SANS FLUORO Optimized: A Case Report of Pulmonary Vein Isolation in a Patient with Cardiac Resynchronization Therapy Defibrillator and the Optimizer™ III Device

DOI: 10.19102/icrm.2020.110502

HYACINTH C. PERCELL, BS,1,2 LYNN SAEGER, BSRN,2 ERIN SHARPE, APRN,2 ROSE SAALFELD, BSRN,2 and ROBERT L. PERCELL, Jr., MD1

1SANS FLUORO Association, Electrophysiology Section, Bryan Heart Institute, Lincoln, NE, USA

2Electrophysiology Section, Bryan Heart Institute, Lincoln, NE, USA

PDF Download PDF
tweeter Follow Us >>

ABSTRACT.We offer the first reported case of a pulmonary vein isolation (PVI) procedure performed in a patient with two devices, specifically a cardiac resynchronization therapy defibrillator (CRT-D) and an Optimizer™ III device (Impulse Dynamics, Mount Laurel, NJ, USA), using the SANS FLUORO technique with zero fluoroscopy. In total, this patient had six leads traversing the right atrium, including two right atrial leads, three right ventricular leads—two associated with the Optimizer™ system and one implantable cardiac defibrillator lead—and a left ventricular lead.

KEYWORDS.Atrial fibrillation, congestive heart failure, implantable cardioverter-defibrillator, Optimizer, pulmonary vein isolation.

Dr. Percell reports that he is on speaker bureaus for Abbott Medical, Biosense Webster, Janssen, Pfizer, and Boehringer Ingelheim, outside the scope of the submitted work. The other authors report no conflicts of interest for the published content. No funding was received for this work.
Manuscript received September 12, 2019. Final version accepted November 18, 2019.
Address correspondence to: Robert L. Percell, Jr., MD, SANS FLUORO Association, Electrophysiology Section, Bryan Heart Institute, 1600 South 48th Street, Lincoln, NE 68506, USA. Email: robertpercelljr@gmail.com.

Introduction

We report the first-ever pulmonary vein isolation (PVI) procedure performed in a patient with two devices, a cardiac resynchronization therapy defibrillator (CRT-D) and an Optimizer™ III device (Impulse Dynamics, Orangeburg, NY, USA) using the SANS FLUORO technique (zero fluoroscopy). This patient had a total of six leads traversing the right atrium: two right atrial (RA) leads, three right ventricular (RV) leads (two for the Optimizer™ III device and one implantable cardioverter-defibrillator lead), and a left ventricular (LV) lead.

Case presentation

The patient was a 69-year-old male with an implantable cardiac monitor, left bundle branch block with an ejection fraction of 25%, and a CRT-D device. He had previously undergone successful implantation of the Optimizer™ III device (Impulse Dynamics, Orangeburg, NY, USA) as part of a clinical trial. Subsequently, he developed severely symptomatic atrial fibrillation (AF) requiring multiple cardioversions.

Procedure

The patient underwent uncomplicated PVI using radiofrequency (RF) via the SANS FLUORO technique. Briefly, groin access was obtained with 7-, 8-, and 9-French (Fr) short sheaths in the right groin. Coronary sinus catheter (Inquiry™; Abbott Laboratories, Chicago, IL, USA), intracardiac echocardiography (ICE) catheter (8-Fr ACUSON Acunav®; Siemens, Berlin, Germany) were adopted, while single transseptal puncture was performed with a long guide catheter (Preface®; Biosense Webster, Diamond Bar, CA, USA) and a HEARTSPAN (Biosense Webster, Diamond Bar, CA, USA) transseptal needle. Right and left atrial mapping was completed with a multielectrode (Advisor HD GRID®; Abbott Laboratories, Chicago, IL, USA) using the EnSite Precision system (Abbott Laboratories, Chicago, IL, USA). PVI was performed with an irrigated contact force-sensing catheter (TactiCath™; Abbott Laboratories, Chicago, IL, USA). Esophageal temperature probe (Medi-Therm® DP400CE Esophageal/Rectal Temperature Probe; Stryker Medical, Kalamazoo, MI, USA) placement was confirmed with the ICE catheter.

