Journal of Innovation in Cardiac Rhythm Management
Articles Articles 2024 June 2024 - Volume 15 Issue 6

Comparative Profiles of the WATCHMAN™ and Amplatzer™ Cardiac Plug/Amplatzer™ Amulet™ Devices for Left Atrial Appendage Closure in Non-valvular Atrial Fibrillation: A Comprehensive Systematic Review and Meta-analysis

DOI: 10.19102/icrm.2024.15061

FNU RAJA, MBBS,1 KHIMYA RANI, MBBS,2 SUNNY KUMAR, MBBS,3 FNU SOMESHWAR, MBBS,3 MUHAMMAD AHSAN NASEER KHAN, MBBS,4 FNU ABUBAKAR, MBBS,4 DHVANI BHATT, MBBS,5 DEEPAK JUNG SUBEDI, MBBS,6 SUJEET SHADMANI, MBBS,7 and FATIMA TUZ ZAHRA ABDULLAH, MBBS8

1Department of Internal Medicine, Federal Medical and Dental College, Islamabad, Pakistan

2Department of Internal Medicine, Chandka Medical College SMBBMU, Larkana, Pakistan

3Department of Internal Medicine, Liaquat National Hospital and Medical College, Karachi, Pakistan

4Department of Internal Medicine, King Edward Medical University, Lahore, Pakistan

5Department of Internal Medicine, American University of Barbados, Bridgetown, Barbados

6Department of Internal Medicine, College of Medical Sciences, Bharatpur, Nepal

7Department of Internal Medicine, Shaheed Mohtarma Benazir Bhutto Medical College, Karachi, Pakistan

8Department of Internal Medicine, Allama Iqbal Medical College, Lahore, Pakistan

PDF Download PDF
tweeter Follow Us >>

ABSTRACT.Atrial fibrillation (AF) is a prevalent cardiac arrhythmia marked by irregular and frequent tachycardic rhythms in the atria, affecting 1%–2% of the general population. The WATCHMAN™ device from Boston Scientific (Marlborough, MA, USA) and the Amplatzer™ Amulet™ device from Abbott (Chicago, IL, USA) are two devices used globally for left atrial appendage closure (LAAC) in non-valvular AF. A systematic search was conducted in PubMed, the Cochrane Library, and Elsevier’s ScienceDirect literature databases to identify studies comparing the WATCHMAN™ procedure with Amulet™ device implantation for LAAC in patients with AF. The analyses were conducted using the random-effects model. A total of 20 studies were identified, with 18 falling into the category of observational studies and 2 being randomized controlled trials. A total of 6310 participants were included in this meta-analysis, with 3198 individuals (50.68%) assigned to the WATCHMAN™ procedure group and 3112 individuals (49.32%) allocated to the Amplatzer™ Cardiac Plug (ACP) group. The analysis revealed a higher risk of stroke associated with the WATCHMAN™ technique (relative risk [RR], 1.14), albeit without statistical significance. Conversely, the WATCHMAN™ approach led to a significantly lower risk of cardiac death (RR, 0.44; P = .04). Notably, the risks of all-cause mortality (RR, 0.89; 95% confidence interval [CI], 0.73–1.08; I2 = 0%; P = .25) and major bleeding (RR, 0.93; 95% CI, 0.65–1.33; I2 = 31%; P = .70) were clinically reduced with the WATCHMAN™ procedure, although statistical significance was not achieved. Compared to Amulet™ device implantation, WATCHMAN™ device implantation decreased the risk of cardiac mortality, while the risks of stroke, systemic embolism, all-cause mortality, and major bleeding were not statistically significant.

KEYWORDS.Amulet™ device, atrial fibrillation, left atrial appendage closure, meta-analysis, WATCHMAN™ device.

The authors report no conflicts of interest for the published content. No funding information was provided.
Manuscript received January 27, 2024. Final version accepted February 19, 2024.
Address correspondence to: Sujeet Shadmani, MBBS, Department of Internal Medicine, Shaheed Mohtarma Benazir Bhutto Medical College, Karachi, Sindh, Pakistan. Email: sujeetshadmani@gmail.com.

