Journal of Innovation in Cardiac Rhythm Management
Articles Articles 2023 July 2023 - Volume 14 Issue 7

Two-year Outcomes of Ventricular-demand Leadless Pacemaker Therapy for Heart Block After Transcatheter Aortic Valve Replacement

DOI: 10.19102/icrm.2023.14062

DAISUKE TOGASHI, MD,1 KENICHI SASAKI, MD, PhD,1 KAZUAKI OKUYAMA, MD, PhD,1 MASAKI IZUMO, MD, PhD,1 IKUTARO NAKAJIMA, MD, PhD,1 TARO SUCHI, MD,1 YUI NAKAYAMA, MD, PhD,1 TOMOO HARADA, MD, PhD,1 and YOSHIHIRO J. AKASHI, MD, PhD1

1Division of Cardiology, Department of Internal Medicine, St. Marianna University School of Medicine, Kawasaki, Japan

PDF Download PDF
tweeter Follow Us >>

ABSTRACT.Ventricular-demand leadless pacemakers (VVI-LPMs) have often been used as an alternative to atrioventricular (AV) synchronous transvenous pacemakers (DDD-TPMs) in patients with high-grade AV block following transcatheter aortic valve replacement (TAVR). However, the clinical outcomes of this unusual usage are not elucidated. Patients who received permanent pacemakers (PPMs) owing to new-onset high-grade AV block after TAVR from September 2017 to August 2020 at a high-volume center in Japan were included in the analysis, and the clinical courses of VVI-LPM and DDD-TPM implants through 2 years of follow-up were compared retrospectively. Out of 413 consecutive patients who underwent TAVR, 51 (12%) patients received a PPM. After excluding 8 patients with chronic atrial fibrillation (AF), 3 with sick sinus syndrome, and 1 with incomplete data, 17 VVI-LPMs and 22 DDD-TPMs were included in our final cohort. The VVI-LPM group had lower serum albumin levels (3.2 ± 0.5 vs. 3.9 ± 0.4 g/dL, P < .01) than the DDD-TPM group. Follow-up revealed no significant differences between the 2 groups in terms of the incidence of late device-related adverse events (0% vs. 5%, log-rank P = .38) and new-onset AF (6% vs. 9%, log-rank P = .75); however, there were increases in the rates of all-cause death (41% vs. 5%, log-rank P < .01) and heart failure rehospitalization (24% vs. 0%, log-rank P = .01) in the VVI-LPM group. This small retrospective study reveals favorable post-procedural complication rates but higher all-cause mortality with VVI-LPM compared to DDD-TPM therapy for high-grade AV block after TAVR at 2 years of follow-up.

KEYWORDS.Atrioventricular block, heart failure, leadless pacemaker, mortality, TAVR.

The authors report no conflicts of interest for the published content. No funding information was provided.
Manuscript received November 16, 2022. Final version accepted January 23, 2023.
Address correspondence to: Kenichi Sasaki, MD, PhD, Division of Cardiology, Department of Internal Medicine, St. Marianna University School of Medicine, 2-16-1 Sugao, Kawasaki 216-8551, Japan. Email: kenichi.sasaki@marianna-u.ac.jp.

Introduction

In accordance with a current guideline,1 a ventricular-demand (VVI) leadless pacemaker (PM) (LPM) (VVI-LPM) has been used mainly for atrial fibrillation (AF) with a slow ventricular response (AF-SVR), and the perioperative complications are deemed comparable to those of the VVI transvenous PM (TPM) (VVI-TPM).24

On the other hand, an LPM could be an attractive therapeutic option for non-AF bradyarrhythmias, such as sick sinus syndrome (SSS) and advanced or complete atrioventricular (AV) block (high-grade AV block), because the device is used in this context to address lead- and pocket-related complications, including vascular occlusion, lead infection, lead fracture, and pocket infection, involved in TPM therapy.5 Therefore, it is inevitable that a large number of VVI-LPMs have been implanted in patients with high-grade AV block following transcatheter aortic valve replacement (TAVR), which is a major complication of TAVR, with an incidence of 4%–24%.6

