with higher mortality, morbidity, and an increase in hospital costs.3

Staphylococcus capitis endocarditis in patients with pacemakers is rare, with very few cases reported in the literature. We observed a patient with S. capitis endocarditis secondary to an implanted pacemaker in whom the catheter could not be removed.

Case report

A 76-year-old white male underwent DDD transvenous pacemaker placement in 1999 due to sick sinus disease. He had a history of Chagas disease and chronic renal failure. He underwent a generator replacement in 2008. He presented with a febrile syndrome with probable respiratory focus; the CT scan showed pulmonary nodules, and the biopsy was negative for neoplastic cells and positive for inflammatory cells (Figure 1). He had been unsuccessfully treated with antibiotics, so he was referred to our institution in May 2009.

crm-03-10-982-f1.jpg

Figure 1: Computed tomography scan showing pulmonary nodules interpreted as multiple septic emboli.

The patient continued to have a mild fever for a few days after hospitalization. The blood cultures drawn on admission to the hospital grew S. capitis. A transesophageal echocardiogram was performed to check for suspected vegetation on the pacemaker leads. The patient commenced cefalotin and lead extraction was indicated.

The echocardiogram showed (Figure 2) a mild increase in left ventricle (LV) size, mild decrease in LV systolic function, severe increase in right ventricular (RV) size, moderate to severe decrease in RV systolic function, severe tricuspid valve regurgitation, and Pulmonary artery systolic pressure of 60 mmHg. The pacemaker lead presented in the right ventricle with a mobile mass of 12×7 mm.

crm-03-10-982-f2.jpg

Figure 2: Echocardiographic subxiphoid view with the abandoned lead in the right ventricle.

During the removal of the catheter, the ventricular lead was broken and it was abandoned in the right ventricle. The infection was microbiologically confirmed by positive cultures from the generator pocket and leads. During hospitalization he suffered acute tubular necrosis due to the antibiotics, so daptomycin was indicated. Owing to the high risk of conventional surgery, the lead was abandoned and suppression antibiotic therapy with cefalexin was indicated. An epicardical catheter was then implanted.

The initial computed topography (CT) images were interpreted as multiple septic emboli, and after appropriate antibiotic treatment the images of the emboli disappeared. RV dysfunction was interpreted as RV overload. The patient was discharged under suppressive antibiotic therapy with cefalexin, and during 2-year follow-up he remained without symptoms or fever.

Discussion

Several studies have identified some characteristics associated with cardiovascular implantable electronic device (CIED) infection. In a single-center case–control study, subjects were more likely to have diabetes mellitus and heart failure, and to have undergone generator replacement; renal dysfunction had the strongest association with device infection.4 This patient had renal dysfunction—one of the most important risk factors of device infection—and a recent generator replacement.

The diagnosis of endocarditis was made using the modified Duke criteria, which includes two positive blood cultures drawn on admission to the hospital and a transesophageal echocardiogram suspicious for vegetation.5 Infection for the same microbiological agent was microbiologically confirmed by positive cultures from the generator pocket and leads.

Pulmonary nodules are rare complications of CIED infections; in some series, the rate of septic pulmonary emboli at admission was 11%. In most cases the diagnosis was made by CT, just like the present patient. Most cases resolved with antibiotic treatment.3

There is abundant evidence that the optimal therapy for CIED combines complete device extraction with a prolonged course of parenteral antibiotics. Reported mortality rates for CIED range from 31% to 66% in cases managed with antibiotics alone; in contrast to 13–21% among patients who undergo complete device removal followed by prolonged treatment with systemic antibiotics.6

Percutaneous techniques are currently the method of choice for transvenous lead extraction; catheter fracture is a rare complication of the percutaneous technique, and is seen in 5% of procedures.7

The risk of incomplete or failed extraction increases with: implant duration, less experienced physicians, ventricular leads implantation, and young patients.8

The patient had undergone a ventricular lead implant 10 years previously, so had two of the risk factors for extraction failure.

The decision to avoid surgical treatment with cardiopulmonary bypass was based on the euroSCORE.9 The calculated euroSCORE mortality for this patient was 37%; therefore, suppression antibiotic therapy was indicated.

The decision to abandon the RV lead was taken because the fracture was proximal, and, to the best of our knowledge, the only way to take out the lead was by conventional surgery with sternotomy. We considered that the risk of conventional surgery was too high and decided to leave the catheter.

With the development of transvenous pacemakers, the incidence of infective endocarditis due to coagulase-negative staphylococci has become more common. S. capitis is present in the normal flora of the skin of the scalp, face, ears, and neck.10 S. capitis accounts for 5% of pathogenic isolates of coagulase-negative staphylococci and has been shown to cause pneumonia, urinary tract infection, catheter-related bacteremia, and cellulitis. Only two cases of pacemaker-related endocarditis caused by S. capitis have been previously described, but the present case is the only one that received suppression antibiotic therapy.10,11 A variety of antibiotic regimens has been used successfully in the management of S. capitis endocarditis. It is generally assumed that S. capitis responds to nafcillin, cephalosporins, and vancomycin with or without rifampin.