Introduction

Atrial fibrillation (AF) is a common heart rhythm disorder characterized by irregular and rapid beating in the atria. It affects 1%–2% of the general population, 6% of those aged ≥65 years, and 13% of individuals aged >80 years.1 The global incidence and prevalence of AF are rising, having quadrupled in the last 50 years, according to the Framingham Heart Study.2 The left atrial appendage (LAA) is the primary site of embolism in about 90% of non-rheumatic AF-related strokes. This is attributed to the lack of contractile function, leading to blood stasis and thrombus formation. As a result, individuals with AF have a more than fivefold increase in the annual risk of significant stroke compared to the general population.3

LAA closure (LAAC) is a non-pharmacological method for lowering the risk of cardioembolic events in AF patients who are at high risk of stroke.4 While oral anticoagulation can reduce the risk of cardioembolism by approximately 70% when used concurrently, such has been related to an elevated risk of severe bleeding.5 The WATCHMAN™ device and the Amplatzer™ Amulet™ are popular LAAC devices.5 Following the Amulet™ IDE trial, the U.S. Food and Drug Administration recently approved the Amplatzer™ Amulet™. Research demonstrated its superiority in LAA occlusion (LAAO), particularly in minimizing leaks >0.5 mm.6 Comparing the Amulet™ to the previous-generation WATCHMAN™ 2.5 device revealed this superiority. At the same time, it is important to note that the most recent iteration, WATCHMAN™ FLX, demands consideration in current clinical practice due to significant architectural enhancements.6 In observational studies, the Amplatzer™ Cardiac Plug (ACP) and second-generation Amplatzer™ Amulet™ devices have been associated with positive outcomes.7 The WATCHMAN™ device reduces the risk of blood clotting and strokes by obstructing the LAA. It includes a self-expanding nickel–titanium frame enveloped in a permeable fabric placed at the ostium of the LAA through a minimally invasive transseptal approach.8 In contrast, the Amulet™, featuring a lobe-and-disk mechanism, enhances LAA sealing and coverage.8,9

Guidelines recommend these devices for AF patients with a significant stroke risk contraindicated for long-term anticoagulation, having subjected both to rigorous clinical evaluation.10 Despite the publication of observational studies and randomized controlled trials (RCTs), conflicting findings and inadequately conducted analyses—particularly regarding infrequent events like stroke—have been documented. This meta-analysis aims to comprehensively compare the safety and effectiveness of the Amplatzer™ and WATCHMAN™ devices. The fundamental differences in design, deployment methods, and clinical outcomes warrant this thorough comparison. While both devices share the same therapeutic goal, their distinct characteristics may lead to variations in efficacy, safety, and compatibility with patient anatomies. The study analyzes data from RCTs, observational studies, and registries to assess stroke-prevention effectiveness, procedural success, complication rates, and long-term outcomes. Through an in-depth analysis of the merits and demerits of each device, our goal was to address the current knowledge gaps and provide accurate suggestions for device selection.

Methodology

This meta-analysis rigorously adheres to the guidelines outlined in the Preferred Reporting Items for Systematic Reviews and Meta-analyses (PRISMA) statement11 to ensure methodological transparency and reliability.

Data sources and search strategy

A systematic and exhaustive search strategy was implemented to identify relevant studies. Electronic databases, including PubMed, the Cochrane Library, and Elsevier’s ScienceDirect, were systemically searched until December 2023 without language restrictions. The search strategy employed a combination of Medical Subject Heading (MeSH) terms and keywords, such as “Watchman device,” “ACP device,” “Amulet device,” “left atrial appendage closure,” and “non-valvular atrial fibrillation.” The detailed search strategy is summarized in Supplementary Table S1.

Supplementary Table S1: Search Strategy

CRM1583_Shadmani-t2.jpg

Study selection

During study selection, a stringent and comprehensive approach was adopted. Two independent reviewers scrutinized the titles and abstracts of identified studies and evaluated their full texts. Any inconsistencies or disparities in the inclusion of studies were addressed through extensive deliberation and, when necessary, consultation with a third reviewer. The predefined inclusion criteria were rigorously adhered to throughout this selection process. Specifically, the following studies were considered for inclusion: first, studies directly comparing the clinical effectiveness and safety profiles of the WATCHMAN™ device against the ACP/Amulet™ device, specifically in the context of percutaneous LAAC, among patients diagnosed with non-valvular AF; second, studies that encompassed diverse research designs, including RCTs, prospective or retrospective cohort studies, case–control studies, and comparative observational studies; and third, studies that included comprehensive reporting of clinically relevant outcomes, encompassing procedural success rates, adverse events, and long-term follow-up data. This stringent selection process aimed to ensure the incorporation of high-quality and pertinent studies that directly addressed these devices’ comparative efficacy and safety in managing non-valvular AF through percutaneous LAAC.