Previous articles have reported that dual-chamber TPM (DDD-TPM) implantation for non-AF bradyarrhythmias does not always lead to a better prognosis than VVI-TPM implantation.7,8 In contrast, others have stated that DDD-TPM has more advantages in preventing AF, PM syndrome, and exercise intolerance.911 However, data comparing VVI-LPM and DDD-TPM implantations for non-AF bradyarrhythmias are scarce,12 and there are much fewer data available from comparing these therapies for high-grade AV block after TAVR.13 Therefore, we sought to examine the clinical courses of patients implanted with a VVI-LPM for high-grade AV block following TAVR and to compare them with those of patients implanted with a DDD-TPM.

Methods

Study design and population

This was a single-center, retrospective, observational study. We enrolled 413 patients who underwent TAVR between September 2017 and August 2020 at St. Marianna University Hospital (Kawasaki, Japan). Of these, 51 (12%) class I or II patients (according to the Japanese Circulation Society/Japanese Heart Rhythm Society guidelines1,14) underwent new permanent pacemaker (PPM) implantation after TAVR. After excluding 8 patients with AF-SVR, 3 with SSS, and 1 with incomplete follow-up data, the final study population consisted of 39 patients with new-onset high-grade AV block, including 17 with VVI-LPMs (45%) and 22 with DDD-TPMs (55%) (Figure 1). The study conformed to the ethical guidelines of the Declaration of Helsinki, and the protocol was approved by the institutional ethics committee (reference no. 5582). Patients were allowed to opt-out of the study.

CRM14501_Sasaki-f1.jpg

Figure 1: Flowchart of the study population. Abbreviations: AF, atrial fibrillation; AV, atrioventricular; DDD-TPM, atrioventricular synchronous transvenous pacemaker; PM, pacemaker; pt, patient; SSS, sick sinus syndrome; TAVR, transcatheter aortic valve replacement; VVI-LPM, ventricular-demand leadless pacemaker.

Transcatheter aortic valve replacement procedure

The indications for TAVR were discussed and carefully decided by a team that included cardiologists, cardiac surgeons, cardiac anesthesiologists, clinical engineers, and medical radiology technicians. Details on the TAVR procedure are provided elsewhere.15 Patients were implanted with a balloon-expandable SAPIEN valve (Edwards Lifesciences, Irvine, CA, USA) or self-expandable Evolut™ valve (Medtronic, Minneapolis, MN, USA). After the procedure, all patients were managed in a general high-care unit for ≥1 day. A temporary PM was placed as a backup for high-grade AV block when necessary. If the AV block persisted or was expected to persist for more than several days after TAVR, attending cardiologists decided whether PPM implantation would be performed during hospitalization.

Transvenous pacemaker and leadless pacemaker system

Standard techniques were used for DDD-TPM implantation. Based on the attending physicians’ decision, atrial and ventricular leads were positioned in the right atrial (RA) and right ventricular (RV) appendage or septum, respectively. Micra™ (Medtronic) is the only LPM available in Japan. The target implant site was the RV mid-septal region. When this region was found to have inadequate pacing and/or sensing threshold or technical difficulties for deployment, a site in the apical or high RV septum was chosen.