There were no issues with catheter placement or transseptal access (Figures 1A1C). The preprocedure chest radiograph from a prior hospitalization is shown in Figure 2. There was increased scar burden in the left atrium (LA) based on voltage mapping, with ablation lesions showing the left atrial roof and posterior line (Figure 3). Post-PVI, entrance and exit blocks were demonstrated. He had no procedural complications and was discharged the following day. The postprocedure CRT-D check revealed no change in lead sensing, threshold, or impedance values. The postprocedure chest radiograph showed no change in lead position (Figure 4). The patient remained free of AF at his three- and six-month follow-up visits.

CRM1206_Percell2-f1.jpg

Figure 1: A: Several device leads in the RV as well as the long guidewire in the inferior vena cava (IVC). B: Transseptal needle with tenting. C: Transseptal sheath in the LA as well as the esophageal temperature probe.

CRM1206_Percell2-f2.jpg

Figure 2: Preprocedure chest radiograph revealing CRT-D and the Optimizer™ III device (Impulse Dynamics, Orangeburg, NY, USA).

CRM1206_Percell2-f3.jpg

Figure 3: Posterior view of the LA revealing four pulmonary veins and the left atrial appendage. Red tags are lesions with lesion size index values of more than 4.5 on the posterior wall and more than 5.0 anteriorly. White, blue, magenta, and green lesions represent impedance decreases of 5%, 10%, 15%, and 20% in ohms, respectively. Yellow points denote the His in the RA. LSPV: left superior pulmonary vein; LIPV: left inferior pulmonary vein; RSPV: right superior pulmonary vein; RIPV: right inferior pulmonary vein; LAAP: left atrial appendage.

CRM1206_Percell2-f4.jpg

Figure 4: Postprocedure anteroposterior chest radiograph, revealing no change occurred in lead position.

Discussion

We report the first PVI procedure in a drug-refractory, persistent, symptomatic AF patient with six leads in the heart using the SANS FLUORO (zero fluoroscopy) technique.1 AF is the most common cardiac arrhythmia and increases dramatically with age. AF frequently coexists with congestive heart failure (CHF), especially in patients with LV dysfunction.2 PVI has been shown to be effective in reducing AF burden and improving mortality in patients with LV dysfunction relative to amiodarone.3

Most PVI procedures are performed with fluoroscopy, whether involving RF or cryoablation, despite recent studies supporting that zero fluoroscopy techniques are feasible, safe, and well-tolerated with similar rates of efficacy.411 Fluoroscopic techniques have been shown to increase the risk of a variety of medical maladies including but not limited to cataracts, head and neck tumors, local erythema, skin desquamation, organ atrophy, birth defects, and bone cancer.12,13 Lead apron use has contributed to multiple orthopedic problems, with nearly half of cardiologists indicating that they are plagued by back, neck, or leg problems.14

The Optimizer™ III (Impulse Dynamics, Orangeburg, NY, USA) is a unique implantable device used in patients with New York Heart Association classes II and III CHF as part of the FIX-HF-5 studies. This rechargeable device delivers impulses to the heart using pacemaker leads during the absolute refractory period in order to modulate cardiac contractility. No mortality benefit has been shown; however; significant improvements in peak ventilatory oxygen uptake and quality of life of patients with moderate to severe heart failure as compared with the effects of the best available medical care were demonstrated.15,16

We believe that all ablation procedures should be performed with zero radiation (SANS FLUORO) to protect patients, operators, and electrophysiology staff from the harmful effects of radiation and protective lead use. In our laboratory, these procedures are commonly performed in device patients without complication. This case provides further evidence that fluoroless complex ablation procedures are safe, even in device patients.

Acknowledgment

Thanks to Timothy Allison, BS, and Travis Swier, BS, at Abbott for their mapping expertise.