Data extraction

Two reviewers performed data extraction independently using a standardized data-extraction form. Extracted data included study characteristics (eg, author, publication year, study design), participant demographics, intervention details, procedural success rates, clinical outcomes (eg, stroke, bleeding events), and adverse events. The outcomes were divided into primary outcomes and secondary outcomes. The primary outcomes included stroke, systemic embolism, all-cause mortality, cardiac death, and major bleeding. Meanwhile, the secondary outcomes included device-related thrombus, peri-device leak (PDL), procedure-related complications, and procedural success. Discrepancies were resolved through discussion or by involving a third reviewer when necessary.

Quality assessment

The methodological quality and risk of bias of included studies were assessed independently by two reviewers using appropriate tools, such as the Cochrane risk-of-bias tool for RCTs12 and the Newcastle–Ottawa scale13 for observational studies. Quality assessment covered aspects related to study design, participant selection, blinding, outcome reporting, and other relevant parameters.

Data synthesis and analysis

The statistical analysis of this meta-analysis was conducted using Review Manager version 5.4.1, a tool developed by The Nordic Cochrane Centre in collaboration with The Cochrane Collaboration in Denmark in 2014. Only comparative studies were included in the analysis. The results were presented through forest plots that display the pooled effect of relative risks (RRs) for dichotomous outcomes and weighted mean differences for continuous outcomes. To ensure the accuracy of the findings, the generic-inverse variance with a random-effects model was used.

The P values were set at <.05 to determine the significance of the results. To assess the possibility of publication bias, funnel plots were created for each primary outcome. Higgins’ I2 test was used to determine the level of heterogeneity, which was categorized as low, moderate, or high. When substantial heterogeneity (>75%)14 was detected, a sensitivity analysis was conducted by omitting one study at a time to identify the impact of individual studies on the overall findings.

All analyses were deemed significant if P < .05, and the authors conducted a comprehensive analysis of the data to ensure the accuracy and reliability of their findings.

Ethical considerations

Ethical approval was not required because this meta-analysis involved synthesizing and analyzing previously published data without directly involving human subjects. All data were extracted from publicly available sources with no identifiable patient information, ensuring compliance with ethical guidelines.

Results

Eligible studies

The initial output of the literature search totaled 3000 articles. Following the removal of duplicates and scrutiny of titles and abstracts, a total of 20 relevant studies1534 were discerned, with 18 falling into the category of observational studies1526,2934 and 2 being RCTs.27,28 Only comparative studies were considered for inclusion in the ensuing meta-analysis. The PRISMA diagram in Figure 1 illustrates the meticulous search strategy employed. All RCTs were characterized by a comparative, multicenter, and double-blind design. Ten studies within the subset of 18 observational studies were prospective, while the remaining 8 were retrospective. The temporal scope of these publications spanned from 2013–2023.

CRM1583_Shadmani-f1.jpg

Figure 1: Preferred Reporting Items for Systematic Reviews and Meta-analyses flow diagram illustrating the search strategy and study-selection process for the meta-analysis. The initial search yielded 3000 articles, which were screened for duplicates and evaluated with regard to title and abstract. Twenty studies, including 2 randomized controlled trials and 18 observational studies, met the inclusion criteria for the analysis, with only comparative studies being included.

Baseline characteristics of the included patients

A total of 6310 participants were included in this meta-analysis, with 3198 individuals (50.68%) assigned to the WATCHMAN™ procedure group and 3112 individuals (49.32%) allocated to the ACP group. The total patient cohort included 3657 women (57.9%). The mean age of all participants was 75 ± 6.3 years. The average CHA2DS2-VASc score was 4.12 ± 1.3 points, while the average HAS-BLED score was 3.4 ± 1.0 points. Examination of the baseline data revealed a high prevalence of diabetes, hypertension, and a prior history of myocardial infarction among the included individuals. Paroxysmal AF was shown to be the most common kind of AF. Table 1 provides a comprehensive overview of the baseline characteristics of the patients enrolled in the study.