Electrocardiogram and echocardiography parameters

The 12-lead electrocardiograms (ECGs), used for evaluating the maximum duration of the QRS during RV pacing, and transthoracic echocardiograms (TTEs) were recorded at baseline, 1–6 month(s), 1 year, and 2 years after PPM implantation. During TTE examination, the following data were extracted: left ventricular (LV) ejection fraction (LVEF), LV dimension, the severity of tricuspid valve regurgitation (TR), mitral valve regurgitation (MR), and the estimated RV systolic pressure (RVSP). The LVEF was calculated using the disk summation method. The diameter of the left ventricle was measured at the level of the mitral valve leaflet tips from the parasternal long-axis image. The grading of TR or MR severity was performed according to the American Society of Echocardiography guidelines16 and categorized into the following 4 groups: 0 = none or trivial, 1 = mild, 2 = moderate, and 3 = severe. The degree of TR was based on the color flow jet area in the RA observed using the apical 4-chamber view in addition to the continuous-wave Doppler pattern, vena contracta width, and hepatic vein flow. Continuous-wave Doppler of the TR jet was used to estimate the RVSP in consonance with the inferior vena cava size using the modified Bernoulli equation and RA pressure. For the determination of the degree of MR, quantitative data from color Doppler involving the color flow jet area in the left atrium and pulmonary vein flow were used.

In addition, we evaluated the development of the severity of TR and MR, the degree of LVEF reduction, and prosthetic valve function during 1 and 2 year(s) of follow-up. If the severity of regurgitation worsened or improved by ≥2 grades between the value observed at baseline and the value obtained at the 1- or 2-year follow-up (or the last follow-up if the patient died before the annual visit), the severity was defined as “increased” or “decreased,” respectively. A change in severity within 1 grade was defined as “equal.” The substantial reduction of LVEF was defined as a >10% drop compared to baseline. The assessment of prosthetic valve function was based on the Valve Academic Research Consortium (VARC)-3 definitions.17

Nutritional assessment

Prior to the TAVR procedure, the Mini Nutritional Assessment—Short Form (MNA-SF) was used to assess the nutritional problems of elderly patients. The MNA-SF consists of a simple measurement and 6 questions.18 The survey collected information on physical measurements (body mass index [BMI], weight loss), global assessment (mobility), dietary questions, and subjective factors (dietary intake, neuropsychological problems, acute illness). According to our recent report that the MNA-SF effectively predicted patient mortality after TAVR,19 we focused on this result in the present study.

Follow-up and endpoints

We evaluated clinical outcomes concerning (1) all-cause mortality, (2) rehospitalization for heart failure (HF), (3) late device-related adverse events (AEs), and (4) new-onset AF during the follow-up period. Basically, clinical events were defined according to the VARC-3 criteria.17 Rehospitalization for HF was defined as any new unplanned overnight stay in a hospital because of worsening signs and/or symptoms of HF that required intensification of medical and/or mechanical therapy after PPM implantation. The conclusive diagnosis of HF was made by attending physicians. All late device-related AEs were defined as accidents (eg, device infection, elevated threshold, lead dislodgement, lead failure, or pericardial effusion) that were attributed to the device or implantation procedure and occurred after the patient’s discharge. Elevation of the pacing threshold was defined as an increase by >2-fold from the value when the PPM was implanted. New-onset AF was defined as AF newly documented on a 12-lead ECG during the follow-up period.

Device checks were also performed at 1- and 12-month follow-ups and annually thereafter in an outpatient clinic where the RV pacing burden was acquired. As setting the lower ventricular pacing rate at a higher level led to a higher ventricular pacing percentage (VP%), we adjusted the VP% using the VP% divided by the set lower rate divided by 60 bpm (which was the generally lower rate), thereby calculating the corrected VP% (cVP%). The maximum value of the cVP% was defined as 100%.

Statistical analysis

Continuous variables are presented as mean ± standard deviation values or median (interquartile range [IQR]) values and were compared using Student’s t-test or the Wilcoxon rank-sum test depending on the variable’s distribution. Categorical values are presented as counts and percentages and were compared using Fisher’s exact test. The cumulative incidences of clinical events at follow-up were assessed using the Kaplan–Meier method and log-rank test. Statistical significance was considered achieved at P < .05, and analysis was performed using JMP® 14 (SAS Institute, Cary, NC, USA).

