Journal of Innovation in Cardiac Rhythm Management
Articles Articles 2017 June 2017 - Volume 8 Issue 6

Paced or Non-Paced? Examining an Electrocardiogram Rythm

DOI: 10.19102/icrm.2017.080602

MARTHA FERRARA, DNP, RN, FHRS, CCDS1

1Pacemaker Center, The Valley Health System, Ridgewood, NJ

PDF Download PDF
tweeter Follow Us >>

ABSTRACT.Many patients who present for pacemaker follow-up one week after device implantation have a “different” presenting rhythm, which can be puzzling to examine for even the most experienced cardiac device clinicians. Novel implanting techniques on the horizon necessitate a thorough understanding of these new technologies to provide safe care in this patient population.

KEYWORDS.His-bundle pacing, electrocardiogram, right ventricular apex pacing.

The author reports no conflicts of interest for the published content.
Manuscript received February 23, 2017. Manuscript accepted May 18, 2017.
Address correspondence to: Martha Ferrara, DNP, RN, FHRS, CCDS, 1855 State Route 32, Modena, NY 12548, USA. E-mail: mayanrn@optonline.net.

Case presentation

An active 79-year-old male with a history of fatigue and progressive shortness of breath with mild activity was seen in consultation due to his abnormal electrocardiogram (ECG), displaying progressive first-degree atrioventricular (AV) block prolongation over a period of years. An echocardiogram at the current visit revealed an ejection fraction of 55% to 65% with no valvular disease. The patient’s history includes hypertension, coronary artery disease, prior inferior wall myocardial infarction, biopsy-proven senile cardiac amyloidosis, and a six-month history of diastolic heart failure. The following question was raised: is this a pacemaker/paced rhythm?

Discussion

Following additional examination, it was determined that this patient’s rhythm was a ventricular paced rhythm (Figure 1); however, the pacemaker lead was not located in the typical right ventricle (RV) apex (RVA), which produces the typical left bundle branch block pattern on an ECG (Figure 2). Studies have shown that RVA pacing is associated with increased morbidity, mortality, heart failure and the development of persistent atrial fibrillation in comparison with native AV conduction.13

crm-08-06-2740-f1.jpg

Figure 1: The patient’s ECG one week after pacemaker implantation.

crm-08-06-2740-f2.jpg

Figure 2: A dual-chamber pacing ECG.

One of the goals of modern pacemaker technology, aside from its primary aim of maintaining a normal heart rate, is to mimic native conduction as closely as possible, thus avoiding a loss of synchrony and any potential long-term ventricular structural and/or functional changes. At the 36th Heart Rhythm Society Annual Scientific session, multiple abstracts were presented illuminating the “new” pacing method of His-bundle region pacing, though the concept has in fact been around for decades.47 This is an exciting and promising novel technique. In pacing the His-bundle area, ventricular activation follows the pattern of the intrinsic, natural electrical wavefront, depolarizing the ventricles, thus obviating the negative effects of RV pacing.8 The lead is generally placed above the tricuspid valve on the His bundle or Hisian region (Figure 3), or as close to this region as possible (para-Hisian) and above the bundle branches, eliciting and promoting “normal” ventricular activation; hence, a narrow QRS is then seen on the ECG, as is presented in this patient’s presenting paced-ECG shown in Figure 1. The current patient’s “natural” or non-paced ECG present prior to the His-bundle pacemaker implant can be seen in Figure 4, with the note that the width of QRS is very similar to the paced QRS post-His-bundle pacemaker implantation. If His-bundle lead placement or capture is not feasible, a para-Hisian location may be acceptable as noted by the ECG demonstrated in Figure 5, which denotes no isoelectric line between the pacemaker spike and the slurring deflection capturing the His-bundle and local ventricular tissue. Of note, the R-wave and T-wave complexes are concordant (ie, both upright) as is seen in the patient’s natural ECG.

crm-08-06-2740-f3.jpg

Figure 3: The cardiac conduction system. Public Domain Heart Conduction Image. Retrieved 2/12/2017 from https://s-media-cacheak0.pinimg.com/originals/22/de/be/22debe23d28fc684464fca3b31acdb2d.jpg.

crm-08-06-2740-f4.jpg

Figure 4: The patient’s ECG before the His-bundle pacemaker implantation.

crm-08-06-2740-f5.jpg

Figure 5: A para-Hisian His-bundle pacing ECG.