Table 1: Baseline Characteristics

CRM1583_Shadmani-t1.jpg

Quality assessment and publication bias

Trials of moderate-to-high quality were discerned through the application of the Cochrane method, which involves a comprehensive assessment of critical elements such as randomization, allocation concealment, blinding, and outcome reporting, employed for the scrutiny of RCTs, as depicted in Supplementary Figure S1. The assessment using the Newcastle–Ottawa scale unveiled a spectrum of quality within the observational studies included in the analysis, spanning from fair to good, as elucidated in Supplementary Table S2. The Newcastle–Ottawa scale is a systematic instrument for evaluating the quality of non-randomized studies—mainly observational studies—by considering selection, comparability, and outcome assessment parameters. The symmetry in the funnel plots instills confidence that the study outcomes remain unaffected by any undue influence of publication bias, as demonstrated in Supplementary Figure S2.

Supplementary Table S2: The Newcastle–Ottawa Scale

CRM1583_Shadmani-t3.jpg

CRM1583_Shadmani-f7.jpg

Supplementary Figure S1: Quality assessment of the included randomized controlled trials. The trials of medium to high quality were identified using the Cochrane method for assessing randomized controlled trials.

CRM1583_Shadmani-f8.jpg

Supplementary Figure S2: Funnel plots of primary outcomes. Funnel plots for (A) stroke, (B) systemic embolism, (C) all-cause mortality, (D) death from cardiogenic causes, and (E) major bleeding. Relative risk was used as an effect measure, and standard error was used as a measure of precision. The funnel plots indicated no evidence of publication bias.

Primary outcomes

The primary outcomes included stroke, systemic embolism, all-cause mortality, cardiac death, and major bleeding.

Stroke. Out of the 20 studies analyzed, 14 studies reported stroke events. The comprehensive analysis revealed that the WATCHMAN™ procedure is associated with a non-significantly increased risk of stroke compared to the ACP method (RR, 1.14; 95% confidence interval [CI], 0.80–1.62; I2 = 0%; P = .46), as depicted in Figure 2.

CRM1583_Shadmani-f2.jpg

Figure 2: Stroke risk comparison (WATCHMAN™ vs. Amplatzer™ Cardiac Plug). The figure demonstrates that the WATCHMAN™ procedure showed a non-significantly increased risk compared to the Amplatzer™ Cardiac Plug device. Abbreviations: CI, confidence interval; M–H, Mantel–Haenszel; RR, relative risk.

Systemic embolism. Data on systemic embolism occurrence were available in 16 of the 20 studies considered. The combined analysis indicated that the WATCHMAN™ procedure is linked to a non-significantly reduced risk of systemic embolism (RR, 0.69; 95% CI, 0.36–1.32; I2 = 0%; P = .26), as depicted in Figure 3.

CRM1583_Shadmani-f3.jpg

Figure 3: Systemic embolism risk (WATCHMAN™ vs. Amplatzer™ Cardiac Plug). The figure illustrates the combined analysis of 16 studies on systemic embolism occurrence. The analysis indicates that the WATCHMAN™ procedure is associated with a clinically reduced risk of systemic embolism, although statistical significance was not reached. Abbreviations: CI, confidence interval; M–H, Mantel–Haenszel; RR, relative risk.

All-cause mortality. Information regarding all-cause mortality was accessible in 16 of the 20 studies included in this study. The combined analysis suggested that the WATCHMAN™ procedure correlates with a non-significantly diminished risk of all-cause mortality (RR, 0.89; 95% CI, 0.73–1.08; I2 = 0%; P = .25), as illustrated in Figure 4.

CRM1583_Shadmani-f4.jpg

Figure 4: All-cause mortality comparison (WATCHMAN™ vs. Amplatzer™ Cardiac Plug). The figure depicts that the combined analysis suggests that employing the WATCHMAN™ procedure correlates with a clinically diminished risk of all-cause mortality, although statistical significance was not achieved. Abbreviations: CI, confidence interval; M–H, Mantel–Haenszel; RR, relative risk.

Cardiac death. Information regarding death from cardiogenic causes was accessible in 11 of the 20 studies evaluated. The comprehensive analysis showed that the WATCHMAN™ procedure is linked to a significantly lower risk of cardiac death compared to the ACP procedure (RR, 0.44; 95% CI, 0.20–0.95; I2 = 73; P = .04), as shown in Figure 5.