Cefalotin was the first-choice treatment, but as the patient suffered acute tubular necrosis the antibiotic was changed to daptomycin.

Daptomycin is a cyclic lipopeptide effective in the treatment of Gram-positive infections, including those caused by multiresistant pathogens. This drug has rapid bactericidal action and low nephrotoxicity. Patients with severe renal failure show a decrease in renal clearance and an increase in the elimination half-life. The presence of renal failure at the start of daptomycin therapy is not associated with an increase in the rates of severe adverse effects. Daptomycin has a good safety and efficacy profile for the treatment of infections in patients with chronic renal insufficiency.12 Daptomycin was administered during the hospitalization and cefalexin was indicated at discharge.

The prognosis of S. capitis endocarditis seems to be good; the two patients with pacemaker endocarditis by S. capitis reported in the literature have survived.10 However, Baman et al13 identified several clinical and echocardiographic variables that identify patients with cardiac device-related infection who were at high-risk of mortality. All-cause 6-month mortality was 18% and independent variables associated with death were systemic embolization, moderate or severe tricuspid regurgitation, abnormal RV function, and abnormal renal function.13 The present patient had all these clinical variables, but in the reported series there was no S. capitis endocarditis. To the best of our knowledge there are no data about S. capitis endocarditis and suppressive treatment, which is why follow-up of this patient is so important. Suppressive antibiotic treatment may be an option for patients with high risk of conventional surgery, especially in patients with incomplete or failed extraction.

References

  1. Di Toro D, Borracci R, Rubio M, Serra J. Estimación del numero de marcapasos y cardiodesfibriladores implantados anualmente en la Republica Argentina. Rev Argent Cirug 2009; 97:85–90.
  2. Baddour LM, et al. Update on cardiovascular implantable electronic device infections and their management: a scientific statement from the American Heart Association. Circulation 2010; 121:458–477. [CrossRef] [PubMed]
  3. Sohail MR, Uslan DZ, Khan AH, Friedman PA, Hayes DL, Wilson WR, et al. Infective endocarditis complicating permanent pacemaker and implantable cardioverter-defibrillator infection. Mayo Clin Proc 2008; 83:46–53. [CrossRef] [PubMed]
  4. Bloom H, Heeke B, Leon A, Mera F, Delurgio D, Beshai J, Langberg J.. Renal insufficiency and the risk of infection from pacemaker or defibrillator surgery. Pacing Clin Electrophysiol 2006; 29:142–145. [CrossRef] [PubMed]
  5. Li JS, Sexton DJ, Mick N, Nettles R, Fowler VG Jr, Ryan T, Bashore T, Corey GR. Proposed modifications to the Duke criteria for the diagnosis of infective endocarditis. Clin Infect Dis 2000; 30:633–638. [CrossRef] [PubMed]
  6. del Río A, Anguera I, Miró JM, Mont L, Fowler VG Jr, Azqueta M, Mestres CA; Hospital Clínic Endocarditis Study Group. Surgical treatment of pacemaker and defibrillator lead endocarditis: the impact of electrode lead extraction on outcome. Chest 2003; 124:1451–1459. [PubMed]
  7. Byrd CL, Wilkoff BL, Love CJ, Sellers TD, Turk KT, Reeves R, et al. Intravascular extraction of problematic or infected permanent pacemaker leads: 1994–1996. U.S. Extraction Database, MED Institute. Pacing Clin Electrophysiol 1999; 22:1348–1357. [CrossRef] [PubMed]
  8. Smith M. C., Love C. J.. Extraction of transvenous pacing and ICD leads. Pacing Clin Electrophysiol 2008; 31:736–752. [CrossRef] [PubMed]
  9. Kobayashi KJ, Williams JA, Nwakanma LU, Weiss ES, Gott VL, Baumgartner WA, Conte JV.l. EuroSCORE predicts short- and mid-term mortality in combined aortic valve replacement and coronary artery bypass patients. J Card Surg 2009; 24:637–643. [CrossRef] [PubMed]
  10. Sandoe JA, Kerr KG, Reynolds GW, Jain S. Staphylococcus capitis endocarditis: two cases and review of the literature. Heart 1999; 82:e1. [CrossRef] [PubMed]
  11. Cone LA, Sontz EM, Wilson JW, Mitruka SN. Staphylococcus capitis endocarditis due to a transvenous endocardial pacemaker infection: case report and review of Staphylococcus capitis endocarditis. Int J Infect Dis 2005; 9:335–339. [CrossRef] [PubMed]
  12. Sarriá Cepeda C. [Daptomycin in the treatment of Gram-positive infections in patients with chronic renal failure]. Enferm Infec Microbiol Clin 2012; 30(Suppl 1):38–42. [CrossRef] [PubMed]
  13. Baman TS, Gupta SK, Valle JA, Yamada E. Risk factors for mortality in patients with cardiac device-related infection. Circ Arrhythm Electrophysiol 2009;2:129–134. [CrossRef] [PubMed]