Results

Baseline characteristics

Comparisons of baseline characteristics between the VVI-LPM and DDD-TPM groups are summarized in Table 1. Clinical parameters, such as age, sex, and BMI, were not significantly different between the study groups. Estimated glomerular filtration rate (eGFR) was higher in the VVI-LPM group (71 ± 30 vs. 47 ± 15 mL/min/1.73 m2, P < .01). It seemed that the higher eGFR in the VVI-LPM group might be associated with (1) the operators’ intention to exclude low-eGFR patients from VVI-LPM implantation because it requires more contrast media or (2) a smaller amount of muscle in VVI-LPM patients, who were malnourished; both the serum albumin level (3.2 ± 0.5 vs. 3.9 ± 0.4 g/dL, P < .01) and the MNA-SF score measured preoperatively (5.6 ± 0.8 vs. 8.3 ± 0.8 points, P = .03) were significantly lower in the VVI-LPM group.

Table 1: Patient Characteristics

CRM14501_Sasaki-t1.jpg

Reasons for precluding the use of a transvenous pacemaker

The type of PPM implant (VVI-LPM or DDD-TPM) was determined by the patient’s attending physician or an arrhythmia heart team. Generally, a VVI-LPM was implanted if the following unfavorable conditions for a DDD-TPM were observed: (1) susceptibility to device infection due to the patient’s frailty, steroid use, or other pathology (n = 8); (2) problems with venous access after breast cancer surgery (n = 2); and (3) concern for patients with dementia or delirium who may interfere with the PM pocket (n = 5). In the remaining 2 patients, the indications were based on the physician’s prediction that the treated arrhythmia, like an intermittent AV block, would not require frequent ventricular pacing.

Electrocardiogram, transthoracic echocardiogram, and procedural parameters

Table 2 shows comparisons of ECG, TTE, and procedural parameters between the VVI-LPM and DDD-TPM groups. Both groups did not observe a preoperative left bundle branch block pattern. Echocardiographic parameters also did not show significant differences between the 2 groups. Procedure- or device-related AEs during hospitalization occurred in 5 patients (13%). In the DDD-TPM group, 2 cases of pocket hematoma, 1 case of atrial lead dislodgement, and 1 case of elevated pacing threshold were observed, respectively. On the other hand, 1 case of arteriovenous fistula was observed in the VVI-LPM group.

Table 2: Electrocardiography/Transthoracic Echocardiography/Procedural Parameters

CRM14501_Sasaki-t2.jpg

Changes in echocardiographic and ventricular pacing parameters

Regarding the development of TR and MR and the incidence of reduction of LVEF during 1- and 2-year follow-ups, we found no significant difference between the VVI-LPM and DDD-TPM groups (Table 3). Similarly, no significant differences were observed between these groups concerning post-TAVR prosthetic dysfunctions, such as paravalvular regurgitation and prosthesis–patient mismatch.

Table 3: Changes in Transthoracic Echocardiography Parameters Through 1 and 2 Year(s) of Follow-upa

CRM14501_Sasaki-t3.jpg

In regard to the pacing percentage after TAVR, the cVP% in the VVI-LPM group tended to be lower than that in the DDD-TPM group at the 1-year (52% [IQR, 32%–82%] vs. 75% [IQR, 5%–100%]; P = .09) and 2-year (53% [IQR, 62%–100%] vs. 79% [IQR, 8%–99%]; P = .07) follow-up visits, but the differences were not statistically significant.