CRM1583_Shadmani-f5.jpg

Figure 5: Cardiac death risk (WATCHMAN™ vs. Amplatzer™ Cardiac Plug). The figure presents a comprehensive analysis revealing that the WATCHMAN™ procedure is associated with a significantly lower risk of death from cardiac causes than the Amplatzer™ Cardiac Plug device. Abbreviations: CI, confidence interval; M–H, Mantel–Haenszel; RR, relative risk.

Major bleeding. Data about significant bleeding events were available in 16 of the 20 included studies. The combined analysis indicated that the WATCHMAN™ procedure is associated with a non-significantly reduced risk of major bleeding compared to the ACP procedure (RR, 0.93; 95% CI, 0.65–1.33; I2 = 31%; P = .70), as illustrated in Figure 6.

CRM1583_Shadmani-f6.jpg

Figure 6: Major bleeding risk (WATCHMAN™ vs. Amplatzer™ Cardiac Plug). The figure depicts that the combined analysis suggests that the WATCHMAN™ procedure is associated with a clinically reduced risk of major bleeding compared to the Amplatzer™ Cardiac Plug device; however, statistical significance was not attained. Abbreviations: CI, confidence interval; M–H, Mantel–Haenszel; RR, relative risk.

Secondary outcomes

The secondary outcomes included device-related thrombus, PDL, procedure-related complications, and procedural success. Fourteen out of 20 studies provided information on device-related thrombus, and the pooled analysis indicated that the WATCHMAN™ procedure was associated with a non-significantly increased risk of device-related thrombus (RR, 1.38; 95% CI, 0.82–2.33; P = .22). Thirteen out of 20 studies reported data on PDL, revealing a non-significantly increased risk associated with the WATCHMAN™ procedure (RR, 1.52; 95% CI, 0.82–2.81; P = .18). Fifteen out of 20 studies reported data on procedure-related complications, demonstrating a non-significantly decreased risk with the WATCHMAN™ procedure (RR, 0.71; 95% CI, 0.47–1.08; P = .11). Nine out of 20 studies provided data regarding procedural success, showing no significant difference between the WATCHMAN™ procedure and ACP device groups (RR, 1.04; 95% CI, 0.96–1.13; P = .37) in the pooled analysis.

Discussion

This systematic review and meta-analysis of the WATCHMAN™ and ACP/Amulet™ devices for percutaneous LAAC consisting of 6310 patients from 20 different studies found numerous essential findings. First, the WATCHMAN™ device was associated with an increased stroke risk and a clinically reduced systemic embolism risk, but these differences were not statistically significant. However, the most recent research done by Tan et al.35 has yielded extensive insights into the feasibility and efficacy of the WATCHMAN™ device as a substitute for stroke prevention in specific patient populations. Unlike standard anticoagulant therapies for non-valvular AF, these studies suggest that the WATCHMAN™ device may reduce bleeding duration and, in most cases, carry a stroke risk that is either equivalent to or lower than that of conventional anticoagulants.36,37

Per our analysis, the WATCHMAN™ device reduced all-cause mortality, although without statistical significance. Moreover, it significantly lowered the risk of cardiac death compared to the ACP device. Despite a decrease in the likelihood of severe bleeding with WATCHMAN™ treatment, the association lacked statistical significance. Yerasi et al. conducted a study summarizing the Amulet™ multicenter registry findings, revealing a procedure fatality rate of 0.2%. The procedure fatality rate in the U.S. LARIAT multicenter registry was notably lower, at 0.006%.37 Noteworthy findings from the WATCHMAN™ Left Atrial Appendage System for Embolic Protection in Patients with Atrial Fibrillation (PROTECT-AF) trial indicated that severe bleeding, according to Bleeding Academic Research Consortium (BARC) criteria, occurred exclusively in the pericardial area, with a frequency of 4.8%. In the Amulet™ multicenter registry, significant bleeding occurred in 2.4% of patients, with pericardial bleeding accounting for half of these. It is worth noting that access site bleeding, while relatively infrequent across all devices, occurred in 0.3% of patients.37 Secondary findings indicated that the WATCHMAN™ approach could increase device-related thrombus and leaks near the device, although statistical significance was not observed. The WATCHMAN™ device showed a statistically insignificant decrease in the likelihood of complications connected to the surgery while maintaining a procedural success rate comparable to the ACP. The results of our study reveal subtle variations in the safety and effectiveness characteristics of the WATCHMAN™ and ACP/Amulet™ systems. The WATCHMAN™ device demonstrates certain therapeutic benefits. However, many of these findings still need to reach statistical significance.