References

  1. Percell J, Sharpe E, Percell R. SANS FLUORO (SAy No Series to FLUOROscopy). J Innov Cardiac Rhythm Manage. 2016;7(2016):2529–2534. [CrossRef]
  2. Agner E, Aguinaga L, Andersen HB, et al. Rationale and design of a study assessing treatment strategies of atrial fibrillation in patients with heart failure: the Atrial Fibrillation and Congestive Heart Failure (AF-CHF) trial. Am Heart J. 2002;144(4):597–607. [CrossRef] [PubMed]
  3. Chen MS, Marrouche NF, Khaykin Y, et al. Pulmonary vein isolation for the treatment of atrial fibrillation in patients with impaired systolic function. J Am Coll Cardiol. 2004;43(6):1004–1009. [CrossRef] [PubMed]
  4. Voskoboinik A, Kalman ES, Savicky Y, et al. Reduction in radiation dose for atrial fibrillation ablation over time: a 12-year single-center experience of 2344 patients. Heart Rhythm. 2017;14(6):810–816. [CrossRef] [PubMed]
  5. Reddy VY, Morales G, Ahmed H, et al. Catheter ablation of atrial fibrillation without the use of fluoroscopy. Heart Rhythm. 2010;7(11):1644–1653. [CrossRef] [PubMed]
  6. Wagner A. Interview with Sandeep Gautam, MD, FHRS. J Innov Cardiac Rhythm Manage. 2016;7(7):2435–2437. [CrossRef]
  7. Lickfett L, Mahesh M, Vasamreddy C, et al. Radiation exposure during catheter ablation of atrial fibrillation. Circulation. 2004;110(19):3003–3010. [CrossRef] [PubMed]
  8. Liu X, Palmer J. Outcomes of 200 consecutive, fluoroless atrial fibrillation ablations using a new technique. Pacing Clin Electrophysiol. 2018;41(11):1404–1411. [CrossRef] [PubMed]
  9. Bulava A, Hanis J, Eisenberger M. Catheter ablation of atrial fibrillation using zero-fluoroscopy technique: a randomized trial. Pacing Clin Electrophysiol. 2015;38(7):797–806. [CrossRef] [PubMed]
  10. Knecht S, Sticherling C, Reichlin T, et al. Effective reduction of fluoroscopy duration by using an advanced electroanatomic-mapping system and a standardized procedural protocol for ablation of atrial fibrillation: “the unleaded study”. Europace. 2015;17(11):1694–1699. [CrossRef] [PubMed]
  11. Ernst S, Castellano I. Radiation exposure and safety for the electrophysiologist. Curr Cardiol Rep. 2013;15(10):402. [CrossRef] [PubMed]
  12. Perisinakis K, Damilakis J, Theocharopoulos N, Manios E, Vardas P, Gourtsoyiannis N. Accurate assessment of patient effective radiation dose and associated detriment risk from radiofrequency catheter ablation procedures. Circulation. 2001;104(1):58–62. [CrossRef] [PubMed]
  13. Jacob S, Boveda S, Bar O, et al, Interventional cardiologists and risk of radiation-induced cataract: results of a French multicenter observational study. Int J Cardiol. 2013;167(5):1843–1847. [CrossRef] [PubMed]
  14. Roguin A, Goldstein J, Bar O, Goldstein JA. Brain and neck tumors among physicians performing interventional procedures. Am J Cardiol. 2013;111(9):1368–1372. [CrossRef] [PubMed]
  15. CardioSurve Survey, 2018–2019. Washington, DC: American College of Cardiology; 2019.
  16. Abraham WT, Lindenfeld J, Reddy VY, et al. A randomized controlled trial to evaluate the safety and efficacy of cardiac contractility modulation in patients with moderately reduced left ventricular ejection fraction and a narrow QRS duration: study rationale and design. J Cardiac Fail. 2015;21(1):16–23. [CrossRef] [PubMed]
 
Banner BSC 09 2024 Resonate JICRM 160x600