Clinical outcomes of patients with ventricular-demand leadless pacemaker or atrioventricular synchronous transvenous pacemaker following transcatheter aortic valve replacement

Clinical outcomes of patients with VVI-LPM or DDD-TPM following TAVR at 1 year of follow-up showed no difference between these groups (Table 4). At 2 years of follow-up, however, there was a higher incidence of all-cause death and HF readmission in the LPM group (Figure 2). During 2 years, all-cause death was observed in 6 cases in the VVI-LPM group and 1 case in the DDD-TPM group (41% vs. 5%, log-rank P < .01). Among them, 3 cases in the VVI-LPM group and 1 case in the DDD-TPM group had sudden cardiovascular death. On the other hand, HF rehospitalization occurred in 4 cases in the VVI-LPM group but none in the DDD-TPM group (24% vs. 0%, log-rank P = .01). The trigger for HF onset was severe mitral stenosis in 1 case, severe TR in 1 case, LV diastolic dysfunction in 1 case, and AV dyssynchrony in 1 case requiring an upgrade to DDD-TPM therapy. Late device-related AEs were observed in just 1 patient in the DDD-TPM group (5% vs. 0%, log-rank P = .38); this patient developed infective endocarditis, which required complete removal of the device system. Finally, new-onset AF was observed in 1 patient in the VVI-LPM group and 2 patients in the DDD-TPM group (6% vs. 9%, log-rank P = .75).

CRM14501_Sasaki-f2.jpg

Figure 2: Kaplan–Meier survival curves for (A) all deaths, (B) HF rehospitalization, (C) late device-related AEs, and (D) new-onset AF. Abbreviations: AE, adverse event; AF, atrial fibrillation; DDD-TPM, atrioventricular synchronous transvenous pacemaker; HF, heart failure; VVI-LPM, ventricular-demand leadless pacemaker.

Table 4: One- and 2-year Clinical Outcomes After Pacemaker Implantationa

CRM14501_Sasaki-t4.jpg

Discussion

Although many previous reports have described the clinical outcomes of VVI-TPM therapy compared to DDD-TPM therapy for high-grade AV block, data on VVI-LPM versus DDD-TPM therapy are scarce and limited in post-TAVR patients. Indeed, the data we previously provided were limited due to the short follow-up duration.13 The present study is more informative, offering more precise data regarding clinical outcomes and the changes in AV valve and LV function through a longer follow-up duration. The main findings of this study were as follows. First, (1) the VVI-LPM group had significantly higher rates of all-cause death and HF rehospitalization compared to the DDD-TPM group at 2 years of follow-up; however, there were no significant differences in the incidence of device-related late AEs and new-onset AF. Second, (2) according to the TTE findings, rates of TR and MR onset and the reduction of LVEF through 1 and 2 year(s) of follow-up after PPM implantation were similar between the study groups.

The present study showed that, during the 2-year follow-up period, there were significant differences appeared between the VVI-LPM and DDD-TPM groups in terms of mortality and HF readmission. On the contrary, the majority of previous studies did not demonstrate a better prognosis of DDD-TPM therapy for high-grade AV block than VVI-TPM therapy despite the hemodynamic superiority of DDD pacing.8,20,21 The potential reason why patients with VVI pacing had poor clinical outcomes in our study is as follows. As described before, preoperative serum albumin and the MNA-SF score were lower in the VVI-LPM group. The meta-analysis revealed that pre-procedural low albumin levels (cut-off value, 3.5 g or 4 g/dL) were significantly associated with short- and mid-term all-cause mortality in patients after TAVR.22 In addition to a lower preoperative serum albumin level of 3.2 ± 0.5 g/dL, the VVI-LPM group had lower MNA-SF scores. Both albumin level and MNA-SF score are useful markers for quantifying nutritional status; thus, it seems reasonable that undernourishment could have an adverse effect on patient prognosis in the VVI-LPM group. Lower albumin levels or MNA-SF scores in the VVI-LPM group might arise from the selection bias by the attending physicians precluding DDD-TPM implantation for apparently frail or malnourished patients. Causes of non-cardiac death, such as infection and cancer, reportedly resulted in a large proportion of deaths after TAVR.23 This finding also might confirm our hypothesis that malnutrition was associated with higher all-cause mortality in the VVI-LPM group. Nevertheless, our data observed no significant difference in clinical frailty scores between the study groups. This seems to be due to the inter-rater reliability or small sample size of our study.