Appendages with proximal lobes, large ostia, or a shallow depth may make complete sealing difficult and result in PDLs. Furthermore, it has been demonstrated that poor use of the LAAO device increases the risk of PDLs.38 It has been shown that individuals with PDLs are more likely to have recurrent thromboembolism or require continuing oral anticoagulant medication, which can diminish LAAO therapeutic efficacy. Physicians usually recommend oral anticoagulants to patients with severe PDLs until the leak size is reduced or a secondary surgery is performed to occlude, coil, or ablate the leak.39

Our research is subject to specific limitations. First, few studies provide information on neck diameter, fluoroscopy duration, residual flow, device size, LAA diameter, or procedural time that might be used to choose the suitable device for LAAO. Additionally, the diverse follow-up periods (ranging from 45 days to 4 years) between studies may have had different effects on the observed outcomes. More extended follow-up periods might have revealed cumulative or delayed effects. In contrast, shorter follow-up periods might have captured early effects.

Moreover, the results may be affected, and their applicability is limited by differences in baseline variables, including age, average CHA2DS2-VASc score, HAS-BLED score, comorbidities, and therapeutic interventions. Additionally, the high rates of cardiac death, major bleeding, and incomplete or inadequately reported data in some studies make it difficult to conduct a thorough analysis and draw reliable conclusions. These factors and differences in study designs, patient populations, and methodology among the included studies introduce heterogeneity in some outcomes.

Conclusion

Overall, the use of WATCHMAN™ devices has demonstrated a higher likelihood of stroke and a lower likelihood of systemic embolism, all-cause mortality, cardiac death, and major bleeding, respectively. However, these findings were not statistically significant and, therefore, cannot be clinically applied to determine whether WATCHMAN™ devices are superior or inferior to Amplatzer™ Amulet™ devices. The only statistically significant difference was observed in the risk of cardiac death, which may have clinical implications. Additional RCTs are necessary to comprehensively compare the two devices, facilitating better-informed decision-making for cardiologists in the future.

