Journal of Innovation in Cardiac Rhythm Management
Articles Articles 2024 November 2024 - Volume 15 Issue 11

Failure of Ventricular Fibrillation Sensing During Subcutaneous Implantable Defibrillator Testing: A Twitchy Situation

DOI: 10.19102/icrm.2024.15112

MUSSA SAAD, MD,1 DEEPTI RANGANATHAN, MBCHB,1 and SHELDON M. SINGH, MD1,2

1Sunnybrook Health Sciences Centre, Toronto, Ontario, Canada

2Department of Medicine, Temerty Faculty of Medicine, University of Toronto, Toronto, Ontario, Canada

PDF Download PDF
tweeter Follow Us >>

ABSTRACT.Oversensing of extra-cardiac noise may inhibit delivery of subcutaneous cardiac implantable defibrillator (S-ICD) therapy. We report a case of diaphragmatic tetany resulting in the inhibition of S-ICD therapy at the time of defibrillator testing without the use of muscle relaxants. Clinicians should be aware of this phenomenon.

KEYWORDS.Defibrillation threshold testing, diaphragmatic tetany, subcutaneous implantable cardioverter-defibrillator, undersensing.

The authors report no conflicts of interest for the published content. No funding information was provided.
Manuscript received March 26, 2024. Final version accepted June 10, 2024.
Address correspondence to: Sheldon Singh, MD, 2075 Bayview Ave, Toronto, Ontario, M4N 3M5, Canada. Email: sheldon.singh@sunnybrook.ca.

Case presentation

A 66-year-old man was admitted to our center following a syncopal event during which he sustained a C5 spinal cord injury. During his admission, he was referred for a subcutaneous cardiac implantable defibrillator (S-ICD) generator replacement. As such, we elected for conscious sedation during the S-ICD replacement procedure to avoid endotracheal intubation in the setting of the patient’s unstable cervical spine fracture.

As per guideline recommendations, defibrillation threshold testing was performed.1 Prior to ventricular fibrillation (VF) induction, we flushed with saline to expel air from the incision sites, closed both incisions, and ensured that we tested for noise with massaging of the lead/generator. Using the Vector Select Algorithm (Boston Scientific, Marlborough, MA, USA), the device was programmed with a primary sensing vector (Figure 1A). VF induction was performed with a 200-mA burst at 50 Hz for 5 s. VF was induced but not sensed appropriately by the device (Figure 1B). After 30 s without delivery of therapy, an external shock was delivered. A review of the associated electrogram suggested inhibition of sensing due to the presence of noise. At this point, the sensing vector was manually programmed to the secondary vector, and VF induction was repeated. With this sensing configuration, noise was not detected (Figure 1C). Sensing of VF was appropriate, and the therapy was committed 8 s into the episode and delivered 8 s later.

CRM1614_Saad-f1.jpg

Figure 1: A: Location of the subcutaneous cardiac implantable defibrillator coil and can. The primary (red) and secondary (yellow) sensing vectors are shown. The left diaphragm is indicated in white and in the region of the primary sensing vector. B: Ventricular fibrillation (VF) induction in the primary sensing mode. Noise related to muscle tetany noted with an N preventing appropriate sensing indicated with a T. Criteria for the detection of VF not achieved at 36 s. C: VF induction in the secondary sensing mode. Appropriate T sensing with a charge committed at 8 s.

Discussion

This case highlights the phenomenon of diaphragmatic tetany during VF induction with an S-ICD system. In the absence of the use of muscle relaxants, VF induction with a 50-Hz burst can stimulate adjacent extra-cardiac muscles—in this case, the diaphragm—resulting in muscular tetany. If the selected S-ICD sensing vector encompasses this muscle group, then the muscle activity can be sensed and interpreted as noise, thereby resulting in the therapy being withheld. Non-cardiac muscular tetany is unlikely to occur when muscle relaxants are used or when ventricular arrhythmia is induced with programmed electrical stimulation using a transvenous electrophysiology catheter.2 Furthermore, this phenomenon would not be expected to occur in the setting of spontaneous ventricular arrhythmia.

With the increasing use of S-ICD and the move away from the use of general anesthesia for S-ICD implantation, clinicians need to be aware of this phenomenon to avoid unnecessary repetitive VF induction or lead revision.

References

  1. Wilkoff BL, Fauchier L, Stiles MK, et al. 2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing. Europace. 2016;18(2):159–183. [CrossRef] [PubMed]
  2. Baalman SWE, Kooiman KM, Quast ABE, Postema PG, Donnelley S, Knops RE. Noise caused by involuntary muscle response may inhibit detection of ventricular fibrillation during defibrillation testing. HeartRhythm Case Rep. 2019;6(3):126–130. [CrossRef] [PubMed]
Banner BSC 09 2024 Resonate JICRM 160x600