Technical difficulties in lead positioning, a lack of appropriate tools, limited clinical data, and the novelty of the procedure have been cited as some of the barriers to the widespread use of this approach by physicians implanting pacemakers.4,9 His-bundle pacemaker SelectSecure pacing leads (model 3830; Medtronic, Minneapolis, MN, USA) and a fixed-shape catheter (C-315; Medtronic, Minneapolis, MN, USA) used to deliver the pacing lead to the His-bundle region are currently the only such tools on the market relevant for use during this procedure. Up-to-date guidelines for pacemaker implants are followed, with patients selected for His-bundle pacing procedures based on their adherence to the criteria: for example, patients with advanced infiltrative or infra-Hisian disease may not qualify for this type of pacing lead. During implantation, the electrophysiologist can make this assessment, as was demonstrated by Sharma et al.8 in a study that demonstrated the lead’s safety and feasibility without necessity of RV back-up leads or His-bundle mapping catheters. However, new studies also show that careful consideration should still be given to patient selection as is the case with the use of any novel technology.10

Patients presenting to the device clinic for the first follow-up, typically one week post implant for a wound check, have a “different” presenting rhythm. At first, it appears as though no ventricular pacing is present due to the “natural” R-wave look (narrow), which closely mimics the patient’s intrinsic R-wave, especially if the lead is located at the His-bundle (Hisian) region. As such, the device clinician must be aware of the type of pacemaker and lead(s) implanted. The operative report is almost always helpful for confirming device implant technique and location of lead placement. A chest X-ray is also helpful for identifying lead placement, as the ventricular lead shows in a different anatomic position than a typical RV apical lead (Figure 6). During clinical follow-up, a 12-lead ECG performed during threshold measurement can help identify His-bundle and RV threshold capture, though this is normally not done during standard RV pacemaker follow-up evaluation. Device programming is also different from customary settings; notably, R-waves are usually lower in amplitude from the His-bundle region, and thus acceptable programming varies in ventricular sensitivity (less sensitive), threshold (slightly higher) and polarity (unipolar) from typical RVapex settings. The importance of educating the heath-care team to novel technologies in order to deliver appropriate care cannot be overstated.

crm-08-06-2740-f6.jpg

Figure 6: Locations of the His-bundle pacemaker (left) and RV apex pacemaker (right).

References

  1. Wilkoff B, Cook J, Epstein AE, et al. Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial. JAMA. 2002;288(24):3115–3123. [CrossRef] [PubMed]
  2. Sweeney MO, Hellkamp AS, Ellenbogen KA, et al. Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation. 2003;107(23):2932–2937. [CrossRef] [PubMed]
  3. Mulpuru SK, Cha YM, Asirvatham SJ. Synchronous ventricular pacing with direct capture of the atrioventricular conduction system: functional anatomy, terminology, and challenges. Heart Rhythm. 2016;13(11):2237–2246. [CrossRef] [PubMed]
  4. Mandrola J. His-bundle pacing: It’s easier than you think. Medscape. May 15, 2015. Retrieved April 2017 from http://www.medscape.com/viewarticle/844740.
  5. Vijayaraman P, Naperkowski A, Lustgarten DL, et al. Long-term effects of permanent His bundle pacing: are there any adverse effects on distal His-Purkinje conduction? Heart Rhythm Society 2015 Scientific Sessions; May 13, 2015; Boston, MA. Abstract PO01-111.
  6. Vijayaraman P, Sharma PS, Koneru JN, et al. Chronic bundle branch blocks are due to longitudinal dissociation in the main His bundle: electrophysiologic observations from correction of BBB by permanent His bundle pacing. Heart Rhythm Society 2015 Scientific Sessions; May 13, 2015; Boston, MA. Abstract PO01-117.
  7. Vijayaraman P, Dandamudi G, Ellenbogen KA. Electro-physiological observations of acute his bundle injury during permanent His bundle pacing. Heart Rhythm Society 2015 Scientific Sessions; May 13, 2015; Boston, MA. Abstract PO02-185.
  8. Sharma PS, Dandamudi G, Naperkowski A, et al. Permanent His-bundle pacing is feasible, safe, and superior to right ventricular pacing in routine clinical practice. Heart Rhythm. 2015;12(2):305–312. [CrossRef] [PubMed]
  9. Vijayaraman P, Dandamudi G, Worsnick S, Ellenbogen KA. Acute His-bundle injury current during permanent His-bundle pacing predicts excellent pacing outcomes. Pacing Clin Electrophysiol. 2015;38(5)540–546. [CrossRef] [PubMed]
  10. Bhatt A, Musat D, Preminger M, et al. Impact of heart block type and presence of bundle branch block on ability to achieve His-bundle pacing. Heart Rhythm Society 2017 Scientific Sessions; May 10, 2017; Chicago, IL. Abstract P002-25.
 
Banner BSC 09 2024 Resonate JICRM 160x600