References

  1. Fahmy P, Spencer R, Tsang M, Gooderham P, Saw J. Left atrial appendage closure for atrial fibrillation is safe and effective after intracranial or intraocular hemorrhage. Can J Cardiol. 2016;32(3):349–354. [CrossRef] [PubMed]
  2. Kornej J, Börschel CS, Benjamin EJ, Schnabel RB. Epidemiology of atrial fibrillation in the 21st century: novel methods and new insights. Circ Res. 2020;127(1):4–20. [CrossRef] [PubMed]
  3. Martín Gutiérrez E, Castaño M, Gualis J, et al. Beneficial effect of left atrial appendage closure during cardiac surgery: a meta-analysis of 280 585 patients. Eur J Cardiothorac Surg. 2020;57(2):252–262. [CrossRef] [PubMed]
  4. Osmancik P, Herman D, Neuzil P, et al. 4-Year outcomes after left atrial appendage closure versus nonwarfarin oral anticoagulation for atrial fibrillation. J Am Coll Cardiol. 2022;79(1):1–14. [CrossRef] [PubMed]
  5. Turagam MK, Osmancik P, Neuzil P, Dukkipati SR, Reddy VY. Left atrial appendage closure versus oral anticoagulants in atrial fibrillation: a meta-analysis of randomized trials. J Am Coll Cardiol. 2020;76(23):2795–2797. [CrossRef] [PubMed]
  6. Della Rocca DG, Magnocavallo M, Gianni C, et al. Procedural and short-term follow-up outcomes of Amplatzer Amulet occluder versus Watchman FLX device: a meta-analysis. Heart Rhythm. 2022;19(6):1017–1018. [CrossRef] [PubMed]
  7. Qiao J, Zhang B, Wang J, et al. Comparison between Amplatzer and Watchman left atrial appendage closure devices for stroke prevention in atrial fibrillation: a systemic review and meta-analysis. Cardiology. 2022;147(3):290–297. [CrossRef] [PubMed]
  8. Basu Ray I, Khanra D, Shah S, et al. Meta-analysis comparing Watchman TM and Amplatzer devices for stroke prevention in atrial fibrillation. Front Cardiovasc Med. 2020;7:89. [CrossRef] [PubMed]
  9. Pietrasik GM, Huang HD, Rodriguez JM, Sharma PS, Trohman RG, Krishnan K. Safety and feasibility of radiofrequency redo pulmonary vein isolation ablation for atrial fibrillation after amulet implantation and device electrical characteristics. HeartRhythm Case Rep. 2020;6(7):415–418. [CrossRef] [PubMed]
  10. Lugtu IC, Cheng WH, Chang SL, Liu SH, Lo LW, Chen SA. Percutaneous left atrial appendage closure using an occluder device: a single center experience. Acta Cardiol Sin. 2022;38(3):352–361. [CrossRef] [PubMed]
  11. Page MJ, McKenzie JE, Bossuyt PM, et al. The PRISMA 2020 statement: an updated guideline for reporting systemic reviews. BMJ. 2021;372:n71. [CrossRef] [PubMed]
  12. Higgins JP, Altman DG, Gøtzsche PC, et al.; Cochrane Bias Methods Group; Cochrane Statistical Methods Group. The Cochrane Collaboration’s tool for assessing risk of bias in randomized trials. BMJ. 2011;343:d5928. [CrossRef] [PubMed]
  13. Wells GA, Shea B, O’Connell D, Peterson J, Welch V, Losos M. The Newcastle-Ottawa Scale (NOS) for assessing the quality of nonrandomized studies in meta-analyses. 2020. Available at: http://www.ohri.ca/programs/clinical_epidemiology/oxford.asp. Accessed January 10, 2023.
  14. Higgins JP, Thompson SG, Deeks JJ, Altman DG. Measuring inconsistency in meta-analyses. BMJ. 2003;327(7414):557–560. [CrossRef] [PubMed]
  15. Chun KR, Bordignon S, Urban V, et al. Left atrial appendage closure followed by 6 weeks of antithrombotic therapy: a prospective single-center experience. Heart Rhythm. 2013;10(12):1792–1799. [CrossRef] [PubMed]
  16. Cruz-Gonzalez I, Perez-Rivera A, Lopez-Jimenez R, et al. Significance of the learning curve in left atrial appendage occlusion with two different devices. Catheter Cardiovasc Interv. 2014;83(4):642–646. [CrossRef] [PubMed]
  17. Gafoor S, Franke J, Bertog S, et al. Left atrial appendage occlusion in octogenarians: short-term and 1-year follow-up. Catheter Cardiovasc Interv. 2014;83(5):805–810. [CrossRef] [PubMed]
  18. Kim JS, Lee H, Suh Y, et al. Left atrial appendage occlusion in non-valvular atrial fibrillation in a korean multi-center registry. Circ J. 2016;80(5):1123–1130. [CrossRef] [PubMed]
  19. Figini F, Mazzone P, Regazzoli D, et al. Left atrial appendage closure: a single center experience and comparison of two contemporary devices. Catheter Cardiovasc Interv. 2017;89(4):763–772. [CrossRef] [PubMed]
  20. Fastner C, Hoffmann L, Aboukoura M, et al. Real-world experience comparing two common left atrial appendage closure devices. BMC Cardiovasc Disord. 2018;18(1):171. [CrossRef] [PubMed]
  21. Chen S, Chun KRJ, Bordignon S, et al. Left atrial appendage occlusion using LAmbre Amulet and Watchman in atrial fibrillation. J Cardiol. 2019;73(4):299–306. [CrossRef] [PubMed]