The incidence of HF readmission was higher in the VVI-LPM group than in the DDD-TPM group during the 2-year follow-up period. Beurskens et al. suggested that LPM therapy is associated with biventricular and AV valve dysfunction,24 while our data did not confirm these findings. As mentioned above, there are conflicting descriptions of worsening cardiac and AV valve function after VVI-LPM implantation.12,2426 To eliminate the discrepancy, we need further studies prospectively investigating TTE parameters with the aid of 3-dimensional echocardiography. Other factors assumed to contribute to the occurrence of decompensated HF, such as wider paced QRS or RV pacing burden, did not differ between VVI-LPM and DDD-TPM therapies in the present study. Therefore, it might be possible that the difference in pacing mode, ie, VVI or DDD, was associated with HF development, although previous studies did not support the hypothesis.8,20,21 Then, we focused on the patient’s nutritional condition again. Arquès et al. reported that hypoalbuminemia caused low colloid osmotic pressure and facilitated the onset of pulmonary edema in patients with diastolic HF.27 This finding could explain the reason for HF development regardless of LV or AV valve function deterioration.

There were a few late device-related AEs in the present study population. Among DDD-TPM patients, device infection occurred in 1 case at 5 months, and the patient required complete device removal, whereas no late device-related AEs were observed in the VVI-LPM group. It was noteworthy that the VVI-LPM group did not experience any device infection despite their poorer nutrition. It was reported that LPM is more advantageous for patients at high infection risk than TPM because of the resistance to bacteremia.28,29 This seems to a reason why the VVI-LPM group did not experience significantly more complications than the DDD-TPM group.

It has been frequently described that VVI pacing for high-grade AV block is associated with a higher incidence of new-onset AF compared to DDD pacing.10,11 The reason for increased AF rates with VVI pacing might be because left atrial stretching originating from AV asynchronous RV pacing could increase the pulmonary artery wedge pressure. We did not observe a higher incidence of new AF in the VVI-LPM group; however, this may be responsible for the underestimation of AF due to the inability of LPMs to record arrhythmic events.

Based on these findings, we must discuss whether we should seek to implant DDD-TPMs first in patients with high-grade AV block after TAVR. However, we observed that patients provided with VVI-LPM therapy were likely to be frail and vulnerable; therefore, DDD-TPM implantation in the cohort would result in more device-related AEs and cancel the benefit of physiologic pacing. Considering that the clinical outcomes of both groups were comparable at 1 year of follow-up, we could say that VVI-LPM implantation is reasonable for patients with a limited life expectancy. In any case, it is expected that a recently launched AV synchronous LPM will resolve the dilemma related to device selection, and we should confirm this expectation in the future.

Study limitation

Our study has several limitations. First, it was a single-center, retrospective, observational study with a relatively small sample size. Thus, the results are likely to be statistically underpowered. Second, there was a selection bias. This study was performed on Japanese post-TAVR patients, so the application of its results to subjects with other clinical backgrounds should be completed with caution. Third, an LPM is incapable of recording arrhythmic events, unlike a TPM, so we cannot ignore the possibility of having underestimated new AF in the VVI-LPM group, especially when AF is asymptomatic. Lastly, the complication rate may have been underestimated because experienced operators in our hospital performed VVI-LPM implantations.

Conclusion

In this small retrospective study of elderly Japanese patients with high-grade AV block following TAVR, VVI-LPM therapy worsened overall mortality and the HF readmission rate in comparison to DDD-TPM therapy through 2 years of follow-up. Although the results may be affected by different clinical backgrounds, we should continue to be careful when implanting a VVI-LPM in these patients.

Acknowledgments

We would like to thank Editage (www.editage.com) for English language editing.