  22. Cheung GS, So KC, Chan CK, et al. Comparison of three left atrial appendage occlusion devices for stroke prevention in patients with non-valvular atrial fibrillation: a single-centre seven-year experience with WATCHMAN, AMPLATZER cardiac Plug/Amulet, LAmbre: comparison of three LAAO devices for stroke prevention. AsiaIntervention. 2019;5(1):57–63. [CrossRef] [PubMed]
  23. Ledwoch J, Franke J, Akin I, et al. WATCHMAN versus ACP or Amulet devices for left atrial appendage occlusion: a sub-analysis of the multicentre LAARGE registry. EuroIntervention. 2020;16(11):e942–e949. [CrossRef] [PubMed]
  24. Kleinecke C, Yu J, Neef P, et al. Clinical outcomes of Watchman vs. Amplatzer occluders for left atrial appendage closure (WATCH at LAAC). Europace. 2020;22(6):916–923. [CrossRef] [PubMed]
  25. Davtyan K, Simonyan G, Topchyan A, et al. Comparative safety and efficacy of left atrial appendage occlusion with the Watchman device and Amplatzer cardiac plug: results of the Russian National Registry. Biomed Res Int. 2020;2020:2352648. [CrossRef] [PubMed]
  26. Chiu FC, Huang PS, Chen JJ, Hwang JJ, Tsai CT. Long-term outcomes of percutaneous left atrial appendage closure for the prevention of stroke in patients with atrial fibrillation: Asia-Pacific experience. J Formos Med Assoc. 2022;121(8):1442–1449. [CrossRef] [PubMed]
  27. Lakkireddy D, Thaler D, Ellis CR, et al. Amplatzer Amulet left atrial appendage occluder versus Watchman device for stroke prophylaxis (Amulet IDE): a randomized, controlled trial. Circulation. 2021;144(19):1543–1552. [CrossRef] [PubMed]
  28. Mansour MJ, Harnay E, Al Ayouby A, et al. One year outcome and analysis of peri-device leak of left atrial appendage occlusion devices. J Interv Card Electrophysiol. 2022;64(1):27–34. [CrossRef] [PubMed]
  29. Saad M, Risha O, Sano M, et al. Comparison between Amulet and Watchman left atrial appendage closure devices: a real-world, single center experience. Int J Cardiol Heart Vasc. 2021;37:100893. [CrossRef] [PubMed]
  30. Radinovic A, Falasconi G, Marzi A, et al. Long-term outcome of left atrial appendage occlusion with multiple devices. Int J Cardiol. 2021;344:66–72. [CrossRef] [PubMed]
  31. Teiger E, Thambo JB, Defaye P, et al.; French National Left Atrial Appendage Closure Registry (FLAAC) Investigators. Left atrial appendage closure for stroke prevention in atrial fibrillation: final report from the French left atrial appendage closure registry. Catheter Cardiovasc Interv. 2021;98(4):788–799. [CrossRef] [PubMed]
  32. Galea R, Mahmoudi K, Gräni C, et al. Watchman FLX vs. Watchman 2.5 in a dual-center left atrial appendage closure cohort: the WATCH-DUAL study. Europace. 2022;24(9):1441–1450. [CrossRef] [PubMed]
  33. Kretzler L, Mues C, Wunderlich C, et al. Short term outcome after left atrial appendage occlusion with the AMPLATZER Amulet and WATCHMAN device: results from the ORIGINAL registry (saxOnian RegIstry analyzinG and followINg left atrial Appendage cLosure). BMC Cardiovasc Disord. 2022;22(1):271. [CrossRef] [PubMed]
  34. Korsholm K, Kramer A, Andersen A, et al. Left atrial appendage sealing performance of the Amplatzer Amulet and Watchman FLX device. J Interv Card Electrophysiol. 2023;66(2):391–401 [Erratum in: J Interv Card Electrophysiol. 2022]. [CrossRef] [PubMed]
  35. Tan BE, Wong PY, Lee JZ, Tan NY, Rao M, Cheung JW. Direct oral anticoagulant versus warfarin after left atrial appendage closure with WATCHMAN: updated systemic review and meta-analysis. Curr Probl Cardiol. 2022;47(11):101335. [CrossRef] [PubMed]
  36. Kleinecke C, Cheikh-Ibrahim M, Schnupp S, et al. Long-term clinical outcomes of Amplatzer cardiac plug versus Amulet occluders for left atrial appendage closure. Catheter Cardiovasc Interv. 2020;96(3):E324–E331. [CrossRef] [PubMed]
  37. Yerasi C, Lazkani M, Kolluru P, et al. An updated Systemic review and meta-analysis of early outcomes after left atrial appendage occlusion. J Interv Cardiol. 2018;31(2):$197–206. [CrossRef] [PubMed]
  38. Lakkireddy D, Nielsen-Kudsk JE, Windecker S, et al. Mechanisms, predictors, and evolution of severe peri-device leaks with two different left atrial appendage occluders. Europace. 2023;25(9):euad237. [CrossRef] [PubMed]
  39. Saw J, Tzikas A, Shakir S, et al. Incidence and clinical impact of device-associated thrombus and peri-device leak following left atrial appendage closure with the amplatzer cardiac plug. JACC Cardiovasc Interv. 2017;10(4):391–399. [CrossRef] [PubMed]