References

  1. Nogami A, Kurita T, Kusano K, et al. JCS/JHRS 2021 guideline focused update on non-pharmacotherapy of cardiac arrhythmias. Circ J. 2022;86(2):337–363. [CrossRef] [PubMed]
  2. Cantillon DJ, Dukkipati SR, Ip JH, et al. Comparative study of acute and mid-term complications with leadless and transvenous cardiac pacemakers. Heart Rhythm. 2018;15(7):1023–1030. [CrossRef] [PubMed]
  3. Tjong FVY, Knops RE, Udo EO, et al. Leadless pacemaker versus transvenous single-chamber pacemaker therapy: a propensity score-matched analysis. Heart Rhythm. 2018;15(9):1387–1393. [CrossRef] [PubMed]
  4. Piccini JP, El-Chami M, Wherry K, et al. Contemporaneous comparison of outcomes among patients implanted with a leadless vs transvenous single-chamber ventricular pacemaker. JAMA Cardiol. 2021;6(10):1187–1195. [CrossRef] [PubMed]
  5. Reddy VY, Knops RE, Sperzel J, et al. Permanent leadless cardiac pacing: results of the LEADLESS trial. Circulation. 2014;129(14):1466–1471. [CrossRef] [PubMed]
  6. van Rosendael PJ, Delgado V, Bax JJ. Pacemaker implantation rate after transcatheter aortic valve implantation with early and new-generation devices: a systematic review. Eur Heart J. 2018;39(21):2003–2013. [CrossRef] [PubMed]
  7. Dretzke J, Toff WD, Lip GY, Raftery J, Fry-Smith A, Taylor R. Dual chamber versus single chamber ventricular pacemakers for sick sinus syndrome and atrioventricular block. Cochrane Database Syst Rev. 2004;2004(2):CD003710. [CrossRef] [PubMed]
  8. Toff WD, Camm AJ, Skehan JD. United Kingdom Pacing and Cardiovascular Events Trial Investigators. Single-chamber versus dual-chamber pacing for high-grade atrioventricular block. N Engl J Med. 2005;353(2):145–155. [CrossRef] [PubMed]
  9. Hesselson AB, Parsonnet V, Bernstein AD, Bonavita GJ. Deleterious effects of long-term single-chamber ventricular pacing in patients with sick sinus syndrome: the hidden benefits of dual-chamber pacing. J Am Coll Cardiol. 1992;19(7):1542–1549. [CrossRef] [PubMed]
  10. Kerr CR, Connolly SJ, Abdollah H, et al. Canadian Trial of Physiological Pacing: effects of physiological pacing during long-term follow-up. Circulation. 2004;109(3):357–362. [CrossRef] [PubMed]
  11. Said S, Cooper CJ, Alkhateeb H, et al. Incidence of new onset atrial fibrillation in patients with permanent pacemakers and the relation to the pacing mode. Med Sci Monit. 2014;20:268–273. [CrossRef] [PubMed]
  12. Sasaki K, Togashi D, Nakajima I, et al. Clinical outcomes of non-atrial fibrillation bradyarrhythmias treated with a ventricular demand leadless pacemaker compared with an atrioventricular synchronous transvenous pacemaker – a propensity score-matched analysis. Circ J. 2022;86(8):1283–1291. [CrossRef] [PubMed]
  13. Okuyama K, Izumo M, Sasaki K, et al. Comparison in clinical outcomes between leadless and conventional transvenous pacemaker following transcatheter aortic valve implantation. J Invasive Cardiol. 2020;32(10):400–404. [PubMed]
  14. Nogami A, Kurita T, Abe H, et al. JCS/JHRS 2019 guideline on non-pharmacotherapy of cardiac arrhythmias. Circ J. 2021;85(7):1104–1244. [CrossRef] [PubMed]
  15. Webb JG, Chandavimol M, Thompson CR, et al. Percutaneous aortic valve implantation retrograde from the femoral artery. Circulation. 2006;113:842–850. [CrossRef] [PubMed]
  16. Zoghbi WA, Adams D, Bonow RO, et al. Recommendations for noninvasive evaluation of native valvular regurgitation: a report from the American Society of Echocardiography developed in collaboration with the Society for Cardiovascular Magnetic Resonance. J Am Soc Echocardiography. 2017;30(4):303–371. [CrossRef] [PubMed]
  17. VARC-3 WRITING COMMITTEE, Généreux P, Piazza N, et al. Valve Academic Research Consortium 3: updated endpoint definitions for aortic valve clinical research. Eur Heart J. 2021;42(19):1825–1857. [CrossRef] [PubMed]
  18. Rubenstein LZ, Harker JO, Salvà A, Guigoz Y, Vellas B. Screening for undernutrition in geriatric practice: developing the short-form mini-nutritional assessment (MNA-SF). J Gerontol A Biol Sci Med Sci. 2001;56(6):M366–M372. [CrossRef] [PubMed]
  19. Doi S, Kohei K, Mika W, et al. Prognostic value of mini nutritional assessment-short form with aortic valve stenosis following transcatheter aortic valve implantation. ESC Heart Failure. 2020;7(6):4024–4031. [CrossRef] [PubMed]
  20. Lamas GA, Orav EJ, Stambler BS, et al. Quality of life and clinical outcomes in elderly patients treated with ventricular pacing as compared with dual-chamber pacing. N Engl J Med. 1998;338(16):1097–1104. [CrossRef] [PubMed]
  21. Jahangir A, Shen WK, Neubauer SA, et al. Relation between mode of pacing and long-term survival in the very elderly. J Am Coll Cardiol. 1999;33(5):1208–1216. [CrossRef] [PubMed]
  22. Liu G, Hu X, Long M, Du ZM, Li Y, Hu CH. Meta-analysis of the impact of pre-procedural serum albumin on mortality in patients undergoing transcatheter aortic valve replacement. Int Heart J. 2020;61(1):67–76. [CrossRef] [PubMed]
  23. Van Mieghem NM, van der Boon RM, Nuis RJ, et al. Cause of death after transcatheter aortic valve implantation. Catheter Cardiovasc Interv. 2014;83(7):E277–E282. [CrossRef] [PubMed]
  24. Beurskens NEG, Tjong FVY, de Bruin-Bon RHA, et al. Impact of leadless pacemaker therapy on cardiac and atrioventricular valve function through 12 months of follow-up. Circ Arrhythm Electrophysiol. 2019;12(5):e007124. [CrossRef] [PubMed]
  25. Salaun E, Tovmassian L, Simonnet B, et al. Right ventricular and tricuspid valve function in patients chronically implanted with leadless pacemakers. Europace. 2018;20(5):823–828. [CrossRef] [PubMed]
  26. Sanchez R, Nadkarni A, Buck B, et al. Incidence of pacing-induced cardiomyopathy in pacemaker-dependent patients is lower with leadless pacemakers compared to transvenous pacemakers. J Cardiovasc Electrophysiol. 2021;32(2):477–483. [CrossRef] [PubMed]
  27. Arquès S, Ambrosi P, Gélisse R, Luccioni R, Habib G. Hypoalbuminemia in elderly patients with acute diastolic heart failure. J Am Coll Cardiol. 2003;42(4):712–716. [CrossRef] [PubMed]
  28. El-Chami MF, Soejima K, Piccini JP, et al. Incidence and outcomes of systemic infections in patients with leadless pacemakers: data from the Micra IDE study. Pacing Clin Electrophysiol. 2019;42(8):1105–1110. [CrossRef] [PubMed]
  29. Garweg C, Vandenberk B, Jentjens S, et al. Bacteraemia after leadless pacemaker implantation. J Cardiovasc Electrophysiol. 2020;31(9):2440–2447. [CrossRef] [PubMed]
 
Banner BSC 09 2024 Resonate JICRM 160x600